Sonablate (R) 500 (SB-500) HIFU devices have been successfully used to treat prostate cancer non-invasively. In addition, Visually Directed HIFU with the SB-500 has demonstrated higher efficacy. Visually Directed HIFU works by displaying hyperechoic changes on the B-mode ultrasound images. However, small changes in the grey-scale images are not detectable by Visually Directed HIFU. To detect all tissue changes reliably, the SB-500 was enhanced with quantitative, real-time Tissue Change Monitoring (TCM) software. TCM uses pulse-echo ultrasound backscattered RF signals in 2D to estimate changes in the tissue properties caused by HIFU. The RF signal energy difference is calculated in selected frequency bands (pre and post HIFU) for each treatment site. The results are overlaid on the real-time ultrasound image in green, yellow and orange to represent low, medium and high degree of change in backscattered energy levels. The color mapping scheme was derived on measured temperature and backscattered RF signals from in vitro chicken tissue experiments. The TCM software was installed and tested in a clinical device to obtain human RF data. Post HIFU contrast enhanced MRI scans verified necrotic regions of the prostate. The color mapping success rate at higher HIFU power levels was 94% in the initial clinical test. Based on these results, TCM software has been released for wider usage. The clinical studies with TCM in Japan and The Bahamas have provided the following PSA (ng/ml) results. Japan (n=97), PSA pre-treatment/post-treatment; minimum 0.7/0.0, maximum 76.0/4.73, median 6.89/0.07, standard deviation 11.19/0.62. The Bahamas (n=59), minimum 0.4/0.0, maximum 13.0/1.4, median 4.7/0.1, standard deviation 2.8/0.3.