Commissioning an Exradin W2 plastic scintillation detector for clinical use in small radiation fields

被引:15
作者
Jacqmin, Dustin J. [1 ,2 ]
Miller, Jessica R. [1 ,2 ]
Barraclough, Brendan A. [1 ,3 ]
Labby, Zacariah E. [1 ]
机构
[1] Univ Wisconsin, Dept Human Oncol, 600 Highland Ave, Madison, WI 53726 USA
[2] Univ Wisconsin, Dept Med Phys, 1530 Med Sci Ctr, Madison, WI 53706 USA
[3] Duke Univ, Med Ctr, Dept Radiat Oncol, Durham, NC USA
关键词
Exradin W2; plastic scintillation detector; radiation measurement; small field dosimetry; stereotactic radiosurgery; MED; PHYS; 42; W1; SCINTILLATOR; DOSIMETRY;
D O I
10.1002/acm2.13728
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The purpose of this work is to evaluate the Standard Imaging Exradin W2 plastic scintillation detector (W2) for use in the types of fields used for stereotactic radiosurgery. Methods Prior to testing the W2 in small fields, the W2 was evaluated in standard large field conditions to ensure good detector performance. These tests included energy dependence, short-term repeatability, dose-response linearity, angular dependence, temperature dependence, and dose rate dependence. Next, scan settings and calibration of the W2 were optimized to ensure high quality data acquisition. Profiles of small fields shaped by cones and multi-leaf collimator (MLCs) were measured using the W2 and IBA RAZOR diode in a scanning water tank. Output factors for cones (4-17.5 mm) and MLC fields (1, 2, 3 cm) were acquired with both detectors. Finally, the dose at isocenter for seven radiosurgery plans was measured with the W2 detector. Results W2 exhibited acceptable warm-up behavior, short-term reproducibility, axial angular dependence, dose-rate linearity, and dose linearity. The detector exhibits a dependence upon energy, polar angle, and temperature. Scanning measurements taken with the W2 and RAZOR were in good agreement, with full-width half-maximum and penumbra widths agreeing to within 0.1 mm. The output factors measured by the W2 and RAZOR exhibited a maximum difference of 1.8%. For the seven point-dose measurements of radiosurgery plans, the W2 agreed well with our treatment planning system with a maximum deviation of 2.2%. The Cerenkov light ratio calibration method did not significantly impact the measurement of relative profiles, output factors, or point dose measurements. Conclusion The W2 demonstrated dosimetric characteristics that are suitable for radiosurgery field measurements. The detector agreed well with the RAZOR diode for output factors and scanned profiles and showed good agreement with the treatment planning system in measurements of clinical treatment plans.
引用
收藏
页数:16
相关论文
共 20 条
[1]   CERENKOV LIGHT GENERATED IN OPTICAL FIBERS AND OTHER LIGHT PIPES IRRADIATED BY ELECTRON-BEAMS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (04) :925-935
[2]   Comment on "Characterization of the Exradin W1 scintillator for use in radiotherapy" [Med. Phys. 42, 297-304 (2015)] [J].
Beierholm, A. R. ;
Behrens, C. F. ;
Andersen, C. E. .
MEDICAL PHYSICS, 2015, 42 (07) :4414-4416
[3]   Dosimetric characterization of the Exradin W1 plastic scintillator detector through comparison with an in-house developed scintillator system [J].
Beierholm, A. R. ;
Behrens, C. F. ;
Andersen, C. E. .
RADIATION MEASUREMENTS, 2014, 69 :50-56
[4]   The practical application of scintillation dosimetry in small-field photon-beam radiotherapy [J].
Burke, Elisa ;
Poppinga, Daniela ;
Schoenfeld, Andreas A. ;
Harder, Dietrich ;
Poppe, Bjoern ;
Looe, Hui Khee .
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2017, 27 (04) :324-333
[5]   Response to "Comment on 'Characterization of the Exradin W1 scintillator for use in radiotherapy' " [Med. Phys. 42, 297-304 (2015)] [J].
Carrasco, P. ;
Jornet, N. ;
Jordi, O. ;
Lizondo, M. ;
Latorre-Musoll, A. ;
Eudaldo, T. ;
Ruiz, A. ;
Ribas, M. .
MEDICAL PHYSICS, 2015, 42 (07) :4417-4418
[6]   Characterization of the Exradin W1 scintillator for use in radiotherapy [J].
Carrasco, P. ;
Jornet, N. ;
Jordi, O. ;
Lizondo, M. ;
Latorre-Musoll, A. ;
Eudaldo, T. ;
Ruiz, A. ;
Ribas, M. .
MEDICAL PHYSICS, 2015, 42 (01) :297-304
[7]   Characterisation of a plastic scintillation detector to be used in a multicentre stereotactic radiosurgery dosimetry audit [J].
Dimitriadis, A. ;
Patallo, I. Silvestre ;
Billas, I. ;
Duane, S. ;
Nisbet, A. ;
Clark, C. H. .
RADIATION PHYSICS AND CHEMISTRY, 2017, 140 :373-378
[8]  
Fontbonne JM, 2002, IEEE T NUCL SCI, V49, P2223, DOI 10.1109/TNS.2002.803680
[9]   Variation of kQclin, Qmsrfclin, fmsr for the small-field dosimetric parameters percentage depth dose, tissue-maximum ratio, and off-axis ratio [J].
Francescon, Paolo ;
Beddar, Sam ;
Satariano, Ninfa ;
Das, Indra J. .
MEDICAL PHYSICS, 2014, 41 (10)
[10]   Spectral discrimination of Cerenkov radiation in scintillating dosimeters [J].
Frelin, AM ;
Fontbonne, JM ;
Ban, G ;
Colin, J ;
Labalme, M ;
Batalla, A ;
Isambert, A ;
Vela, A ;
Leroux, T .
MEDICAL PHYSICS, 2005, 32 (09) :3000-3006