Torsional ultrasound modality for hard nucleus phacoemulsification cataract extraction

被引:56
作者
Zeng, M. [1 ,2 ]
Liu, X. [1 ]
Liu, Y. [1 ]
Xia, Y. [1 ]
Luo, L. [1 ]
Yuan, Z. [1 ]
Zeng, Y. [1 ]
Liu, Y. [1 ]
机构
[1] Sun Yat Sen Univ, Zhongshan Ophthalm Ctr, Guangzhou 510060, Guangdong, Peoples R China
[2] Second Affiliated Hosp, Guangzhou Med Coll, Guangzhou, Guangdong, Peoples R China
关键词
D O I
10.1136/bjo.2007.128504
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aim: To evaluate the efficacy and safety of phacoemulsification using torsional modality with different parameter settings for hard nucleus cataract extraction. Design: A prospective, randomised clinical study. Methods: A clinical practice study conducted at the Cataract Service, Zhongshan Ophthalmic Center, Sun-Yat-Sen University, and Guangzhou. One eye each from 198 consecutive patients with cataract density grade IV according to the Emery-Little system classification system, requiring phacoemulsification and intraocular lens implantation, was included. Eyes were randomly assigned to the Linear Torsional combined with Ultrasound power group (Linear Tor+US group, n = 66), 100% Fixed Torsional group (Fixed Tor group, n = 65) and conventional Ultrasound burst group (US group, n = 67). All surgeries were performed by a single experienced surgeon and outcomes evaluated by another surgeon masked to treatment. Intraoperative parameters were Ultrasound Time (UST), Cumulative Dissipated Energy (CDE) and surgical complications. Patients were examined on post-op days 1, 7 and 30. Postoperative outcomes were final best corrected visual acuity (BCVA), average central and incisional corneal thickness and central endothelial cell counts. Results: The mean UST was lower in the Fixed Tor group than in the US group and in the Lin US+Tor group (p <= 0.0001). The mean CDE was lower in the Lin Tor+US group and in the Fixed Tor group than in the US group (p <= 0.0001). Comparing with the two Tor group, the US group had a lower average BCVA on post-op 1, 7 (p <= 0.0001) and 30 (p>0.01), greater average central corneal and incisional thickness on days 1, 7 (p <= 0.0001) and 30 (p>0.01), and higher average corneal endothelial cell losses on day 7 and 30 days (p <= 0.0001). Conclusions: Torsional combined with ultrasound power or high fixed torsional amplitude can yield more effective hard nucleus phacoemulsification than conventional ultrasound modality.
引用
收藏
页码:1092 / 1096
页数:5
相关论文
共 21 条
  • [1] Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy for assessment of the anterior segment
    Dada, Tanuj
    Sihota, Ramanjit
    Gadia, Ritu
    Aggarwal, Anand
    Mandal, Subrata
    Gupta, Viney
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (05) : 837 - 840
  • [2] Performance comparison of the Alcon Legacy 20000 1.1 mm TurboSonics and 0.9 mm Aspiration Bypass System tips
    Davison, JA
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1999, 25 (10) : 1386 - 1391
  • [3] Domingues Flavia Guedes Pinto, 2005, Arq. Bras. Oftalmol., V68, P109, DOI 10.1590/S0004-27492005000100020
  • [4] Fluidics and heat generation of Alcon Infiniti and Legacy, Bausch&Lomb Millennium, and Advanced Medical Optics Sovereign phacoemulsification systems
    Floyd, Michael S.
    Valentine, Jeremy R.
    Olson, Randall J.
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 142 (03) : 387 - 392
  • [5] Gimbel Howard V, 2002, Curr Opin Ophthalmol, V13, P30, DOI 10.1097/00055735-200202000-00007
  • [6] Refractive lens exchange in high myopia: long term follow up
    Horgan, N
    Condon, PI
    Beatty, S
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2005, 89 (06) : 670 - 672
  • [7] Phacoemulsification parameters:: Series 20000 Legacy versus Legacy with AdvanTec software and NeoSoniX handpiece
    Jirásková, N
    Rozsíval, P
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (01) : 144 - 148
  • [8] Phacoemulsification with a bevel-down phaco tip: Phaco-drill
    Joo, CK
    Kim, YH
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1997, 23 (08) : 1149 - 1152
  • [9] Corneal endothelial damage by air bubbles during phacoemulsification
    Kim, EK
    Cristol, SM
    Geroski, DH
    McCarey, BE
    Edelhauser, HF
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1997, 115 (01) : 81 - 88
  • [10] Liu SM, 2004, ACTA METALL SIN, V40, P225