Thin microelectrodes reduce GFAP expression in the implant site in rodent somatosensory cortex

被引:65
作者
Stice, Paula [1 ]
Gilletti, Aaron
Panitch, Alyssa
Muthuswamy, Jit
机构
[1] Arizona State Univ, Harrington Dept Bioengn, Tempe, AZ 85287 USA
[2] Johnson & Johnson, Res & Dev, Boston, MA USA
关键词
D O I
10.1088/1741-2560/4/2/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The objective of this study was to test the hypothesis that neural implants with reduced cross-sectional areas will have less glial scarring associated with implantation injury in long-term experiments. In this study, we implanted nine adult rats with two different implants of 12 mu m (n = 6), and 25 mu m (n = 6) diameters (cross-sectional areas of 68 mu m(2), 232 mu m(2) respectively) and the expression of glial fibrilliary acidic protein (GFAP) was assessed after 2 weeks and 4 weeks of implantation. In order to facilitate implantation, the 12 Am diameter implants were coated with poly-glycolic acid (PGA), a biodegradable polymer that degraded within minutes of implantation. In n = 3 animals, 25 Am diameter implants also coated with PGA were implanted and assessed for GFAP expression at the end of 4 weeks of implantation. Statistical analysis of the GFAP expression around the different implants demonstrated that after 2 weeks of implantation there is no statistically significant difference in GFAP expression between the 12 Am and the 25 Am diameter implants. However, after 4 weeks of implantation the implant site of 12 Am diameter implants exhibited a statistically significant reduction in GFAP expression when compared to the implant sites of the 25 Am diameter implants (both with and without the PGA coating). We conclude that in neural implants that are tethered to the skull, implant cross-sectional areas of 68 Am 2 and smaller could lead to a reduced glial scarring under chronic conditions. Future studies with longer implant durations can confirm if this observation remains consistent beyond 4 weeks.
引用
收藏
页码:42 / 53
页数:12
相关论文
共 31 条
[1]   HISTOPATHOLOGIC EVALUATION OF PROLONGED INTRACORTICAL ELECTRICAL-STIMULATION [J].
AGNEW, WF ;
YUEN, TGH ;
MCCREERY, DB ;
BULLARA, LA .
EXPERIMENTAL NEUROLOGY, 1986, 92 (01) :162-185
[2]   STIMULATION OF THE VENTRAL INTERMEDIATE THALAMIC NUCLEUS IN TREMOR DOMINATED PARKINSONS-DISEASE AND ESSENTIAL TREMOR [J].
ALESCH, F ;
PINTER, MM ;
HELSCHER, RJ ;
FERTL, L ;
BENABID, AL ;
KOOS, WT .
ACTA NEUROCHIRURGICA, 1995, 136 (1-2) :75-81
[3]  
[Anonymous], 1990, NEURAL PROSTHESES FU
[4]   Micromechanical resonator array for an implantable bionic ear [J].
Bachman, M ;
Zeng, FG ;
Xu, T ;
Li, GP .
AUDIOLOGY AND NEURO-OTOLOGY, 2006, 11 (02) :95-103
[5]   Geometric control of cell life and death [J].
Chen, CS ;
Mrksich, M ;
Huang, S ;
Whitesides, GM ;
Ingber, DE .
SCIENCE, 1997, 276 (5317) :1425-1428
[6]   FACTORS INFLUENCING THE BIOCOMPATIBILITY OF INSERTABLE SILICON MICROSHAFTS IN CEREBRAL-CORTEX [J].
EDELL, DJ ;
TOI, VV ;
MCNEIL, VM ;
CLARK, LD .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (06) :635-643
[7]   The glial scar and central nervous system repair [J].
Fawcett, JW ;
Asher, RA .
BRAIN RESEARCH BULLETIN, 1999, 49 (06) :377-391
[8]  
Fitch MT, 1999, J NEUROSCI, V19, P8182
[9]   Brain micromotion around implants in the rodent somatosensory cortex [J].
Gilletti, Aaron ;
Muthuswamy, Jit .
JOURNAL OF NEURAL ENGINEERING, 2006, 3 (03) :189-195
[10]   Differential degradation rates in vivo and in vitro of biocompatible poly(lactic acid) and poly(glycolic acid) homo- and co-polymers for a polymeric drug-delivery microchip [J].
Grayson, ACR ;
Voskerician, G ;
Lynn, A ;
Anderson, JM ;
Cima, MJ ;
Langer, R .
JOURNAL OF BIOMATERIALS SCIENCE-POLYMER EDITION, 2004, 15 (10) :1281-1304