Frequent embolization in peripheral angioplasty:: Detection with an embolism protection device (AngioGuard) and electron microscopy

被引:47
作者
König, CW
Pusich, B
Tepe, G
Wendel, HP
Hahn, U
Schneider, W
Claussen, CD
Duda, SH
机构
[1] Univ Tubingen, Dept Diagnost Radiol, D-72076 Tubingen, Germany
[2] Univ Tubingen, Clin Thorac & Cardiovasc Surg, D-72076 Tubingen, Germany
关键词
angioplasty; peripheral vessels; embolization; embolization protection; BALLOON ANGIOPLASTY; CAROTID ANGIOPLASTY; STENT PLACEMENT; CEREBRAL PROTECTION; PREVENTION; STENOSIS; EVENTS;
D O I
10.1007/s00270-003-2656-3
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose: To evaluate the deliverability and protection capabilities of an embolism protection filter in angioplasty of peripheral arteries. Methods: The Angioguard emboli capture guidewire system was applied in 11 patients with femoropopliteal lesions (6 stenoses, 3 occlusions, 2 controls). Data on lesion crossing, flow deceleration and macroembolization were recorded. Filter membranes were evaluated with scanning electron microscopy (SEM). Results: System delivery was successful in all patients. Primary lesion crossing was feasible in four of six stenoses; predilatation was required in two of six. Marked flow deceleration was recorded in six patients. Emboli next to the filter were detected in each patient with concentric plaques, but could not reliably be removed with the filter. Downstream macroembolization was also present in all patients with concentric stenoses, but in none with chronic occlusion. None of the patients had clinical signs of ischemia. SEM analysis demonstrated only small particles on control group filters and non-obliterating fibrinous conglomerates on filters used in chronic occlusion. Substantial obliteration was seen on several filters used in stenotic lesions. Conclusion: Microembolization of fibrin aggregates is a common incident in balloon angioplasty of femoropopliteal stenoses. Macroembolization occurred more frequently than previously reported. The use of embolism protection filters aided in the detection but not in the removal of larger emboli.
引用
收藏
页码:334 / 339
页数:6
相关论文
共 18 条
[1]   Asymptomatic peripheral embolism during balloon angioplasty [J].
Abraham, P ;
Lhoste, P ;
PascoPapon, A ;
Saumet, JL .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 166 (05) :1230-1231
[2]   Detection of distal emboli in patients with peripheral arterial stenosis before and after iliac angioplasty: A prospective study [J].
Al-Hamali, S ;
Baskerville, P ;
Fraser, S ;
Walters, H ;
Markus, HS .
JOURNAL OF VASCULAR SURGERY, 1999, 29 (02) :345-351
[3]   NONCORONARY ANGIOPLASTY [J].
BECKER, GJ ;
KATZEN, BT ;
DAKE, MD .
RADIOLOGY, 1989, 170 (03) :921-940
[4]   Push and park. An alternative strategy for management of embolic complication during balloon angioplasty [J].
Higginson, A ;
Alaeddin, F ;
Fishwick, G ;
Bolia, A .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2001, 21 (03) :279-282
[5]   Microemboli detected by transcranial Doppler monitoring in patients during carotid angioplasty versus carotid endarterectomy [J].
Jordan, WD ;
Voellinger, DC ;
Doblar, DD ;
Plyushcheva, NP ;
Fisher, WS ;
McDowell, HA .
CARDIOVASCULAR SURGERY, 1999, 7 (01) :33-38
[6]   Stent placement versus percutaneous transluminal angioplasty of human carotid arteries in cadavers in situ:: Distal embolization and findings at intravascular US, MR imaging, and histopathologic analysis' [J].
Manninen, HI ;
Räsänen, HT ;
Vanninen, RL ;
Vainio, P ;
Hippeläinen, M ;
Kosma, VM .
RADIOLOGY, 1999, 212 (02) :483-492
[7]   Role of the distal balloon protection technique in the prevention of cerebral embolic events during carotid stent placement [J].
Martin, JB ;
Pache, JC ;
Treggiari-Venzi, M ;
Murphy, KJ ;
Gailloud, P ;
Puget, E ;
Pizzolato, G ;
Sugiu, K ;
Guimaraens, L ;
Théron, J ;
Rüfenacht, DA .
STROKE, 2001, 32 (02) :479-484
[8]  
McCleary AJ, 1998, BRIT J SURG, V85, P771
[9]   COMPLICATIONS OF ANGIOPLASTY [J].
MCDERMOTT, JC ;
CRUMMY, AB .
SEMINARS IN INTERVENTIONAL RADIOLOGY, 1994, 11 (02) :145-149
[10]   Cerebral protection devices for treatment of carotid stenoses:: Where do we stand today? [J].
Müller-Hülsbeck, S .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (08) :945-954