Toroidal coil counter-current chromatography - Achievement of high resolution by optimizing flow-rate, rotation speed, sample volume and tube length

被引:23
作者
Matsuda, K
Matsuda, S
Ito, Y [1 ]
机构
[1] NHLBI, Biophys Chem Lab, NIH, Bethesda, MD 20892 USA
[2] NHLBI, Pulm Crit Care Med Branch, NIH, Bethesda, MD 20892 USA
关键词
counter-current chromatography; optimization; toroidal coil centrifuge; amino acids;
D O I
10.1016/S0021-9673(98)00114-9
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This paper deals with optimization of a new seal-free compact toroidal coil centrifuge to achieve high resolution in analytical counter-current chromatography (CCC). Toroidal coil CCC (hydrostatic motion) has advantages compared with high-speed CCC (efficiently mixing solution with planetary motion) in the separation of protein or easily emulsified samples. A toroidal coil separation column of 0.4 mm I.D. PTFE tubing was accommodated around the periphery of the cylindrical centrifuge bowl. Using a two-phase solvent system composed of chloroform-acetic acid-0.1 M hydrochloric acid (2:2,1, v/v/v) and a set of dinitrophenyl-amino acids as test samples, a series of experiments was performed with parameters such as the column length, sample volume, flow-rate, elution mode of the mobile phase and rotation speed. The highest efficiency, over 10 000 theoretical plates, was achieved with a 100 m long coiled tube and an 11 ml total capacity at a flow-rate of 0.01 ml/min at 800 rpm. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:95 / 104
页数:10
相关论文
共 10 条
[1]   TOROIDAL COIL COUNTERCURRENT CHROMATOGRAPHY IN THE AFFINITY PARTITIONING OF NICOTINIC CHOLINERGIC RECEPTOR ENRICHED MEMBRANES [J].
FLANAGAN, SD ;
JOHANSSON, G ;
YOST, B ;
ITO, Y ;
SUTHERLAND, IA .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1984, 7 (02) :385-402
[2]  
ITO Y, 1986, CRIT REV ANAL CHEM, V17, P65, DOI 10.1080/10408348608542792
[3]   COUNTERCURRENT CHROMATOGRAPHY . LIQUID-LIQUID PARTITION CHROMATOGRAPHY WITHOUT SOLID SUPPORT [J].
ITO, Y ;
BOWMAN, RL .
SCIENCE, 1970, 167 (3916) :281-&
[4]   Effects of coriolis force on countercurrent chromatography [J].
Ito, Y ;
Ma, Y .
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 1998, 21 (1-2) :1-17
[5]   Toroidal coil counter-current chromatography study of the mass transfer rate of proteins in aqueous-aqueous polymer phase system [J].
Ito, Y ;
Matsuda, K ;
Ma, Y ;
Qi, L .
JOURNAL OF CHROMATOGRAPHY A, 1998, 802 (02) :277-283
[6]   NEW FLOW-THROUGH CENTRIFUGE WITHOUT ROTATING SEALS APPLIED TO PLASMAPHERESIS [J].
ITO, Y ;
SUAUDEAU, J ;
BOWMAN, RL .
SCIENCE, 1975, 189 (4207) :999-1000
[7]   HIGH-SPEED PREPARATIVE COUNTERCURRENT CHROMATOGRAPHY WITH A COIL PLANET CENTRIFUGE [J].
ITO, Y ;
SANDLIN, J ;
BOWERS, WG .
JOURNAL OF CHROMATOGRAPHY, 1982, 244 (02) :247-258
[8]   COUNTERCURRENT CHROMATOGRAPHY WITH FLOW-THROUGH CENTRIFUGE WITHOUT ROTATING SEALS [J].
ITO, Y ;
BOWMAN, RL .
ANALYTICAL BIOCHEMISTRY, 1978, 85 (02) :614-617
[9]  
Ito Y, 1996, HIGH SPEED COUNTERCU, P3
[10]  
Mandava N.B., 1988, COUNTER CURRENT CHRO, V44, P79