Deuterium/hydrogen isotope ratio measurement of water and organic samples by continuous-flow isotope ratio mass spectrometry using chromium as the reducing agent in an elemental analyser

被引:27
作者
Kelly, SD [1 ]
Heaton, KD [1 ]
Brereton, P [1 ]
机构
[1] Cent Sci Lab, York YO41 1LZ, N Yorkshire, England
关键词
D O I
10.1002/rcm.303
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A rapid continuous-flow technique for quantitative determination of hydrogen isotope ratios in water and organic materials at natural abundance levels is described. Water and organic samples were reduced in a helium stream at temperatures in excess of 1000 degreesC over chromium metal. a delta H-2 parts per thousand values of water and organic samples were determined by calibration against International Atomic Energy Agency reference materials V-SMOW and SLAP water. The accuracy of the method was demonstrated through the analysis of the intermediate water standard GISP and IAEA water intercomparison materials OH-1, CH-2 and OH-3. Values obtained using this technique compared well with reference values (maximum difference 2.2 parts per thousand). The precision of water analyses was less than 2.3 parts per thousand, (1 sigma or 1 standard deviation) in all cases. No apparent memory effect was observed when measuring samples at the natural abundance level. The application of the technique to organic molecules and the salts of organic acids was successfully demonstrated by measuring the delta H-2 parts per thousand values of an n-hexadecane laboratory reference and anhydrous calcium formate versus water calibration materials. (C) Crown copyright 2001. Reproduced with permission of the Controller of Her Majesty's Stationery Office. Published by John Wiley & Sons, Ltd.
引用
收藏
页码:1283 / 1286
页数:4
相关论文
共 24 条
[1]  
Barrie S., 1984, SPECTROSC-INT J, V3, P439
[2]  
COTTON FA, 1988, ADV INORG CHEM, P681
[3]  
Farquhar GD, 1997, RAPID COMMUN MASS SP, V11, P1554, DOI 10.1002/(SICI)1097-0231(199709)11:14<1554::AID-RCM980>3.0.CO
[4]  
2-I
[5]   Sample preparation device for quantitative hydrogen isotope analysis using chromium metal [J].
Gehre, M ;
Hoefling, R ;
Kowski, P ;
Strauch, G .
ANALYTICAL CHEMISTRY, 1996, 68 (24) :4414-4417
[6]   STANDARDS FOR STABLE ISOTOPE MEASUREMENTS IN NATURAL COMPOUNDS [J].
GONFIANTINI, R .
NATURE, 1978, 271 (5645) :534-536
[7]   AUTOMATED-ANALYSIS OF N-15 AND C-14 IN BIOLOGICAL SAMPLES [J].
HARRIS, D ;
PAUL, EA .
COMMUNICATIONS IN SOIL SCIENCE AND PLANT ANALYSIS, 1989, 20 (9-10) :935-947
[8]  
Hilkert AW, 1999, RAPID COMMUN MASS SP, V13, P1226, DOI 10.1002/(SICI)1097-0231(19990715)13:13<1226::AID-RCM575>3.0.CO
[9]  
2-9
[10]  
HORITA J, 1989, APPL RADIAT ISOTOPES, V40, P801, DOI 10.1016/0883-2889(89)90100-7