First Image From a Combined Positron Emission Tomography and Field-Cycled MRI System

被引:13
作者
Bindseil, Geron A. [1 ]
Gilbert, Kyle M. [1 ]
Scholl, Timothy J. [1 ]
Handler, William B. [1 ]
Chronik, Blaine A. [1 ]
机构
[1] Univ Western Ontario, Dept Phys & Astron, London, ON N6A 3K7, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
field-cycled MRI; positron emission tomography; multimodality imaging; PET/MRI development; hybrid imaging; PET; SCANNER; ACQUISITION;
D O I
10.1002/mrm.22793
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Combining positron emission tomography and MRI modalities typically requires using either conventional MRI with a MR-compatible positron emission tomography system or a modified MR system with conventional positron emission tomography. A feature of field-cycled MRI is that all magnetic fields can be turned off rapidly, enabling the use of conventional positron emission tomography detectors based on photomultiplier tubes. In this demonstration, two photomultiplier tube-based positron emission tomography detectors were integrated with a field-cycled MRI system (0.3 T/4 MHz) by placing them into a 9-cm axial gap. A positron emission tomography-MRI phantom consisting of a triangular arrangement of positron-emitting point sources embedded in an onion was imaged in a repeating interleaved sequence of similar to 1 sec MRI then 1 sec positron emission tomography. The first multimodality images from the combined positron emission tomography and field-cycled MRI system show no additional artifacts due to interaction between the systems and demonstrate the potential of this approach to combining positron emission tomography and MRI. Magn Reson Med 66:301-305, 2011. (C) 2011 Wiley-Liss, Inc.
引用
收藏
页码:301 / 305
页数:5
相关论文
共 25 条
  • [1] SCINTILLATION CAMERA
    ANGER, HO
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1958, 29 (01) : 27 - 33
  • [2] Boone John M, 2004, Mol Imaging, V3, P149, DOI 10.1162/1535350042380326
  • [3] Simultaneous in vivo positron emission tomography and magnetic resonance imaging
    Catana, Ciprian
    Procissi, Daniel
    Wu, Yibao
    Judenhofer, Martin S.
    Qi, Jinyi
    Pichler, Bernd J.
    Jacobs, Russell E.
    Cherry, Simon R.
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2008, 105 (10) : 3705 - 3710
  • [4] The integration of positron emission tomography with magnetic resonance imaging
    Cherry, Simon R.
    Louie, Angelique Y.
    Jacobs, Russell E.
    [J]. PROCEEDINGS OF THE IEEE, 2008, 96 (03) : 416 - 438
  • [5] Chun S Y., 2010, Proceedings of the 51st Annual Meeting of the Society of Nuclear Medicine, P80
  • [6] PET/MRI system design
    Delso, Gaspar
    Ziegler, Sibylle
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 : 86 - 92
  • [7] Fueger BJ, 2006, J NUCL MED, V47, P999
  • [8] Evaluation of a Positron Emission Tomography (PET)-Compatible Field-Cycled MRI (FCMRI) Scanner
    Gilbert, Kyle M.
    Scholl, Timothy J.
    Handler, William B.
    Alford, Jamu K.
    Chronik, Blaine A.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2009, 62 (04) : 1017 - 1025
  • [9] Simultaneous PET-MRI:: a new approach for functional and morphological imaging
    Judenhofer, Martin S.
    Wehrl, Hans F.
    Newport, Danny F.
    Catana, Ciprian
    Siegel, Stefan B.
    Becker, Markus
    Thielscher, Axel
    Kneilling, Manfred
    Lichy, Matthias P.
    Eichner, Martin
    Klingel, Karin
    Reischl, Gerald
    Widmaier, Stefan
    Roecken, Martin
    Nutt, Robert E.
    Machulla, Hans-Juergen
    Uludag, Kamil
    Cherry, Simon R.
    Claussen, Claus D.
    Pichler, Bernd J.
    [J]. NATURE MEDICINE, 2008, 14 (04) : 459 - 465
  • [10] Prepolarized fast spin-echo pulse sequence for low-field MRI
    Kegler, C.
    Seton, H. C.
    Hutchison, J. M. S.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2007, 57 (06) : 1180 - 1184