Speech Understanding and Sound Localization with a New Nonimplantable Wearing Option for Baha

被引:19
作者
Gawliczek, Tom [1 ]
Wimmer, Wilhelm [1 ,2 ]
Munzinger, Fabio [2 ]
Caversaccio, Marco [1 ,2 ]
Kompis, Martin [2 ]
机构
[1] Univ Bern, ARTORG Ctr Biomed Engn Res, Hearing Res Lab, CH-3008 Bern, Switzerland
[2] Univ Bern, Inselspital, Dept ENT Head & Neck Surg, CH-3010 Bern, Switzerland
关键词
ANCHORED HEARING-AIDS; BONE-CONDUCTION IMPLANT; DEVICES; DEAFNESS; SYSTEMS; STIMULATION; CHILDREN; ADULTS; USERS; CROS;
D O I
10.1155/2018/5264124
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Objective. To measure the audiological benefit of the Baha SoundArc, a recently introduced nonimplantable wearing option for bone conduction sound processor, and to compare it with the known softband wearing option in subjects with normal cochlear function and a purely conductive bilateral hearing loss. Methods. Both ears of 15 normal hearing subjects were occluded for the time of the measurement, yielding an average unaided threshold of 49 dB HL (0.5 - 4 kHz). Soundfield thresholds, speech understanding in quiet and in noise, and sound localization were measured in unaided conditions and with 1 or 2 Baha 5 sound processors mounted on either a softband or a SoundArc device. Results. Soundfield thresholds and speech reception thresholds were improved by 19.5 to 24.8 dB (p<.001), when compared to the unaided condition. Speech reception thresholds in noise were improved by 3.7 to 4.7 dB (p<.001). Using 2 sound processors rather than one improved speech understanding in noise for speech from the direction of the 2nd device and sound localization error by 23 degrees to 28 degrees. No statistically significant difference was found between the SoundArc and the softband wearing options in any of the tests. Conclusions. Bone conduction sound processor mounted on a SoundArc or on a softband resulted in considerable improvements in hearing and speech understanding in subjects with a simulated, purely conductive, and bilateral hearing loss. No significant difference between the 2 wearing options was found. Using 2 sound processors improves sound localization and speech understanding in noise in certain spatial settings.
引用
收藏
页数:8
相关论文
共 28 条
[1]  
Bosman AJ, 2001, AUDIOLOGY, V40, P158
[2]   Predisposing factors for adverse skin reactions with percutaneous bone anchored hearing devices implanted with skin reduction techniques [J].
Candreia, Claudia ;
Birrer, Ruth ;
Fistarol, Susanna ;
Kompis, Martin ;
Caversaccio, Marco ;
Arnold, Andreas ;
Stieger, Christof .
EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2016, 273 (12) :4185-4192
[3]  
Dun CAJ, 2011, ADV OTO-RHINO-LARYNG, V71, P22, DOI 10.1159/000323577
[4]   Bone conduction implants - a rapidly developing field [J].
Edmiston, R. C. ;
Aggarwal, R. ;
Green, K. M. J. .
JOURNAL OF LARYNGOLOGY AND OTOLOGY, 2015, 129 (10) :936-940
[5]   Intracochlear Measurements of Interaural Time and Level Differences Conveyed by Bilateral Bone Conduction Systems [J].
Farrell, Nyssa F. ;
Hartl, Renee M. Banakis ;
Benichoux, Victor ;
Brown, Andrew D. ;
Cass, Stephen P. ;
Tollin, Daniel J. .
OTOLOGY & NEUROTOLOGY, 2017, 38 (10) :1476-1483
[6]   Unilateral and Bilateral Audiological Benefit With an Adhesively Attached, Noninvasive Bone Conduction Hearing System [J].
Gawliczek, Tom ;
Munzinger, Fabio ;
Anschuetz, Lukas ;
Caversaccio, Marco ;
Kompis, Martin ;
Wimmer, Wilhelm .
OTOLOGY & NEUROTOLOGY, 2018, 39 (08) :1025-1030
[7]   THE BONE-ANCHORED HEARING-AID - PRINCIPAL DESIGN AND A PSYCHOACOUSTICAL EVALUATION [J].
HAKANSSON, B ;
TJELLSTROM, A ;
ROSENHALL, U ;
CARLSSON, P .
ACTA OTO-LARYNGOLOGICA, 1985, 100 (3-4) :229-239
[8]  
Hakansson B., 1990, ANN OTOL RHINOL LA S, V151, P1
[9]   Pilot study on the effectiveness of the conventional CROS, the transcranial CROS and the BAHA transcranial CROS in adults with unilateral inner ear deafness [J].
Hol, Myrthe K. S. ;
Kunst, Sylvia J. W. ;
Snik, Ad F. M. ;
Cremers, Cor W. R. J. .
EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2010, 267 (06) :889-896
[10]  
HOLM S, 1979, SCAND J STAT, V6, P65