Anodic stripping voltammetry of sulphide at a nickel film: towards the development of a reagentless sensor

被引:49
作者
Giovanelli, D
Lawrence, NS
Wilkins, SJ
Jiang, L
Jones, TGJ
Compton, RG
机构
[1] Univ Oxford, Phys & Theoret Chem Lab, Oxford OX1 3QZ, England
[2] Univ Oxford, Dept Mat, Oxford OX1 3PH, England
[3] Schlumberger Cambridge Res Ltd, Cambridge CB3 0EL, England
关键词
nickel deposition; glassy carbon; screen-printed carbon electrode; sulphide;
D O I
10.1016/S0039-9140(03)00269-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The determination of sulphide at an electrochemically generated nickel oxide layer at glassy carbon and screen-printed electrodes in acidic media has been examined and appraised. The NiO layer was found to produce a stripping-like signal to sulphide and gave a linear peak current response from 20 to 90 muM. The response was further enhanced by repetitive cycling allowing accumulation of nickel sulphide at the electrode surface such that lower micromolar levels of sulphide (i.e. 5 muM) can be determined. The response at the NiO layer to sulphide is shown to be reproducible over a period of 24 h, thereby offering the development of a disposable amperometric sensor for sulphide. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:211 / 220
页数:10
相关论文
共 35 条