Electromagnetic effects on the biological tissue surrounding a transcutaneous transformer for an artificial anal sphincter system

被引:1
作者
Zan, Peng [1 ]
Yang, Bang-hua [1 ]
Shao, Yong [1 ]
Yan, Guo-zheng [2 ]
Liu, Hua [2 ]
机构
[1] Shanghai Univ, Coll Mechatron Engn & Automat, Shanghai Key Lab Power Stn Automat Technol, Shanghai 200072, Peoples R China
[2] Shanghai Jiao Tong Univ, Sch Elect Informat & Elect Engn, Shanghai 200240, Peoples R China
基金
中国国家自然科学基金;
关键词
Artificial anal sphincter; Transcutaneous energy transmission; Current density; Specific absorption rate; CONDUCTIVITY; ABSORPTION; MODEL;
D O I
10.1631/jzus.B1000058
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
This paper reports on the electromagnetic effects on the biological tissue surrounding a transcutaneous transformer for an artificial anal sphincter. The coupling coils and human tissues, including the skin, fat, muscle, liver, and blood, were considered. Specific absorption rate (SAR) and current density were analyzed by a finite-length solenoid model. First, SAR and current density as a function of frequency (10-10(7) Hz) for an emission current of 1.5 A were calculated under different tissue thickness. Then relations between SAR, current density, and five types of tissues under each frequency were deduced. As a result, both the SAR and current density were below the basic restrictions of the International Commission on Non-Ionizing Radiation Protection (ICNIRP). The results show that the analysis of these data is very important for developing the artificial anal sphincter system.
引用
收藏
页码:931 / 936
页数:6
相关论文
共 18 条
[1]  
Ahlbom A, 1998, HEALTH PHYS, V74, P494
[2]   Acoustic emission during DC operations of the ITER Central Solenoid model coil [J].
Arai, K ;
Ninomiya, A ;
Ishigohka, T ;
Takano, K ;
Matsui, K ;
Michael, PC ;
Vieira, RF ;
Martovetsky, NN ;
Kaiho, K ;
Nakajima, H ;
Takahashi, Y ;
Kato, T ;
Ando, T ;
Tsuji, H ;
Okuno, K .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2002, 12 (01) :504-507
[3]  
Cristina S., 2008, Institution of Engineering and Technology 7th International Conference on Computation in Electromagnetics, CEM 2008, P208, DOI 10.1049/cp:20080264
[4]   A novel artificial sphincter prosthesis driven by a four-membrane silicon micropump [J].
Doll, Alexander F. ;
Wischke, Martin ;
Geipel, Andreas ;
Goldschmidtboeing, Frank ;
Ruthmann, Olaf ;
Hopt, Ulrich T. ;
Schrag, Hans-Juergen ;
Woias, Peter .
SENSORS AND ACTUATORS A-PHYSICAL, 2007, 139 (1-2) :203-209
[5]   Outcome after implantation of a novel prosthetic anal sphincter in humans [J].
Finlay, IG ;
Richardson, W ;
Hajivassiliou, CA .
BRITISH JOURNAL OF SURGERY, 2004, 91 (11) :1485-1492
[6]   Electrical conductivity of tissue at frequencies below 1 MHz [J].
Gabriel, C. ;
Peyman, A. ;
Grant, E. H. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (16) :4863-4878
[7]   The dielectric properties of biological tissues .3. Parametric models for the dielectric spectrum of tissues [J].
Gabriel, S ;
Lau, RW ;
Gabriel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2271-2293
[8]  
Guralnick ML, 2003, ATLAS UROL CLIN, V11, P83, DOI DOI 10.1016/S1063-5777(02)00067-1
[9]  
ICNIRP, 2009, EXP HIGH FREQ EL FIE
[10]   Temperature control of SMA artificial anal sphincter [J].
Kakubari, Y ;
Sato, F ;
Matsuki, H ;
Sato, T ;
Luo, Y ;
Takagi, T ;
Yambe, T ;
Nitta, SI .
IEEE TRANSACTIONS ON MAGNETICS, 2003, 39 (05) :3384-3386