摘要
The purpose of this study was to evaluate the corrosion resistance of experimental dental Ti-Ag alloys in 0.9% NaCl solution. Open-circuit potential (OCP) measurement and elution tests of the alloys with 5-30% Ag were performed. The amounts of both Ti and Ag ions released from the alloys with Ag <= 20% were below the detection limit. A very small amount of Ti ions was released from some of the 22.5% Ag specimens and some of the 25% Ag specimens. The time for the Ti-Ag alloys with 5-25% Ag to become the stable potential was earlier than that for titanium, and the OCP of the alloys was higher than that of titanium. These results indicated that the corrosion resistance of the alloys with 5-25% Ag was equivalent to that of titanium. On the other hand, it was suggested that the precipitation of TiAg deteriorated the corrosion resistance of Ti-Ag alloys because TiAg dissolved preferentially in the NaCl solution. Ti-Ag alloys with 5-25% Ag can be used not only as dental restorative materials but also as dental implant materials.
引用
收藏
页码:367 / 371
页数:5
相关论文
共 50 条
  • [1] Corrosion Resistance of Dental Ti-Ag Alloys in NaCl Solution
    Takahashi, Masatoshi
    Kikuchi, Masafumi
    Takada, Yukyo
    Okuno, Osamu
    MATERIALS TRANSACTIONS, 2010, 51 (04) : 762 - 766
  • [2] Corrosion behavior of Ti-Ag alloys used in dentistry in lactic acid solution
    Masatoshi Takahashi
    Masafumi Kikuchi
    Yukyo Takada
    Metals and Materials International, 2011, 17 : 175 - 179
  • [3] Corrosion Behavior of Ti-Ag Alloys Used in Dentistry in Lactic Acid Solution
    Takahashi, Masatoshi
    Kikuchi, Masafumi
    Takada, Yukyo
    METALS AND MATERIALS INTERNATIONAL, 2011, 17 (01) : 175 - 179
  • [4] Grindability of dental cast Ti-Ag and Ti-Cu alloys
    Kikuchi, M
    Takahashi, M
    Okabe, T
    Okuno, O
    DENTAL MATERIALS JOURNAL, 2003, 22 (02) : 191 - 205
  • [5] Properties of Ti-Ag alloys for biomedical and dental applications.
    Shim, HM
    Oh, KT
    Kim, KN
    JOURNAL OF DENTAL RESEARCH, 2002, 81 : A262 - A262
  • [6] Effect of Ag on the corrosion behavior of Ti-Ag alloys in artificial saliva solutions
    Zhang, B. B.
    Zheng, Y. F.
    Liu, Y.
    DENTAL MATERIALS, 2009, 25 (05) : 672 - 677
  • [7] Mechanical properties and microstructures of dental cast Ti-Ag and Ti-Cu alloys
    Takahashi, M
    Kikuchi, M
    Takada, Y
    Okuno, O
    DENTAL MATERIALS JOURNAL, 2002, 21 (03) : 270 - 280
  • [8] Machinability of experimental Ti-Ag alloys
    Kikuchi, Masafumi
    Takahashi, Masatoshi
    Okuno, Osamu
    DENTAL MATERIALS JOURNAL, 2008, 27 (02) : 216 - 220
  • [9] Mechanical properties of dental Ti-Ag alloys with 223, 25, 273, and 30 mass% Ag
    Takahashi, Masatoshi
    Kikuchi, Masafumi
    Takada, Yukyo
    DENTAL MATERIALS JOURNAL, 2015, 34 (04) : 503 - 507
  • [10] Electrochemical behavior of cast Ti-Ag alloys
    Takahashi, Masatoshi
    Kikuchi, Masafumi
    Takada, Yukyo
    Okabe, Toru
    Okuno, Osamu
    DENTAL MATERIALS JOURNAL, 2006, 25 (03) : 516 - 523