Copper nano-clusters prepared by one-step electrodeposition and its application on nitrate sensing

被引:30
作者
Li, Y. [1 ]
Sun, J. Z. [1 ]
Bian, C. [1 ]
Tong, J. H. [1 ]
Dong, H. P. [1 ]
Zhang, H. [1 ]
Xia, S. H. [1 ]
机构
[1] Chinese Acad Sci, Inst Elect, State Key Lab Transducer Technol, Beijing, Peoples R China
关键词
CHEMICAL OXYGEN-DEMAND; GLUCOSE; NANOSTRUCTURES; SENSORS; GOLD;
D O I
10.1063/1.4905712
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
This paper describes the fabrication and characterization of copper nano-clusters prepared by a simple one-step electrodeposition process on platinum microelectrode, and the application for nitrate determination. The one-step electrodepostion process was performed by chronoamperometry scan in acidic copper sulphate electrolyte directly. The SEM and electrochemical examination showed that the morphologies and microstructures of deposited copper layers can be precisely controlled by using different deposition voltages. It was found that the copper layer is porous when the deposition voltage is higher than -500 mV, and this porous layer has a larger effective surface area compared with the corresponding smooth flat copper layer deposited under voltage less than -300 mV. Under the optimized deposition voltage, copper clusters constructed by uniform nanoparticles with an average diameter of about 100 nm can be obtained. The mechanism of electrodeposition process for this method was also speculated. The copper layers deposited under different voltages are used in a series of tests in order to evaluate their performance for nitrate sensing. The experimental results reveal that the microelectrode modified by fixed potential deposition under -700 mV had a higher sensitivity of 39.31 mu A/mmolL(-1) for nitrate detection within the concentration ranging from 0.1 mmolL(-1) to 4.0 mmolL(-1). (C) 2015 Author(s). All article content, except where otherwise noted, is licensed under a Creative Commons Attribution 3.0 Unported License.
引用
收藏
页数:7
相关论文
共 21 条
[1]   Electrochemical Conversion of Copper-Based Hierarchical Micro/Nanostructures to Copper Metal Nanoparticles and Their Testing in Nitrate Sensing [J].
Andreoli, Enrico ;
Annibaldi, Valeria ;
Rooney, Denise A. ;
Liao, Kang-Shyang ;
Alley, Nigel J. ;
Curran, Seamus A. ;
Breslin, Carmel B. .
ELECTROANALYSIS, 2011, 23 (09) :2164-2173
[2]   The use of nanoparticles in electroanalysis: an updated review [J].
Campbell, Fallyn W. ;
Compton, Richard G. .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2010, 396 (01) :241-259
[3]   Large-scale synthesis of high-quality ultralong copper nanowires [J].
Chang, Y ;
Lye, ML ;
Zeng, HC .
LANGMUIR, 2005, 21 (09) :3746-3748
[4]   Fabrication of Cu-modified boron-doped diamond microband electrodes and their application for selective detection of glucose [J].
Chiku, Masanobu ;
Watanabe, Takeshi ;
Einaga, Yasuaki .
DIAMOND AND RELATED MATERIALS, 2010, 19 (7-9) :673-679
[5]   A renewable copper electrode as an amperometric flow detector for nitrate determination in mineral water and soft drink samples [J].
Gamboa, Juan C. M. ;
Pena, Roselyn C. ;
Paixao, Thiago R. L. C. ;
Bertotti, Mauro .
TALANTA, 2009, 80 (02) :581-585
[6]   Copper nanocluster-based fluorescent sensors for sensitive and selective detection of kojic acid in food stuff [J].
Gao, Zhao ;
Su, Rongxin ;
Qi, Wei ;
Wang, Libing ;
He, Zhimin .
SENSORS AND ACTUATORS B-CHEMICAL, 2014, 195 :359-364
[7]   Hydrogen bubble dynamic template synthesis of porous gold for nonenzymatic electrochemical detection of glucose [J].
Li, Ying ;
Song, Yan-Yan ;
Yang, Chen ;
Xia, Xing-Hua .
ELECTROCHEMISTRY COMMUNICATIONS, 2007, 9 (05) :981-988
[8]   Scanning electrochemical microscopy investigation of nitrate reduction at activated copper cathodes in acidic medium [J].
Lima, Alex S. ;
Salles, Maiara O. ;
Ferreira, Tiago L. ;
Paixao, Thiago R. L. C. ;
Bertotti, Mauro .
ELECTROCHIMICA ACTA, 2012, 78 :446-451
[9]   Shape and size controlled fabrication of copper nanopowders from industrial electrolytes by pulse electrodeposition [J].
Lukomska, Aneta ;
Plewka, Anna ;
Los, Przemyslaw .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2009, 637 (1-2) :50-54
[10]   Simple one step electrochemical preparation of copper nanostructures [J].
Mignani, Adriana ;
Ballarin, Barbara ;
Boanini, Elisa ;
Cassani, Maria Cristina .
ELECTROCHIMICA ACTA, 2014, 115 :537-545