W-shaped volume curve with gated myocardial perfusion single photon emission computed tomography

被引:5
作者
Kasai, T [1 ]
DePuey, EG [1 ]
Sponder, I [1 ]
机构
[1] Columbia Univ, Coll Phys & Surg, Div Nucl Med, Dept Radiol,St Lukes Roosevelt Hosp Ctr, New York, NY 10027 USA
关键词
gating errors; myocardial SPECT; left ventricular function; volume curve; time-activity curve;
D O I
10.1007/BF02986338
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: Gating errors (GEs) with ECG gated myocardial SPECT (G-SPECT) may occur irrespective of the presence or absence of arrhythmias. We evaluated the impact of GEs on both reconstructed tomograms and left ventricular ejection fraction (LVEF) derived from G-SPECT, and searched for clues to identify these errors. Methods: We studied 2 GE patients, 10 normal subjects (NLs), and 10 atrial fibrillation patients. Stress technetium-99m sestamibi G-SPECT was performed. Left ventricular (LV) contraction was evaluated in the beating slices. LVEF was calculated with G-SPECT using QGS (Cedars-Sinai, Los Angels) and p-FAST (Sapporo Medical University, Japan), and compared with that obtained by echocardiography (ECHO). LV volume curves were created by QGS and p-FAST. ne heart rates (HRs) were calculated from the acquired images, and compared with their resting HRs. The mean count density of the myocardium was measured and time-activity curves were created. Results: In patients with GEs, bi-phasic LV contraction was demonstrated with fading-out towards end-diastole. G-SPECT underestimated LVEF compared to ECHO by 10% or more. The volume curves appeared "W-shaped." The HRs from the images were slower than the resting HRs. The count density decrement from frame #1 to #8 was larger than that of NLs. The time-activity curves were different in shape from those of NLs. Conclusions: G-SPECT underestimates LVEF in patients with GEs. These errors can be identified with a combination of visual inspection of beating slices, time-volume curves, and time-activity curves. Monitoring the HR is a clue for anticipating and avoiding these errors.
引用
收藏
页码:59 / 64
页数:6
相关论文
共 17 条
  • [1] Artifactual reverse distribution pattern in myocardial perfusion SPECT with technetium-99m sestamibi
    Araujo, W
    DePuey, EG
    Kamran, M
    Undavia, M
    Friedman, M
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2000, 7 (06) : 633 - 638
  • [2] The additive value of gated SPET myocardial perfusion imaging in patients with known and suspected coronary artery disease
    Bavelaar-Croon, CDL
    Atsma, DE
    Van der Wall, EE
    Dibbets-Schneider, P
    Zwinderman, AH
    Pauwels, EKJ
    [J]. NUCLEAR MEDICINE COMMUNICATIONS, 2001, 22 (01) : 45 - 55
  • [3] DEPUEY EG, 1908, DIAGNOSTIC NUCL MED, P355
  • [4] ANALYSIS OF PROBABILITY AS AN AID IN THE CLINICAL-DIAGNOSIS OF CORONARY-ARTERY DISEASE
    DIAMOND, GA
    FORRESTER, JS
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1979, 300 (24) : 1350 - 1358
  • [5] Feng B, 2002, J NUCL MED, V43, P786
  • [6] Reverse redistribution of 99mTc-sestamibi after direct percutaneous transluminal coronary angioplasty in acute myocardial infarction:: relationship with wall motion and functional response to dobutamine stimulation
    Fujiwara, S
    Takeishi, Y
    Hirono, O
    Fukui, A
    Okuyama, M
    Yamaguchi, S
    Ito, M
    Kaneko, K
    Shishido, T
    Miyamoto, T
    Yuki, K
    Miyashita, T
    Takahashi, K
    Minamihaba, O
    Tomoike, H
    Yamauchi, S
    Fatema, K
    [J]. NUCLEAR MEDICINE COMMUNICATIONS, 2001, 22 (11) : 1223 - 1230
  • [7] GERMANO G, 1995, J NUCL MED, V36, P2138
  • [8] Impact of gating errors with electrocardiography gated myocardial perfusion SPECT
    Kasai, T
    DePuey, EG
    Shah, AA
    Merla, VC
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2003, 10 (06) : 709 - 711
  • [9] Assessment of myocardial washout of Tc-99m-sestamibi in patients with chronic heart failure: Comparison with normal control
    Kumita, S
    Seino, Y
    Cho, K
    Nakajo, H
    Toba, M
    Fukushima, Y
    Okamoto, N
    Takano, T
    Kumazaki, T
    [J]. ANNALS OF NUCLEAR MEDICINE, 2002, 16 (04) : 237 - 242
  • [10] Two- and three-dimensional assessments of myocardial perfusion and function by using technetium-99m sestamibi gated SPECT with a combination of count- and image-based techniques
    Nakata, T
    Katagiri, Y
    Odawara, Y
    Eguchi, M
    Kuroda, M
    Tsuchihashi, K
    Hareyama, M
    Shimamato, K
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2000, 7 (06) : 623 - 632