Reduction of Radiation Exposure in Atrial Fibrillation Ablation Using a New Image Integration Module: A Prospective Randomized Trial in Patients Undergoing Pulmonary Vein Isolation

被引:17
作者
Akbulak, Ruken Oezge [1 ]
Schaeffer, Benjamin [1 ]
Jularic, Mario [1 ]
Moser, Julia [1 ]
Schreiber, Doreen [1 ]
Salzbrunn, Tim [1 ]
Meyer, Christian [1 ]
Eickholt, Christian [1 ]
Kuklik, Pawel [1 ]
Hoffmann, Boris A. [1 ]
Willems, Stephan [1 ]
机构
[1] Univ Hosp Hamburg Eppendorf, Univ Heart Ctr, Dept Cardiol Electrophysiol, D-20246 Hamburg, Germany
关键词
atrial fibrillation; catheter ablation; fluoroscopy; IIM-CartoUnivu; image integration; X-ray; 3D-mapping system; RADIOFREQUENCY CATHETER ABLATION; FLUOROSCOPY; EXPERIENCE;
D O I
10.1111/jce.12673
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Reduced Radiation Exposure in AF Ablation IntroductionRecently, a new image integration module (IIM, CartoUnivu Module) has been introduced to combine and merge fluoroscopy images with 3-dimensional-(3D)-electroanatomical maps (Carto (R) 3 System) into an accurate 3D view. The aim of the study was to investigate the influence of IIM on the fluoroscopy exposure during pulmonary vein isolation (PVI) for paroxysmal atrial fibrillation (PAF) in a prospective randomized trial. Methods and resultsBetween June and November 2014, a total of 60 patients with PAF (73.3% male, 64.0 9.2 years), who underwent PVI with the endpoint of unexcitability of the ablation line, were randomized to either a conventional 3D mapping system (Carto (R) 3 System) or to an additional IIM on the basis of an assumed reduction of fluoroscopy exposure by the use of IIM. There were no significant differences in baseline characteristics. The median ablation procedure time was identical in both groups (140.7 +/- 27.8 minutes vs. 140.8 +/- 39.5 minutes; P = 0.851). A significant decrease of mean fluoroscopy time from 11.9 +/- 2.1 to 7.4 +/- 2.6 minutes (P < 0.0006) and median fluoroscopy dose from 882.9 to 476.5 cGycm(2) (P < 0.001) was achieved. The main reduction of radiation could be realized during creation of the 3D-map. No major complications occurred during the procedures. After a median follow-up of 125.7 +/- 45.6 days 80% of the patients were free from any atrial arrhythmias. ConclusionCartoUnivu module easily integrates into the workflow of PVI with the endpoint of unexcitability of the ablation line without prolonging the procedure time. It is associated with a marked reduction in fluoroscopic dose when compared to a conventional 3D mapping system.
引用
收藏
页码:747 / 753
页数:7
相关论文
共 21 条
  • [1] ESC-EURObservational Research Programme: the Atrial Fibrillation Ablation Pilot Study, conducted by the European Heart Rhythm Association
    Arbelo, Elena
    Brugada, Josep
    Hindricks, Gerhard
    Maggioni, Aldo
    Tavazzi, Luigi
    Vardas, Panos
    Anselme, Frederic
    Inama, Giuseppe
    Jais, Pierre
    Kalarus, Zbigniew
    Kautzner, Josef
    Lewalter, Thorsten
    Mairesse, Georges
    Perez-Villacastin, Julian
    Riahi, Sam
    Taborsky, Milos
    Theodorakis, George
    Trines, Serge
    [J]. EUROPACE, 2012, 14 (08): : 1094 - 1103
  • [2] Camm AJ, 2012, EUR HEART J, V33
  • [3] Camm AJ, 2010, EUROPACE, V12, P1360, DOI [10.1093/europace/euq350, 10.1093/eurheartj/ehq278]
  • [4] Electrical isolation of pulmonary veins in patients with atrial fibrillation:: reduction of fluoroscopy exposure and procedure duration by the use of a non-fluoroscopic navigation system (NavX®)
    Estner, Heidi Luise
    Deisenhofer, Isabel
    Luik, Armin
    Ndrepepa, Gjin
    von Bary, Christian
    Zrenner, Bernhard
    Schmitt, Claus
    [J]. EUROPACE, 2006, 8 (08): : 583 - 587
  • [5] European Commission, 2008, RAD PROT, V154
  • [6] Flugelman Moshe Y, 2008, EuroIntervention, V4, P158, DOI 10.4244/EIJV4I1A25
  • [7] Ionizing Radiation in Cardiac Imaging A Science Advisory From the American Heart Association Committee on Cardiac Imaging of the Council on Clinical Cardiology and Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and Intervention
    Gerber, Thomas C.
    Carr, J. Jeffrey
    Arai, Andrew E.
    Dixon, Robert L.
    Ferrari, Victor A.
    Gomes, Antoinette S.
    Heller, Gary V.
    McCollough, Cynthia H.
    McNitt-Gray, Michael F.
    Mettler, Fred A.
    Mieres, Jennifer H.
    Morin, Richard L.
    Yester, Michael V.
    [J]. CIRCULATION, 2009, 119 (07) : 1056 - 1065
  • [8] Harold JG, 2014, J AM COLL CARDIOL, V63, pE57, DOI [10.1016/j.jacc.2014.02.537, 10.1016/j.jacc.2014.02.536, 10.1016/j.jtcvs.2014.05.014]
  • [9] Practical ways to reduce radiation dose for patients and staff during device implantations and electrophysiological procedures
    Heidbuchel, Hein
    Wittkampf, Fred H. M.
    Vano, Eliseo
    Ernst, Sabine
    Schilling, Richard
    Picano, Eugenio
    Mont, Lluis
    [J]. EUROPACE, 2014, 16 (07): : 946 - 964
  • [10] Impact of Structural Heart Disease on the Acute Complication Rate in Atrial Fibrillation Ablation: Results from the German Ablation Registry
    Hoffmann, Boris A.
    Kuck, Karl-Heinz
    Andresen, Dietrich
    Spitzer, Stefan G.
    Hoffmann, Ellen
    Schumacher, Burghard
    Eckardt, Lars
    Brachmann, Johannes
    Becker, Ruediger
    Steven, Daniel
    Rostock, Thomas
    Juenger, Claus
    Senges, Jochen
    Willems, Stephan
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2014, 25 (03) : 242 - 249