Open source deformable image registration system for treatment planning and recurrence CT scans

被引:0
作者
Zukauskaite, Ruta [1 ,2 ]
Brink, Carsten [1 ,3 ]
Hansen, Christian Ronn [3 ,4 ]
Bertelsen, Anders [3 ,4 ]
Johansen, Jorgen [2 ]
Grau, Cai [5 ]
Eriksen, Jesper Grau [2 ]
机构
[1] Univ Southern Denmark, Inst Clin Res, Odense, Denmark
[2] Odense Univ Hosp, Dept Oncol, Sdr Blvd 29, DK-5000 Odense C, Denmark
[3] Odense Univ Hosp, Lab Radiat Phys, Odense, Denmark
[4] Univ Sydney, Inst Med Phys, Sch Phys, Sydney, NSW, Australia
[5] Aarhus Univ Hosp, Dept Oncol, Aarhus, Denmark
关键词
Computed tomography; Validation; elastix; head and neck; ADAPTIVE RADIATION-THERAPY; CONE-BEAM CT; NECK-CANCER; HEAD; VALIDATION; RADIOTHERAPY;
D O I
10.1007/s00066-016-0998-4
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Clinical application of deformable registration (DIR) of medical images remains limited due to sparse validation of DIR methods in specific situations, e. g. in case of cancer recurrences. In this study the accuracy of DIR for registration of planning CT (pCT) and recurrence CT (rCT) images of head and neck squamous cell carcinoma (HNSCC) patients was evaluated. Twenty patients treated with definitive IMRT for HNSCC in 2010-2012 were included. For each patient, a pCT and an rCT scan were used. Median interval between the scans was 8.5 months. One observer manually contoured eight anatomical regions-of-interest (ROI) twice on pCT and once on rCT. pCT and rCT images were deformably registered using the open source software elastix. Mean surface distance (MSD) and Dice similarity coefficient (DSC) between contours were used for validation of DIR. A measure for delineation uncertainty was estimated by assessing MSD from the re-delineations of the same ROI on pCT. DIR and manual contouring uncertainties were correlated with tissue volume and rigidity. MSD varied 1-3 mm for different ROIs for DIR and 1-1.5 mm for re-delineated ROIs performed on pCT. DSC for DIR varied between 0.58 and 0.79 for soft tissues and was 0.79 or higher for bony structures, and correlated with the volumes of ROIs (r = 0.5, p < 0.001) and tissue rigidity (r = 0.54, p < 0.001). DIR using elastix in HNSCC on planning and recurrence CT scans is feasible; an uncertainty of the method is close to the voxel size length of the planning CT images.
引用
收藏
页码:545 / 551
页数:7
相关论文
共 20 条
  • [1] Quantification of volumetric and geometric changes occurring during fractionated radiotherapy for head-and-neck cancer using an integrated CT/linear accelerator system
    Barker, JL
    Garden, AS
    Ang, KK
    O'Daniel, JC
    Wang, H
    Court, LE
    Morrison, WH
    Rosenthal, DI
    Chao, KSC
    Tucker, SL
    Mohan, R
    Dong, L
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (04): : 960 - 970
  • [2] Radiation dose response of normal lung assessed by Cone Beam CT - A potential tool for biologically adaptive radiation therapy
    Bertelsen, Anders
    Schytte, Tine
    Bentzen, Soren M.
    Hansen, Olfred
    Nielsen, Morten
    Brink, Carsten
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 100 (03) : 351 - 355
  • [3] Locoregional Control of Non-Small Cell Lung Cancer in Relation to Automated Early Assessment of Tumor Regression on Cone Beam Computed Tomography
    Brink, Carsten
    Bernchou, Uffe
    Bertelsen, Anders
    Hansen, Olfred
    Schytte, Tine
    Bentzen, Soren M.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 89 (04): : 916 - 923
  • [4] Brock KK, 2014, IMAG MED DIAGN THER, P41
  • [5] Comparison of 12 deformable registration strategies in adaptive radiation therapy for the treatment of head and neck tumors
    Castadot, Pierre
    Lee, John Aldo
    Parraga, Adriane
    Geets, Xavier
    Macq, Benoit
    Gregoire, Vincent
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 89 (01) : 1 - 12
  • [6] MEASURES OF THE AMOUNT OF ECOLOGIC ASSOCIATION BETWEEN SPECIES
    DICE, LR
    [J]. ECOLOGY, 1945, 26 (03) : 297 - 302
  • [7] Methods for estimating the site of origin of locoregional recurrence in head and neck squamous cell carcinoma
    Due, A. K.
    Vogelius, I. R.
    Aznar, M. C.
    Bentzen, S. M.
    Berthelsen, A. K.
    Korreman, S. S.
    Kristensen, C. A.
    Specht, L.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2012, 188 (08) : 671 - 676
  • [8] Feasibility of Multimodal Deformable Registration for Head and Neck Tumor Treatment Planning
    Fortunati, Valerio
    Verhaart, Rene F.
    Angeloni, Francesco
    van der Lugt, Aad
    Niessen, Wiro J.
    Veenland, Jifke F.
    Paulides, Margarethus M.
    van Walsum, Theo
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 (01): : 85 - 93
  • [9] Evaluation of Tumor Shape Variability in Head-and-Neck Cancer Patients Over the Course of Radiation Therapy Using Implanted Gold Markers
    Hamming-Vrieze, Olga
    van Kranen, Simon Robert
    van Beek, Suzanne
    Heemsbergen, Wilma
    van Herk, Marcel
    van den Brekel, Michiel Wilhelmus Maria
    Sonke, Jan-Jakob
    Rasch, Coenraad Robert Nico
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (02): : E201 - E207
  • [10] Accuracy quantification of a deformable image registration tool applied in a clinical setting
    Hoffmann, Christoph
    Krause, Sonja
    Stoiber, Eva Maria
    Mohr, Angela
    Rieken, Stefan
    Schramm, Oliver
    Debus, Jurgen
    Sterzing, Florian
    Bendl, Rolf
    Giske, Kristina
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (01): : 237 - 245