DEPTH DOSE DETERMINATION FOR A MIXED RADIATION FIELD USING A THIN PLASTIC SCINTILLATOR DOSIMETRY SYSTEM

被引:1
作者
Aydarous, Abdulkadir [1 ,2 ]
Waker, Anthony [1 ]
机构
[1] Univ Ontario, Fac Energy Syst & Nucl Sci, Inst Technol, Oshawa, ON, Canada
[2] Taif Univ, Dept Phys, Fac Sci, Al Hawiah, Taif, Saudi Arabia
关键词
plastic scintillator; MCNP code; mixed field radiation; dose;
D O I
10.2298/NTRP1201020A
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Plastic scintillators, due to their favorable characteristics compared with other dosimetry techniques, were used as detectors to estimate dose distributions in high gradient dose fields. In this study, a thin plastic scintillator (type BC-408) was coupled to a photomultiplier tube and multichannel analyzer as a technique for real-time dose measurements. The well-defined beta, gamma, and beta-gamma emitters (Cs-137, Ba-133, Na-22, Cd-109, Fe-55, and Am-241) have enabled parallel depth dose measurements with Monte-Carlo calculations to be critically compared. The measurements of doses were made for depths range of 0.1 mm to 5 mm. The MCNP dose results were comparable with the plastic scintillator detector and can be used to approximately estimate the dose rate values from mixed electron-photon fields. The minimum dose rate that can be measured by the plastic scintillator system was similar to 2 mu Gy/h and was for Cd-109 source of activity 222 Bq.
引用
收藏
页码:20 / 27
页数:8
相关论文
共 15 条
[1]  
Aydarous A. S., 2010, P INT WORKSH REALT M, P41
[2]   Dose distribution measurements and calculations for Dounreay hot particles [J].
Aydarous, A. Sh. ;
Charles, M. W. ;
Darley, P. J. .
RADIATION PROTECTION DOSIMETRY, 2008, 128 (02) :146-158
[3]   Development of an ICCD-scintillator system for measurement of spatial dose distributions around 'hot particles' [J].
Aydarous, AS ;
Darley, PJ ;
Charles, MW .
RADIATION PROTECTION DOSIMETRY, 2004, 108 (04) :317-326
[4]   A wide dynamic range, high-spatial-resolution scanning system for radiochromic dye films [J].
Aydarous, AS ;
Darley, PJ ;
Charles, MW .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (05) :1379-1389
[5]   Plastic scintillation dosimetry and its application to radiotherapy [J].
Beddar, A. S. .
RADIATION MEASUREMENTS, 2006, 41 :S124-S133
[6]   Monte Carlo calculations of the absorbed dose and energy dependence of plastic scintillators [J].
Beddar, AS ;
Briere, TM ;
Mourtada, FA ;
Vassiliev, ON ;
Liu, HH ;
Mohan, R .
MEDICAL PHYSICS, 2005, 32 (05) :1265-1269
[7]  
Briesmeister J., 2000, GEN MONTE CARLO N PA
[8]  
CHARLES MW, 1991, RADIAT PROT DOSIM, V39, P39
[9]  
DARLEY PJ, 1991, RADIAT PROT DOSIM, V39, P61
[10]  
*INT COMM RAD UN M, 1997, 56 ICRU INT COMM RAD