Dynamic reaction cell ICP-MS for determination of total As, Cr, Se and V in complex matrices: Still a challenge? A review

被引:119
作者
D'Ilio, Sonia [1 ]
Violante, Nicola [1 ]
Majorani, Costanza [1 ]
Petrucci, Francesco [1 ]
机构
[1] Ist Super Sanita, Dept Environm & Prevent, I-00161 Rome, Italy
关键词
DRC-ICP-MS; Interferences suppression; Reaction gases; Trace elements; PLASMA-MASS SPECTROMETRY; ROOM-TEMPERATURE KINETICS; ATOMIC TRANSITION-METAL; BANDPASS REACTION CELL; GAS-PHASE REACTIONS; MAIN-GROUP CATIONS; OXIDE-BASED INTERFERENCES; TRACE-ELEMENTS; DRC-MS; MULTIELEMENT ANALYSIS;
D O I
10.1016/j.aca.2011.04.052
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Mass interferences, caused by atomic or polyatomic species and having the same mass/charge ratio of the analyte, can be a severe limit for a reliable assay of trace and ultratrace elements by ICP-MS. The DRC (TM) technology uses a reaction gas to overcome these interferences. Reactions of charge exchange, atom transfer, adduct formation, condensation and analyte association/condensation are the main mechanisms. Interfering ions tend to react with the gas exothermally, while, the analyte reacts endothermally. Selecting the most appropriate reaction gas in DRC-ICP-MS is the very critical point for the determination of strongly interfered elements. A careful evaluation of the reaction mechanisms and the chemistry involved are required. The DRC allows the use of different gases, among them, ammonia (NH3), methane (CH4), hydrogen (H-2) and oxygen (O-2) are the most known, but there are other potentially useful gases like nitrous oxide (N2O), nitrogen oxide (NO), carbon dioxide (CO2), fluoromethane (CH3F), sulphur hexafluoride (SF6) and carbon disulfide (CS2). This paper provides a review on the analytical challenges for a reliable assay of As, Cr, Se and V by DRC-ICP-MS and illustrates different approaches and mechanisms involved in the analysis of polymers, biological fluids (serum, urine and whole blood), rock, soil and particulate matter. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:6 / 13
页数:8
相关论文
共 76 条
[1]   Speciation of chromium in cow's milk by solid-phase extraction/dynamic reaction cell inductively coupled plasma mass spectrometry( DRC-ICP-MS) [J].
Ambushe, Abayneh A. ;
McCrindle, Robert I. ;
McCrindle, Cheryl M. E. .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2009, 24 (04) :502-507
[2]  
ANICICH VG, 2003, NASA
[3]   Gas-phase ion-molecule reactions for resolution of atomic isobars: AMS and ICP-MS perspectives [J].
Bandura, Dmitry R. ;
Baranov, Vladimir I. ;
Litherland, A. E. ;
Tanner, Scott D. .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2006, 255 :312-327
[4]   Reaction chemistry and collisional processes in multipole devices for resolving isobaric interferences in ICP-MS [J].
Bandura, DR ;
Baranov, VI ;
Tanner, SD .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 370 (05) :454-470
[5]   A dynamic reaction cell for inductively coupled plasma mass spectrometry (ICP-DRC-MS) - Part 1. The rf-field energy contribution in thermodynamics of ion-molecule reactions [J].
Baranov, VI ;
Tanner, SD .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (08) :1133-1142
[6]   Determination of trace elements in biological samples by inductively coupled plasma mass spectrometry with tetramethylammonium hydroxide solubilization at room temperature [J].
Batista, Bruno Lemos ;
Grotto, Denise ;
Rodrigues, Jairo Lisboa ;
de Oliveira Souza, Vanessa Cristina ;
Barbosa, Fernando, Jr. .
ANALYTICA CHIMICA ACTA, 2009, 646 (1-2) :23-29
[7]   Exploiting dynamic reaction cell inductively coupled plasma mass spectrometry (DRC-ICP-MS) for sequential determination of trace elements in blood using a dilute-and-shoot procedure [J].
Batista, Bruno Lemos ;
Rodrigues, Jairo Lisboa ;
Nunes, Juliana Andrade ;
de Oliveira Souza, Vanessa Cristina ;
Barbosa, Fernando, Jr. .
ANALYTICA CHIMICA ACTA, 2009, 639 (1-2) :13-18
[8]   Determination of vanadium by reaction cell inductively coupled plasma mass spectrometry [J].
Bednar, A. J. .
TALANTA, 2009, 78 (02) :453-457
[9]   Comparison of standard and reaction cell inductively coupled plasma mass spectrometry in the determination of chromium and selenium species by HPLC-ICP-MS [J].
Bednar, A. J. ;
Kirgan, R. A. ;
Jones, W. T. .
ANALYTICA CHIMICA ACTA, 2009, 632 (01) :27-34
[10]   Nitric oxide as an electron donor, an atom donor, an atom acceptor, and a ligand in reactions with atomic transition-metal and main-group cations in the gas phase [J].
Blagojevic, V ;
Flaim, E ;
Jarvis, MJV ;
Koyanagi, GK ;
Bohme, DK .