Technical Feasibility and Design of a Shape Memory Alloy Support Device to Increase Ejection Fraction in Patients with Heart Failure

被引:4
作者
Aarnink, K. M. [1 ,2 ]
Halfwerk, F. R. [3 ,4 ]
Said, S. A. M. [5 ]
Grandjean, J. G. [3 ,4 ]
Paulusse, J. M. J. [6 ,7 ]
机构
[1] Univ Twente, TechMed Ctr, Enschede, Netherlands
[2] Isala Hosp, Zwolle, Netherlands
[3] Med Spectrum Twente Hosp, Thorax Ctr Twente, Dept Cardiothorac Surg, POB 50 000, NL-7500 KA Enschede, Netherlands
[4] Univ Twente, Dept Biomech Engn, TechMed Ctr, Enschede, Netherlands
[5] Hosp Grp Twente, Dept Cardiol, Almelo Hengelo, Netherlands
[6] Univ Twente, Dept Biomol Nanotechnol, Enschede, Netherlands
[7] Univ Med Ctr Groningen, Dept Nucl Med & Mol Imaging, Groningen, Netherlands
关键词
Shape memory alloy; Heart failure; Ejection fraction; Cardiac support device; Design configuration;
D O I
10.1007/s13239-018-00399-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
PurposeHeart failure is increasingly prevalent in the elderly. Treatment of patients with heart failure aims at improving their clinical condition, quality of life, prevent hospital (re)admissions and reduce mortality. Unfortunately, only a select group of heart failure patients with reduced ejection fraction are eligible for Cardiac Resynchronization Therapy where 30-40% remain non-responders and need left ventricular support. The aim of this study is to investigate if a shape memory alloy (SMA) is able to increase the ejection fraction of a mono-chamber static heart model by 5%.MethodsA pediatric ventilation balloon was used as a heart model (mono-chamber). Flexinol((R)), a SMA, was placed around the heart model in multiple configurations and activated using pulse width modulation techniques to determine influence of diameter and configuration on volume displacement. Furthermore, pressure within the heart model was measured with a custom-made pressure sensor.ResultsSMA with a diameter of 0.38mm, placed in a spiral shape and activated with a duty cycle of 80% and a frequency of 50/min gave the highest ejection fraction increase of 3.5%.ConclusionsThis study demonstrated the feasibility of volume displacement in a static heart model by activation of SMA-wires. Configuration, duty cycle, frequency, pulse intervals and diameter were identified as important factors affecting the activation of SMA-wires on volume displacement. Future research should include the use of parallel SMA-wires, prototype testing in dynamic or ex vivo bench models.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 26 条
[1]   Incidence and prognosis of vascular complications after percutaneous placement of left ventricular assist device [J].
Abaunza, Miriam ;
Kabbani, Loay S. ;
Nypaver, Timothy ;
Greenbaum, Adam ;
Balraj, Praveen ;
Qureshi, Sherazuddin ;
Alqarqaz, Mohammed A. ;
Shepard, Alexander D. .
JOURNAL OF VASCULAR SURGERY, 2015, 62 (02) :417-423
[2]  
[Anonymous], 2011, TECHNICAL CHARACTERI
[3]   Non-Responders to Cardiac Resynchronization Therapy - The Magnitude of the Problem and the Issues [J].
Auricchio, Angelo ;
Prinzen, Frits W. .
CIRCULATION JOURNAL, 2011, 75 (03) :521-527
[4]  
Benjamin EJ, 2017, CIRCULATION, V135, pE146, DOI [10.1161/CIR.0000000000000558, 10.1161/CIR.0000000000000485, 10.1161/CIR.0000000000000530]
[5]   Precision instrument for characterizing shape memory alloy wires in bias spring actuation [J].
Chikkamaranahalli, S ;
Vallance, RR ;
Khan, A ;
Marsh, ER ;
Rawashdeh, OA ;
Lumpp, JE ;
Walcott, BL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2005, 76 (06)
[6]   Trends in heart failure hospitalizations [J].
Fida N. ;
Piña I.L. .
Current Heart Failure Reports, 2012, 9 (4) :346-353
[7]   SmartFlexA® NiTi Wires for Shape Memory Actuators [J].
Fumagalli, L. ;
Butera, F. ;
Coda, A. .
JOURNAL OF MATERIALS ENGINEERING AND PERFORMANCE, 2009, 18 (5-6) :691-695
[8]   A Contemporary Appraisal of the Heart Failure Epidemic in Olmsted County, Minnesota, 2000 to 2010 [J].
Gerber, Yariv ;
Weston, Susan A. ;
Redfield, Margaret M. ;
Chamberlain, Alanna M. ;
Manemann, Sheila M. ;
Jiang, Ruoxiang ;
Killian, Jill M. ;
Roger, Veronique L. .
JAMA INTERNAL MEDICINE, 2015, 175 (06) :996-1004
[9]  
Hassoulas IA, 2010, HELL J CARDIOL, V51, P301
[10]   Shape memory materials [J].
Huang, W. M. ;
Ding, Z. ;
Wang, C. C. ;
Wei, J. ;
Zhao, Y. ;
Purnawali, H. .
MATERIALS TODAY, 2010, 13 (7-8) :54-61