Compton camera study for high efficiency SPECT and benchmark with Anger system

被引:44
作者
Fontana, M. [1 ]
Dauvergne, D. [2 ]
Letang, J. M. [3 ]
Ley, J-L [1 ]
Testa, E. [1 ]
机构
[1] Univ Claude Bernard Lyon 1, Univ Lyon, CNRS IN2P3, Inst Phys Nucl Lyon, F-69622 Villeurbanne, France
[2] Univ Grenoble Alpes, LSPC, CNRS IN2P3, UMR 5821, Grenoble, France
[3] Univ Claude Bernard Lyon 1, Univ Lyon, INSA Lyon,UMR 5220 U1206, CNRS,INSERM,UJM St Etienne,Ctr Leon Berard,CREATI, F-69373 Lyon, France
关键词
Compton imaging; single photon emission computed tomography; gamma detection; Anger camera; imaging radiotracers;
D O I
10.1088/1361-6560/aa926a
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Single photon emission computed tomography (SPECT) is at present one of the major techniques for non-invasive diagnostics in nuclear medicine. The clinical routine is mostly based on collimated cameras, originally proposed by Hal Anger. Due to the presence of mechanical collimation, detection efficiency and energy acceptance are limited and fixed by the system's geometrical features. In order to overcome these limitations, the application of Compton cameras for SPECT has been investigated for several years. In this study we compare a commercial SPECT-Anger device, the General Electric HealthCare Infinia system with a High Energy General Purpose (HEGP) collimator, and the Compton camera prototype under development by the French collaboration CLaRyS, through Monte Carlo simulations (GATE-GEANT4 Application for Tomographic Emission-version 7.1 and GEANT4 version 9.6, respectively). Given the possible introduction of new radio-emitters at higher energies intrinsically allowed by the Compton camera detection principle, the two detectors are exposed to point-like sources at increasing primary gamma energies, from actual isotopes already suggested for nuclear medicine applications. The Compton camera prototype is first characterized for SPECT application by studying the main parameters affecting its imaging performance: detector energy resolution and random coincidence rate. The two detector performances are then compared in terms of radial event distribution, detection efficiency and final image, obtained by gamma transmission analysis for the Anger system, and with an iterative List Mode-Maximum Likelihood Expectation Maximization (LM-MLEM) algorithm for the Compton reconstruction. The results show for the Compton camera a detection efficiency increased by a factor larger than an order of magnitude with respect to the Anger camera, associated with an enhanced spatial resolution for energies beyond 500 keV. We discuss the advantages of Compton camera application for SPECT if compared to present commercial Anger systems, with particular focus on dose delivered to the patient, examination time, and spatial uncertainties.
引用
收藏
页码:8794 / 8812
页数:19
相关论文
共 20 条
[1]   SCINTILLATION CAMERA [J].
ANGER, HO .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1958, 29 (01) :27-33
[2]  
ANGER HO, 1964, J NUCL MED, V5, P515
[3]  
[Anonymous], 2012, THESIS
[4]  
Bhat M. R., 1992, NUCL DATA SCI TECHNO, P817, DOI DOI 10.1007/978-3-642-58113-7_227
[5]   Analytical and experimental FWHM of a gamma camera: theoretical and practical issues [J].
Cecchin, Diego ;
Poggiali, Davide ;
Riccardi, Lucia ;
Turco, Paolo ;
Bui, Franco ;
De Marchi, Stefano .
PEERJ, 2015, 3
[6]   A quantum theory of the scattering of x-rays by light elements [J].
Compton, AH .
PHYSICAL REVIEW, 1923, 21 (05) :0483-0502
[7]   GAMMA-RADIATION IMAGING-SYSTEM BASED ON COMPTON-EFFECT [J].
EVERETT, DB ;
FLEMING, JS ;
TODD, RW ;
NIGHTINGALE, JM .
PROCEEDINGS OF THE INSTITUTION OF ELECTRICAL ENGINEERS-LONDON, 1977, 124 (11) :995-1000
[8]  
General Electrics HealthCare, 2006, INF 3 8 1 FREE GEOM
[9]   Statistical performance evaluation and comparison of a Compton medical imaging system and a collimated Anger camera for higher energy photon imaging [J].
Han, Li ;
Rogers, W. Leslie ;
Huh, Sam S. ;
Clinthorne, Neal .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (24) :7029-7045
[10]  
Hubbell J H, 1987, 873597 NBSIR NAT I S