A Differential Programmable Josephson Voltage Standard for Low-voltage Measurement

被引:0
作者
Li, Honghui [1 ]
Gao, Yuan [1 ]
Wang, Zengmin [1 ]
机构
[1] Natl Inst Metrol, Beijing, Peoples R China
来源
2016 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2016) | 2016年
关键词
Josephson; standard; voltmeter; microvolt; nanovolt; voltage measurement; digital-analog conversion;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
An intrinsically accurate automated system for low-voltage measurements has been developed at NIM based on a differential programmable Josephson voltage standard. At 1 microvolt level, the typical difference between the mean value of 250 measurement points and the nominal value is 0.17 nV with the standard deviation of 1.4 nV. The experiments for uncertainty estimate have been carried out, which demonstrate the system offers a satisfactory reliability and stability. For fine-increment programmable Josephson voltage, a few expected applications are proposed.
引用
收藏
页数:2
相关论文
共 6 条
[1]   Application of Josephson series arrays to a DC quantum voltmeter [J].
Behr, R ;
Grimm, L ;
Funck, T ;
Kohlmann, J ;
Schulze, H ;
Müller, F ;
Schumacher, B ;
Warnecke, P ;
Niemeyer, J .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2001, 50 (02) :185-187
[2]   Application of the Josephson effect to voltage metrology [J].
Benz, SP ;
Hamilton, CA .
PROCEEDINGS OF THE IEEE, 2004, 92 (10) :1617-1629
[3]  
Honghui Li, 2014, 29th Conference on Precision Electromagnetic Measurements (CPEM 2014), P744, DOI 10.1109/CPEM.2014.6898600
[4]  
Lee J., 2012, 2012 Conference on Precision Electromagnetic Measurements (CPEM 2012), P40, DOI 10.1109/CPEM.2012.6250650
[5]  
Li HH, 2010, 2010 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS CPEM, P143, DOI 10.1109/CPEM.2010.5545010
[6]  
Rufenacht A., 2011, JVS 650 NEW VERSION, P1, DOI [10.1109/IEEESTD.2011.5967871, DOI 10.1109/IEEESTD.2011.5967871]