The effects of prolonged ethanol exposure on the mechanical properties of polyurethane and silicone catheters used for intravascular access

被引:103
作者
Crnich, CJ
Halfmann, JA
Crone, WC
Maki, DG
机构
[1] Univ Wisconsin Hosp, Infect Dis Sect, Dept Med, Madison, WI 53792 USA
[2] Univ Wisconsin, Dept Engn Phys, Madison, WI USA
[3] Univ Wisconsin, Sch Med, Madison, WI USA
[4] Univ Wisconsin, Engn Mech Program, Madison, WI USA
关键词
D O I
10.1086/502607
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
BACKGROUND: Products containing alcohol are commonly used with intravascular devices at insertion, to remove lipids from occluded intravascular devices used during parenteral nutrition, and increasingly for the prevention and treatment of intravascular device-related bloodstream infection. The effects of alcohol on the integrity of intravascular devices remain unknown. METHODS: Two types of widely used commercial peripherally inserted central catheters, one made of polyetherurethane and one made of silicone, were exposed to a 70% ethanol lock solution for up to 10 weeks. Mechanical testing was performed to identify force-at-break, stress, strain, modulus of elasticity, modulus of toughness, and wall area of ethanol-exposed and control catheters. RESULTS: No significant differences between exposed and unexposed catheters were identified for any of the mechanical parameters tested except for a marginal reduction in the modulus of elasticity for both polyetherurethane and silicone catheters and minor increases in the wall area of polyetherurethane catheters. CONCLUSIONS: These data indicate that exposure to a 70% ethanol lock solution does not appreciably alter the integrity of selected commercial polyetherurethane and silicone catheters. Given the greatly expanded use of alcoholic solutions with intravascular devices of all types, we believe that manufacturers would be well advised to subject their catheters and other intravascular devices to formal testing of the type employed in this study.
引用
收藏
页码:708 / 714
页数:7
相关论文
共 20 条
  • [1] Failure of central vein catheters after prolonged use
    Agraharkar, M
    [J]. SEMINARS IN DIALYSIS, 2000, 13 (03) : 205 - 206
  • [2] [Anonymous], 42 INT C ANT AG CHEM
  • [3] *BOST SCI, 2004, VAXC PICC PASV VALV
  • [4] *CTR DEV RAD HLTH, 2004, MAN US FAC DEV EXP D
  • [5] Ethanol-lock technique in the treatment of bloodstream infections in pediatric oncology patients with broviac catheter
    Dannenberg, C
    Bierbach, U
    Rothe, A
    Beer, J
    Körholz, D
    [J]. JOURNAL OF PEDIATRIC HEMATOLOGY ONCOLOGY, 2003, 25 (08) : 616 - 621
  • [6] Effect of skin disinfection with octenidine dihydrochloride on insertion site colonization of intravascular catheters
    Dettenkofer, M
    Jonas, D
    Wiechmann, C
    Rossner, R
    Frank, U
    Zentner, J
    Daschner, FD
    [J]. INFECTION, 2002, 30 (05) : 282 - 285
  • [7] *ISO, 1995, 105551 ISO 1
  • [8] ETHANOL FLUSH FOR THE PREVENTION OF CATHETER OCCLUSION
    JOHNSTON, DA
    WALKER, K
    RICHARDS, J
    PENNINGTON, CR
    [J]. CLINICAL NUTRITION, 1992, 11 (02) : 97 - 100
  • [9] Kluger DM, 2000, INFECT CONT HOSP EP, V21, P95
  • [10] Polyurethane central venous catheters, hydrochloric acid and 70% ethanol: A safety evaluation
    McHugh, GJ
    Wild, DJC
    Havill, JH
    [J]. ANAESTHESIA AND INTENSIVE CARE, 1997, 25 (04) : 350 - 353