Qualification of a chemotherapy-compounding robot

被引:14
作者
Jobard, Marion [1 ]
Brandely-Piat, Marie-Laure [1 ]
Chast, Francois [1 ]
Batista, Rui [1 ]
机构
[1] Hop Univ Paris Ctr, AP HP, Serv Pharm Clin, Paris, France
关键词
Automated compounding; robot; qualification; chemotherapy preparations;
D O I
10.1177/1078155219843322
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
KIRO (R) Oncology (Kiro Grifols, Spain) is a robotic system for automated compounding of sterile injectable drugs including intravenous cytotoxic treatments. The present article describes the qualification procedure applied prior to production phases. Peristaltic pumps which ensure the reconstitution of drugs were tested with water and NaCl 0.9%. The performance of the robot (accuracy and precision) to prepare bags, syringes and elastomeric pumps was evaluated with three placebo solutions (aqueous, foaming and viscous) using gravimetric controls. Microbiological controls were also performed. The pumps met the requirements set for volumes ranging from 5 to 100 mL. A total of 274 preparations was compounded. For the bags, the filling accuracy was within the limit of +/- 10% from 1 to 48 mL with aqueous solution, from 0.6 to 48 mL with foaming solution and from 5 to 48 mL with viscous solution. For all syringes and elastomeric pumps, it was within the limit of +/- 10%. The precision was validated for all preparations, except for bags and syringes prepared with 0.6 and 0.25 mL, respectively. The samples of surfaces and air complied with ISO 5 class environment. Among the 24 gloves tests performed, two presented microbiological growth. All Media fill tests were validated. The qualification procedure led us to exclude injections of any active principle volume strictly lower than 1 mL. The microbiological contamination of operators' gloves remains a critical point. Our operators will be made aware of the issue during the training period.
引用
收藏
页码:312 / 324
页数:13
相关论文
共 12 条
[1]  
Agence nationale de securite du medicament et des produits de sante (ANSM), 2007, BONNES PRATIQUES PRE
[2]  
[Anonymous], 124692000 NF EN
[3]  
European Pharmacopoeia, 2017, EDQM
[4]  
Institut National du Cancer (INCa), 2017, CHIM ANT
[5]  
International Organization for Standardization (iso), 2015, Cleanrooms and Associated Controlled Environments, Part 1: Classification of Air Cleanliness by Particle Concentration, V2nd
[6]   Performance evaluation of the compounding robot, APOTECAchemo, for injectable anticancer drugs in a Japanese hospital [J].
Takuya Iwamoto ;
Takuya Morikawa ;
Miki Hioki ;
Hirofumi Sudo ;
Demis Paolucci ;
Masahiro Okuda .
Journal of Pharmaceutical Health Care and Sciences, 3 (1)
[7]   Media-fill simulation tests in manual and robotic aseptic preparation of injection solutions in syringes [J].
Kraemer, Irene ;
Federici, Matteo ;
Kaiser, Vanessa ;
Thiesen, Judith .
JOURNAL OF ONCOLOGY PHARMACY PRACTICE, 2016, 22 (02) :195-204
[8]   A three-year study of a first-generation chemotherapy-compounding robot [J].
Nurgat, Zubeir ;
Faris, Dima ;
Mominah, Maher ;
Vibar, Arris ;
Al-Jazairi, Abdulrazaq ;
Ewing, Sheena ;
Ashour, Mohammed ;
Qaisi, Shrouq Kamel ;
Balhareth, Sakra ;
Al-Jedai, Ahmed .
AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 2015, 72 (12) :1036-1045
[9]  
Schierl Rudolf, 2016, J Oncol Pharm Pract, V22, P37, DOI 10.1177/1078155214551316
[10]   Environmental contamination, product contamination and workers exposure using a robotic system for antineoplastic drug preparation [J].
Sessink, Paul J. M. ;
Leclercq, Gisele M. ;
Wouters, Dominique-Marie ;
Halbardier, Loic ;
Hammad, Chaima ;
Kassoul, Nassima .
JOURNAL OF ONCOLOGY PHARMACY PRACTICE, 2015, 21 (02) :118-127