Magnetic Resonance Imaging in Patients With Hearing Implants - Follow-up on Prevalence and Complications

被引:6
作者
Loth, Andreas G. [1 ]
Fischer, Kathrin [1 ]
Hey, Anna Katharina [1 ]
Weiss, Roxanne [1 ]
Leinung, Martin [1 ]
Burck, Iris [2 ]
Linke, Albrecht [1 ]
Kramer, Sabine [1 ]
Stover, Timo [1 ]
Helbig, Silke [1 ]
机构
[1] Univ Hosp Frankfurt, Dept Otolaryngol Head & Neck Surg, Theodor Stem Kai 7, D-60590 Frankfurt, Germany
[2] Univ Hosp Frankfurt, Dept Diagnost & Intervent Radiol, Frankfurt, Germany
基金
爱尔兰科学基金会;
关键词
Cochlear implants; Complication; Hearing implants; Magnetic resonance imaging; Prevalence; COCHLEAR IMPLANTS; MRI; DISLOCATION;
D O I
10.1097/MAO.0000000000003249
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: To examine the number of magnetic resonance imaging (MRI) examinations performed in patients with hearing implants and to quantify side effects or complications related to this procedure. Study Design: Questionnaire. Setting: Tertiary referral center, academic hospital. Patients: One thousand four hundred sixty-onepatients with an implanted hearing system. Intervention: Patients were asked to complete a questionnaire either during a visit to the clinic (304) or by mail contact (1,157) between February 2018 and March 2019. Main Outcome Measures: Number of examinations by means of MRI per patient and number of side effects or complications. Results: A total of 711 questionnaires were returned. After excluding nonvalid information on the questionnaire, 12.8% of patients were identified who had undergone an MRI after having received their hearing implant. Within this group of 91 patients, the most common precaution undertaken was a head bandage (69%). Side effects were mainly pain (37%), followed by anxiety (15%) and tinnitus (9%). The MRI had to be aborted in 14% and dislocation of the magnet occurred in 7% of examinations. Conclusions: Our data indicate that patients undergoing hearing implant surgery need better information about the limitations and requirements of MRI. The occurrence of side effects is likely as only half of the patients in our study group were completely free of symptoms. Dislocation of the implant magnet was observed in several cases, hence patients and physicians need to be educated about this potential complication.
引用
收藏
页码:1334 / 1341
页数:8
相关论文
共 17 条
  • [1] Broomfield Stephen J, 2013, Cochlear Implants Int, V14, P51
  • [2] Magnetic Resonance Imaging With Cochlear Implant Magnet in Place: Safety and Imaging Quality
    Carlson, Matthew L.
    Neff, Brian A.
    Link, Michael J.
    Lane, John I.
    Watson, Robert E.
    McGee, Kiaran P.
    Bernstein, Matt A.
    Driscoll, Colin L. W.
    [J]. OTOLOGY & NEUROTOLOGY, 2015, 36 (06) : 965 - 971
  • [3] First MRI With New Cochlear Implant With Rotatable Internal Magnet System and Proposal for Standardization of Reporting Magnet-Related Artifact Size
    Cass, Nathan D.
    Honce, Justin M.
    O'Dell, Alexandra L.
    Gubbels, Samuel P.
    [J]. OTOLOGY & NEUROTOLOGY, 2019, 40 (07) : 883 - 891
  • [4] Cochlear implant magnet displacement during magnetic resonance imaging
    Deneuve, Sophie
    Loundon, Nathalie
    Leboulanger, Nicolas
    Rouillon, Isabelle
    Garabedian, Erea Noel
    [J]. OTOLOGY & NEUROTOLOGY, 2008, 29 (06) : 789 - 790
  • [5] MR Imaging and Cochlear Implants with Retained Internal Magnets: Reducing Artifacts near Highly Inhomogeneous Magnetic Fields
    Edmonson, Heidi A.
    Carlson, Matthew L.
    Patton, Alice C.
    Watson, Robert E.
    [J]. RADIOGRAPHICS, 2018, 38 (01) : 94 - 106
  • [6] Fallpauschalenbezogene Krankenhausstatistik (DRG-Statistik), OP PROZ VOLLST PAT P
  • [7] Grupe G, 2017, HNO, V65, pS35, DOI 10.1007/s00106-016-0129-7
  • [8] Magnet Dislocation: An Increasing and Serious Complication Following MRI in Patients with Cochlear Implants
    Hassepass, F.
    Stabenau, V.
    Arndt, S.
    Beck, R.
    Bulla, S.
    Grauvogel, T.
    Aschendorff, A.
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2014, 186 (07): : 680 - 685
  • [9] Revision Surgery Due to Magnet Dislocation in Cochlear Implant Patients: An Emerging Complication
    Hassepass, Frederike
    Stabenau, Vanessa
    Maier, Wolfgang
    Arndt, Susan
    Laszig, Roland
    Beck, Rainer
    Aschendorff, Antje
    [J]. OTOLOGY & NEUROTOLOGY, 2014, 35 (01) : 29 - 34
  • [10] Adverse Events and Discomfort During Magnetic Resonance Imaging in Cochlear Implant Recipients
    Kim, Bo Gyung
    Kim, Jin Won
    Park, Jeong Jin
    Kim, Sung Huhn
    Kim, Hee Nam
    Choi, Jae Young
    [J]. JAMA OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2015, 141 (01) : 45 - 52