Accuracy of SpaceLabs 90207 is altered by venous blood redistribution

被引:8
作者
Czarkowski, M
Mikulska, M
Baran, A
Zebrowski, M
Rózanowski, K
机构
[1] Med Univ Warsaw, Chair Internal Med & Endocrinol, PL-02097 Warsaw, Poland
[2] Med Univ Warsaw, Dept Internal Med & Endocrinol, SSA, PL-02097 Warsaw, Poland
[3] Mil Inst Aerosp Med, Dept Physiol, Warsaw, Poland
关键词
automated blood pressure monitor; blood pressure measurement; lower body negative pressure (LBNP); oscillometric method; SpaceLabs; 90207;
D O I
10.1080/08037050310022414
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Validation protocols that have been introduced for automated blood pressure (BP) measuring devices recommend procedures carried out at rest only. We aimed to determine whether venous blood redistribution affects the accuracy of the oscillometric method of BP measurement. For this purpose, we chose a popular oscillometric ambulatory BP monitor -- the SpaceLabs 90207 (Osc). Lower body negative pressure (-40 mmHg) (LBNP) was used to simulate changes of body position. Fifty-one young healthy volunteers had their BP measured simultaneously by Osc and mercury sphygmomanometer (HgS) at rest (min 3 and 5), during LBNP (min 7 and 9) and after LBNP (min 11). Differences (Delta, mmHg) between HgS and Osc for systolic (SBP) and diastolic BP (DBP) were calculated for every measurement minute. For SBP, DeltaSBP-7 and DeltaSBP-9 were significantly different from DeltaSBP-5 (0.65 +/- 2.6 and 0.33 +/- 2.4 mmHg vs -0.80 +/- 2.9 mmHg, p < 0.003 and p < 0.02, respectively). DeltaSBP-11 also differed significantly from DeltaSBP-3 and DeltaSBP-5 (1.16 +/- 2.5 mmHg vs -0.06 +/- 3.1 mmHg and -0.80 +/- 2.9, p < 0.01 and p < 0.00007, respectively). When graded according to British Hypertension Society protocol, Osc remained in the A class in every measurement minute. The accuracy of SpaceLabs 90207 is altered by venous blood redistribution. These inaccuracies may constitute an additional limitation of oscillometric ambulatory BP monitoring.
引用
收藏
页码:334 / 339
页数:6
相关论文
共 22 条
[1]   CAROTID AND CARDIOPULMONARY BARORECEPTOR CONTROL OF SPLANCHNIC AND FOREARM VASCULAR-RESISTANCE DURING VENOUS POOLING IN MAN [J].
ABBOUD, FM ;
ECKBERG, DL ;
JOHANNSEN, UJ ;
MARK, AL .
JOURNAL OF PHYSIOLOGY-LONDON, 1979, 286 (JAN) :173-184
[2]  
[Anonymous], 1987, AM NAT STAND EL AUT
[3]   The clinical value of ambulatory blood pressure monitoring [J].
Gibbs, CR ;
Murray, S ;
Beevers, DG .
HEART, 1998, 79 (02) :115-117
[4]  
GOULD BA, 1985, BRIT HEART J, V53, P611
[5]  
Iqbal P., 1996, Blood Pressure, V5, P222, DOI 10.3109/08037059609079675
[6]   HUMAN SPLANCHNIC AND FOREARM VASOCONSTRICTOR RESPONSES TO REDUCTIONS OF RIGHT ATRIAL AND AORTIC PRESSURES [J].
JOHNSON, JM ;
ROWELL, LB ;
NIEDERBERGER, M ;
EISMAN, MM .
CIRCULATION RESEARCH, 1974, 34 (04) :515-524
[7]   COMMENTARY ON THE REVISED BRITISH-HYPERTENSION-SOCIETY PROTOCOL FOR EVALUATION OF BLOOD-PRESSURE MEASURING DEVICES - A CRITIQUE OF ASPECTS RELATED TO 24-HOUR AMBULATORY BLOOD-PRESSURE MEASUREMENT [J].
MANCIA, G ;
PARATI, G .
JOURNAL OF HYPERTENSION, 1993, 11 (06) :595-597
[8]   Evidence based treatment of hypertension - Measurement of blood presssure: an evidence based review [J].
McAlister, FA ;
Straus, SE .
BMJ-BRITISH MEDICAL JOURNAL, 2001, 322 (7291) :908-911
[9]  
MUSGRAVE FS, 1969, AEROSPACE MED, V40, P602
[10]   Use and interpretation of ambulatory blood pressure monitoring: recommendations of the British Hypertension Society [J].
O'Brien, E ;
Coats, A ;
Owens, P ;
Petrie, J ;
Padfield, PL ;
Littler, WA ;
de Swiet, M ;
Mee, F .
BMJ-BRITISH MEDICAL JOURNAL, 2000, 320 (7242) :1128-1134