Characterization of occlusal splint materials: CAD-CAM versus conventional resins

被引:34
作者
Gibreel, Mona [1 ,2 ]
Perea-Lowery, Leila [3 ]
Vallittu, Pekka K. [4 ]
Lassila, Lippo [2 ]
机构
[1] Univ Turku, Inst Dent, Dept Biomat Sci, Turku, Finland
[2] Univ Turku, Inst Dent, Turku Clin Biomat Ctr TCBC, Turku, Finland
[3] Univ Turku, Turku Clin Biomat Ctr TCBC, Inst Dent, Dept Biomat Sci, Turku, Finland
[4] Univ Turku, Biomat Sci Dept, Welf Div, Turku, Finland
关键词
Splints; Polymethyl methacrylate; Flexural strength; Modulus; Toughness; Water sorption; MECHANICAL-PROPERTIES; CAD/CAM POLYMERS; ORAL APPLIANCES; MANAGEMENT; THERAPY; WEAR; PAIN;
D O I
10.1016/j.jmbbm.2021.104813
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Aim: The aim of this in vitro study was to assess the mechanical properties of five commercially available subtractive computer-aided design and computer-aided manufacturing (CAD-CAM) milled splint materials, as well as to compare them with conventional heat-polymerized and autopolymerizing resins used in the construction of conventional splints. Material and methods: Five CAD-CAM milled (ProArt CAD Splint, Therapon Transpa, Temp Premium Flexible Transpa, Cast, and Aqua), one autopolymerizing (Palapress), and one heat-polymerized (Paladon 65) resin materials were evaluated. Flexural strength, E-modulus, Vickers hardness, fracture toughness, fracture work, water sorption, and water solubility were measured. Samples were evaluated after dry and water storage for 30 days at 37 degrees C. Data were collected and statistically analyzed. Results: Under both storage circumstances, the flexural strength values of Paladon 65, Therapon Transpa, Temp Premium Flexible Transpa, and Aqua were statistically non-significant (P=0.055). The polycarbonate-based CAD-CAM material Temp Premium Flexible Transpa had the highest statistically significant values of the fracture toughness and fracture work (P<0.001). Moreover, it exhibited the lowest percentages of water sorption and water solubility among the investigated materials (P<0.001). All of the CAD-CAM materials exhibited dry elastic moduli greater than Palapress and lower than Paladon 65. One of the CAD-CAM materials, Cast, had the highest dry Vickers hardness value, which was non-significant when compared to Therapon Transpa (P=0.762). Conclusion: CAD-CAM polycarbonate-based splint materials exhibit higher fracture toughness and fracture work as well as lower water sorption and solubility than polymethyl methacrylate-based ones. The mechanical characteristics of the assessed CAD-CAM milled splint materials were not typically superior to those of the conventional heat-polymerized resin. However, some of them outperformed the autopolymerizing acrylic resin in terms of flexural strength, surface microhardness, water sorption, and water solubility.
引用
收藏
页数:7
相关论文
共 50 条
[1]   A novel in vitro approach to assess the fit of implant frameworks [J].
Abduo, Jaafar ;
Bennani, Vincent ;
Lyons, Karl ;
Waddell, Neil ;
Swain, Michael .
CLINICAL ORAL IMPLANTS RESEARCH, 2011, 22 (06) :658-663
[2]   A Comparison of the Surface Properties of CAD/CAM and Conventional Polymethylmethacrylate (PMMA) [J].
Al-Dwairi, Ziad N. ;
Tahboub, Kawkab Y. ;
Baba, Nadim Z. ;
Goodacre, Charles J. ;
Ozcan, Mutlu .
JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2019, 28 (04) :452-457
[3]  
Ali Isma Liza, 2008, J Prosthodont, V17, P545, DOI 10.1111/j.1532-849X.2008.00357.x
[4]   Comparison of Flexural Strength of Different CAD/CAM PMMA-Based Polymers [J].
Alp, Gulce ;
Murat, Sema ;
Yilmaz, Burak .
JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2019, 28 (02) :E491-E495
[5]   Fracture strength of temporary fixed partial dentures: CAD/CAM versus directly fabricated restorations [J].
Alt, Vanessa ;
Hannig, Matthias ;
Woestmann, Bernd ;
Balkenhol, Markus .
DENTAL MATERIALS, 2011, 27 (04) :339-347
[6]   Surface roughness and wear behavior of occlusal splint materials made of contemporary and high-performance polymers [J].
Benli, Merve ;
Gumus, Beril Eker ;
Kahraman, Yusuf ;
Gokcen-Rohlig, Bilge ;
Evlioglu, Gulumser ;
Huck, Olivier ;
Ozcan, Mutlu .
ODONTOLOGY, 2020, 108 (02) :240-250
[7]   Comparing the mechanical properties of pressed, milled, and 3D-printed resins for occlusal devices [J].
Berli, Constantin ;
Thieringer, Florian M. ;
Sharma, Neha ;
Muller, Johannes A. ;
Dedem, Philipp ;
Fischer, Jens ;
Rohr, Nadja .
JOURNAL OF PROSTHETIC DENTISTRY, 2020, 124 (06) :780-786
[8]   Mechanical and structural characterization of discontinuous fiber-reinforced dental resin composite [J].
Bijelic-Donova, Jasmina ;
Garoushi, Sufyan ;
Lassila, Lippo V. J. ;
Keulemans, Filip ;
Vallittu, Pekka K. .
JOURNAL OF DENTISTRY, 2016, 52 :70-78
[9]  
BOERO RP, 1989, ANGLE ORTHOD, V59, P165
[10]  
Craig R.G, 2019, RESTORATIVE DENT MAT, V14th, P29