Electrode Array Type and Its Impact on Impedance Fluctuations and Loss of Residual Hearing in Cochlear Implantation

被引:18
作者
Thompson, Nicholas J. [1 ]
Dillon, Margaret T. [1 ]
Buss, Emily [1 ]
Park, Lisa R. [1 ]
Pillsbury, Harold C., III [1 ]
O'Connell, Brendan P. [1 ]
Brown, Kevin D. [1 ]
机构
[1] Univ N Carolina, Dept Otolaryngol Head & Neck Surg, Chapel Hill, NC 27515 USA
关键词
cochlear implant; electrode array; hearing preservation; impedance; residual hearing; MULTICENTER CLINICAL-TRIAL; ANGULAR INSERTION DEPTH; SPEECH-PERCEPTION; ACOUSTIC STIMULATION; PRESERVATION; OUTCOMES; LOCALIZATION; PATHOLOGY; LOCATION; BENEFITS;
D O I
10.1097/MAO.0000000000002457
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Hypothesis/Objective:Determine variables associated with electrode impedance fluctuations and loss of residual hearing in cochlear implant (CI) recipients.Background:CI recipients with postoperative hearing preservation demonstrate superior speech perception with an electric-acoustic stimulation (EAS) device as compared with a CI-alone device. Maintaining superior speech perception over time relies on long-term hearing preservation; therefore, understanding variables that may contribute to loss of residual hearing is needed. Recent reports suggest a relationship between changes in electrode impedance and loss of residual hearing. The variables influencing this relationship have yet to be determined.Methods:Review of pediatric and adult CI cases from 2013 to 2016 who presented with preoperative residual hearing. Regression analysis was performed to evaluate effects of array type (lateral wall vs. perimodiolar), manufacturer, age at implantation, and preoperative hearing on impedance. The correlation between peak impedance change and change in low-frequency hearing was determined.Results:One hundred forty-six CI recipients presented with preoperative residual hearing. A multivariate regression analysis demonstrated a statistically significant association between preoperative hearing thresholds (p=0.017), device manufacturer (p=0.011), and array type (p=0.038) on postoperative impedance changes. Hearing preservation rates and change in impedance differed by electrode array type. The association between peak impedance changes and loss of residual hearing differed between manufacturers (R-2=0.208, p=0.029 vs. R-2=0.016, p=0.609).Conclusion:Impedance fluctuation appears to be a marker for loss of residual hearing for specific electrode array types and manufacturers. Specific arrays may affect the cochlear microenvironment differently, with different effects on postoperative hearing preservation.
引用
收藏
页码:186 / 191
页数:6
相关论文
共 34 条
  • [1] Hearing Preservation and Speech Perception Outcomes With Electric-Acoustic Stimulation After 12 Months of Listening Experience
    Adunka, Oliver F.
    Dillon, Margaret T.
    Adunka, Marcia C.
    King, English R.
    Pillsbury, Harold C.
    Buchman, Craig A.
    [J]. LARYNGOSCOPE, 2013, 123 (10) : 2509 - 2515
  • [2] Scalar Localization by Cone-Beam Computed Tomography of Cochlear Implant Carriers: A Comparative Study Between Straight and Periomodiolar Precurved Electrode Arrays
    Boyer, Eric
    Karkas, Alexandre
    Attye, Arnaud
    Lefournier, Virginie
    Escude, Bernard
    Schmerber, Sebastien
    [J]. OTOLOGY & NEUROTOLOGY, 2015, 36 (03) : 422 - 429
  • [3] Electrode Impedance Fluctuations as a Biomarker for Inner Ear Pathology After Cochlear Implantation
    Choi, June
    Payne, Matthew R.
    Campbell, Luke J.
    Bester, Christo W.
    Newbold, Carrie
    Eastwood, Hayden
    O'Leary, Stephen J.
    [J]. OTOLOGY & NEUROTOLOGY, 2017, 38 (10) : 1433 - 1439
  • [4] An overview of cochlear implant electrode array designse
    Dhanasingh, Anandhan
    Jolly, Claude
    [J]. HEARING RESEARCH, 2017, 356 : 93 - 103
  • [5] Influence of Test Condition on Speech Perception With Electric-Acoustic Stimulation
    Dillon, Margaret T.
    Buss, Emily
    Adunka, Oliver F.
    Buchman, Craig A.
    Pillsbury, Harold C.
    [J]. AMERICAN JOURNAL OF AUDIOLOGY, 2015, 24 (04) : 520 - 528
  • [6] Benefits of Localization and Speech Perception with Multiple Noise Sources in Listeners with a Short-Electrode Cochlear Implant
    Dunn, Camille C.
    Perreau, Ann
    Gantz, Bruce
    Tyler, Richard S.
    [J]. JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2010, 21 (01) : 44 - 51
  • [7] Comparative study of cochlear damage with three perimodiolar electrode designs
    Eshraghi, AA
    Yang, NW
    Balkany, TJ
    [J]. LARYNGOSCOPE, 2003, 113 (03) : 415 - 419
  • [8] Eshraghi Adrien A, 2006, Curr Opin Otolaryngol Head Neck Surg, V14, P323, DOI 10.1097/01.moo.0000244189.74431.df
  • [9] Multicenter clinical trial of the Nucleus Hybrid S8 cochlear implant: Final outcomes
    Gantz, Bruce J.
    Dunn, Camille
    Oleson, Jacob
    Hansen, Marlan
    Parkinson, Aaron
    Turner, Christopher
    [J]. LARYNGOSCOPE, 2016, 126 (04) : 962 - 973
  • [10] Gfeller Kate E, 2006, Audiol Neurootol, V11 Suppl 1, P12, DOI 10.1159/000095608