Comparison of strains for new generation CAD-CAM implant-supported crowns under loading

被引:13
作者
Yilmaz, Burak [1 ]
Alsaery, Amani [2 ]
Altintas, Subutay H. [3 ]
Schimmel, Martin [4 ]
机构
[1] Ohio State Univ, Div Restorat & Prosthet Dent, Columbus, OH 43210 USA
[2] Ohio State Univ, Div Restorat & Prosthet Dent, Grad Prosthodont, Columbus, OH 43210 USA
[3] Karadeniz Tech Univ, Dept Prosthodont, Trabzon, Turkey
[4] Univ Bern, Sch Dent Med, Dept Reconstruct Dent & Gerodontol, Bern, Switzerland
关键词
CAD; CAM technology; fixed implant prosthesis; stress; FINITE-ELEMENT-ANALYSIS; STRESS-DISTRIBUTION; CEMENT; PROSTHESES;
D O I
10.1111/cid.12894
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose The behavior of implant-crowns fabricated from recently introduced CAD-CAM zirconia-reinforced lithium silicate ceramic (ZLS) or a hybrid ceramic containing resin-reinforced glass network (HC) for strains around the implant platform is not well-known. A force absorption capacity of the latter has been claimed by the manufacturer. The aim of this study was to measure and compare recently introduced ZLS and HC with commonly used CAD-CAM implant crown materials for strain distribution around the implant platform. Methods Fourimplants(Legacy 1;Implant Direct) were placed into a resin block. Zirconia abutments (Straight contoured stock abutment; Implant Direct) were torqued into the implant fixtures to support crowns that were milled from a virtual design using four different CAD-CAM materials (Vita Suprinity PC (ZLS), Vita Enamix (HC), IPS Emax, ZirCAD Zirkonzahn) (N = 20). The crowns were cemented with a resin cement, loaded and strain values were recorded. Three-dimensional digital image correlation (3D-DIC) was used to measure compressive and tensile strains around the implant platforms. The tensile and compressive strains were recorded for each test and first analyzed for equality of variance using Levene's test, and further tested using a 2-way ANOVA repeated measures analysis of variance (alpha= .05). Results The data analysis showed no statistically significant effect of crown material on the generated strains (P > .05). Compressive strains were significantly higher than the tensile strains (P < .05). One of the HC crowns fractured during loading. Conclusions Strains generated around implant platform when new generation CAD-CAM crown materials were used was similar to strains observed when CAD-CAM zirconia and lithium disilicate crowns were used. New generation crown materials did not have a significant load absorption effect to change or minimize the strains generated around the implant platform.
引用
收藏
页码:397 / 402
页数:6
相关论文
共 22 条
[1]   A load-to-fracture and strain analysis of monolithic zirconia cantilevered frameworks [J].
Alshahrani, Faris A. ;
Yilmaz, Burak ;
Seidt, Jeremy D. ;
McGlumphy, Edwin A. ;
Brantley, William A. .
JOURNAL OF PROSTHETIC DENTISTRY, 2017, 118 (06) :752-758
[2]  
[Anonymous], 1996, TURKISH J ORAL IMP
[3]  
Bassit R, 2002, INT J ORAL MAX IMPL, V17, P17
[4]  
Benzing U R, 1995, Int J Oral Maxillofac Implants, V10, P188
[5]  
Brunski J B, 1999, Adv Dent Res, V13, P99
[6]   DETERMINING THE FORCE ABSORPTION QUOTIENT FOR RESTORATIVE MATERIALS USED IN IMPLANT OCCLUSAL SURFACES [J].
CIBIRKA, RM ;
RAZZOOG, ME ;
LANG, BR ;
STOHLER, CS .
JOURNAL OF PROSTHETIC DENTISTRY, 1992, 67 (03) :361-364
[7]  
Çiftçi Y, 2001, INT J PROSTHODONT, V14, P406
[8]  
Clelland N L, 1995, J Prosthodont, V4, P95, DOI 10.1111/j.1532-849X.1995.tb00323.x
[9]   Three-Dimensional Image Correlation Analyses for Strains Generated by Cement and Screw-Retained Implant Prostheses [J].
Clelland, Nancy L. ;
Yilmaz, Burak ;
Seidt, Jeremy D. .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2013, 15 (02) :271-282
[10]  
Davidowitz Gary, 2011, Dent Clin North Am, V55, P559, DOI 10.1016/j.cden.2011.02.011