Magnetic resonance imaging with cochlear implants and auditory brainstem implants: Are we truly practicing MRI safety?

被引:48
作者
Shew, Matthew [1 ]
Wichova, Helena [1 ]
Lin, James [1 ]
Ledbetter, Luke N. [2 ]
Staecker, Hinrich [1 ]
机构
[1] Univ Kansas, Sch Med, Dept Otolaryngol Head & Neck Surg, Kansas City, KS USA
[2] Univ Kansas, Sch Med, Dept Radiol, Kansas City, KS USA
关键词
Cochlear implant; MRI safety; magnet displacement; NF2; INTERNAL MAGNET; COMPATIBILITY; DISPLACEMENT;
D O I
10.1002/lary.27516
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objective Our objective is to evaluate the safety in patients with cochlear implants (CIs) and auditory brainstem implants (ABI) undergoing 1.5 Tesla (T) magnetic resonance imaging (MRI). Secondly, we want to raise awareness on CI and MRI safety, and advocate for continued improvement and advancement to minimize morbidity for our CI patients. Methods Retrospective case series from 2006 to 2018 at a single tertiary academic center. Data was collected on patients with CI or auditory brainstem implants undergoing MRI. Outcomes collected include demographic data, age at time of MRI, MRI characteristics, complications, CI manufacturer, and image quality. Results Eighteen patients with CI or ABI collectively underwent a total of 62 MRI scans. Five of 15 (33%) CI patients with magnet had complications: five total of 24 MRI scans (21%). Two patients had magnet removal prior to 29 MRI scans without complications. Four of five MRI-related complications were equipped with a U.S. Food and Drug Administration-approved head wrap. Three of five required a trip to the operating room to explore and reposition the CI magnet; two could not complete MRI secondary to pain. Of the complications, two were Cochlear (Sydney, Australia), two Advanced Bionics (Valencia, CA), and one MED-EL (Innsbruck, Austria). Synchrony model (MED-EL) had 0 of seven complications, with a total of 19 MRI scans, which features a freely rotating and self-aligning magnet. Conclusion Our series offers a diverse number of CI manufacturers and is in accordance with other literature that CI MRI-related adverse events are occurring at an unacceptable frequency. We can promote CI MRI safety through our institutions' MRI CI patient protocols, raise awareness that diagnostic MRI benefits must outweigh CI-related complications, and advocate for continued industry technological innovation.
引用
收藏
页码:482 / 489
页数:8
相关论文
共 28 条
[1]   MRI Information for Commonly Used Otologic Implants Review and Update [J].
Azadarmaki, Roya ;
Tubbs, Rhonda ;
Chen, Douglas A. ;
Shellock, Frank G. .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2014, 150 (04) :512-519
[2]   Clinical application of magnetic resonance imaging in 30 cochlear implant patients [J].
Baumgartner, WD ;
Youssefzadeh, S ;
Hamzavi, J ;
Czerny, C ;
Gstoettner, W .
OTOLOGY & NEUROTOLOGY, 2001, 22 (06) :818-822
[3]  
Broomfield Stephen J, 2013, Cochlear Implants Int, V14, P51
[4]   Magnetic Resonance Imaging With Cochlear Implant Magnet in Place: Safety and Imaging Quality [J].
Carlson, Matthew L. ;
Neff, Brian A. ;
Link, Michael J. ;
Lane, John I. ;
Watson, Robert E. ;
McGee, Kiaran P. ;
Bernstein, Matt A. ;
Driscoll, Colin L. W. .
OTOLOGY & NEUROTOLOGY, 2015, 36 (06) :965-971
[5]   Magnetic Resonance Imaging at 1.5 T After Cochlear Implantation [J].
Crane, Benjamin T. ;
Gottschalk, Barbara ;
Kraut, Michael ;
Aygun, Nafi ;
Niparko, John K. .
OTOLOGY & NEUROTOLOGY, 2010, 31 (08) :1215-1220
[6]  
Cuda D, 2013, ACTA OTORHINOLARYNGO, V33, P133
[7]   Cochlear implant magnet displacement during magnetic resonance imaging [J].
Deneuve, Sophie ;
Loundon, Nathalie ;
Leboulanger, Nicolas ;
Rouillon, Isabelle ;
Garabedian, Erea Noel .
OTOLOGY & NEUROTOLOGY, 2008, 29 (06) :789-790
[8]   Noninvasive management of cochlear implant's inner magnet displacement after magnetic resonance [J].
Di Nardo, Walter ;
Giannantonio, Sara ;
Schinaia, Lorenzo ;
De Corso, Eugenio ;
Paludetti, Gaetano .
LARYNGOSCOPE, 2013, 123 (03) :783-786
[9]   Should patients with brain implants undergo MRI? [J].
Erhardt, Johannes B. ;
Fuhrer, Erwin ;
Gruschke, Oliver G. ;
Leupold, Jochen ;
Wapler, Matthias C. ;
Hennig, Juergen ;
Stieglitz, Thomas ;
Korvink, Jan G. .
JOURNAL OF NEURAL ENGINEERING, 2018, 15 (04)
[10]   Safety study of the Cochlear Nucleus® 24 Device with internal magnet in the 1.5 Tesla magnetic resonance imaging scanner [J].
Gubbels, SP ;
McMenomey, SO .
LARYNGOSCOPE, 2006, 116 (06) :865-871