Comparison of three methods of intraocular pressure measurement and their relation to central corneal thickness

被引:36
作者
Carbonaro, F. [1 ]
Andrew, T.
Mackey, D. A. [2 ]
Spector, T. D.
Hammond, C. J. [3 ]
机构
[1] Kings Coll London, Twin Res & Genet Epidemiol Dept, Sch Med, Twin Res & Genet Epidemiol Unit, London SE1 7EH, England
[2] Univ Melbourne, Royal Victorian Eye & Ear Hosp, Ctr Eye Res Australia, Melbourne, Vic, Australia
[3] Bromley Hosp NHS Trust, Princess Royal Univ Hosp, Orpington, England
关键词
comparison; reliability; ocular response analyser; dynamic contour tonometer; DYNAMIC CONTOUR TONOMETRY; GOLDMANN APPLANATION TONOMETRY; OCULAR PULSE AMPLITUDE; BIOMECHANICAL PROPERTIES; GLAUCOMA; HERITABILITY; AGE; PREVALENCE; AGREEMENT; CUP;
D O I
10.1038/eye.2010.11
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose The purpose of this study was to compare the reliability of the 'gold standard' Goldmann applanation tonometer (GAT), with that of the ocular response analyser (ORA), and the dynamic contour tonometer (DCT). Patients and methods A total of 694 subjects were recruited to participate from the TwinsUK (UK Adult Twin Registry) at St Thomas' Hospital, London. Intraocular pressure (IOP) was measured using GAT, ORA, and the DCT. The agreement between the three methods was assessed using the Bland-Altman method. Repeatability coefficients and coefficient of variation between first and second readings of the same eye were used to assess reliability. Results Mean age was 57.5 years (SD, 13.1; range, 16.1-88.5). The mean IOPs, calculated using the mean of two readings from the right eye were as follows: Goldmann (GAT), 14.1 +/- 2.8 mm Hg; IOPg (ORA), 15.9 +/- 3.2 mm Hg; IOPcc (ORA), 16.6 +/- 3.2 mm Hg; and DCT, 16.9 +/- 2.7 mm Hg. The 95% limits of agreement were for ORA (IOPcc): GAT, -2.07 to 7.18 mm Hg; for DCT: GAT, -0.49 to 6.21 mm Hg; and for DCT: ORA (IOPcc), -3.01 to 4.85 mm Hg. Coefficients of variation for the three tonometers were GAT, 8.3%; ORA, 8.2%; DCT, 6.3%. The repeatability coefficients were 3.4 mm Hg for GAT, 3.57 mm Hg for ORA and 3.09 mm Hg for DCT. GAT and ORA (IOPg) readings showed a positive correlation with central corneal thickness (P<0.005). Conclusions This study found similar reliability in all three tonometers. Bland-Altman plots showed the three instruments to have 95% limits of agreement outside the generally accepted limits, which means they are not interchangeable. GAT measurements were found to be significantly lower than the two newer instruments. Eye (2010) 24, 1165-1170; doi:10.1038/eye.2010.11; published online 12 February 2010
引用
收藏
页码:1165 / 1170
页数:6
相关论文
共 39 条
[1]   Comparison of dynamic contour tonometry and Goldmann applanation tonometry in glaucoma patients and healthy subjects [J].
Barleon, Lorenz ;
Hoffmann, Esther M. ;
Berres, Manfred ;
Pfeiffer, Norbert ;
Grus, Franz H. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 142 (04) :583-590
[2]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Dynamic contour tonometry in comparison to intracameral IOP measurements [J].
Boehm, Andreas G. ;
Weber, Anja ;
Pillunat, Lutz E. ;
Koch, Rainer ;
Spoerl, Eberhard .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2008, 49 (06) :2472-2477
[5]   Prevalence of glaucoma and intraocular pressure distribution in a defined population -: The Egna-Neumarkt study [J].
Bonomi, L ;
Marchini, G ;
Marraffa, M ;
Bernardi, P ;
De Franco, I ;
Perfetti, S ;
Varotto, A ;
Tenna, V .
OPHTHALMOLOGY, 1998, 105 (02) :209-215
[6]  
Brubaker RF, 1999, ARCH OPHTHALMOL-CHIC, V117, P104
[7]   The heritability of corneal hysteresis and ocular pulse amplitude - A twin study [J].
Carbonaro, Francis ;
Andrew, Toby ;
Mackey, David A. ;
Spector, Tim D. ;
Hammond, Chris J. .
OPHTHALMOLOGY, 2008, 115 (09) :1545-1549
[8]   Determinants and heritability of intraocular pressure and cup-to-disc ratio in a defined older population [J].
Chang, TC ;
Congdon, NG ;
Wojciechowski, R ;
Muñoz, B ;
Gilbert, D ;
Chen, P ;
Friedman, DS ;
West, SK .
OPHTHALMOLOGY, 2005, 112 (07) :1186-1191
[9]   A genetic contribution to intraocular pressure: The Beaver Dam Eye Study [J].
Duggal, P ;
Klein, AP ;
Lee, KE ;
Iyengar, SK ;
Klein, R ;
Bailey-Wilson, JE ;
Klein, BEK .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (02) :555-560
[10]  
EHLERS N, 1975, ACTA OPHTHALMOL, V53, P34