Accurate, reproducible measurement of acetone concentration in breath using selected ion flow tube-mass spectrometry

被引:30
作者
Dummer, Jack F. [1 ,2 ]
Storer, Malina K. [3 ]
Hu, Wan-Ping [3 ]
Swanney, Maureen P. [1 ]
Milne, Gordon J. [3 ]
Frampton, Christopher M. [2 ]
Scotter, Jenny M. [3 ]
Prisk, G. Kim [4 ,5 ]
Epton, Michael J. [1 ]
机构
[1] Christchurch Hosp, Resp Serv, Christchurch, New Zealand
[2] Univ Otago, Dept Med, Christchurch, New Zealand
[3] Syft Technol Ltd, Christchurch, New Zealand
[4] Univ Calif San Diego, Dept Med, San Diego, CA 92103 USA
[5] Univ Calif San Diego, Dept Radiol, San Diego, CA 92103 USA
关键词
TRACE GAS-ANALYSIS; SIFT; AIR;
D O I
10.1088/1752-7155/4/4/046001
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Using selected ion flow tube-mass spectrometry (SIFT-MS) for on-line analysis, we aimed to define the optimal single-exhalation breathing manoeuvre from which a measure of expired acetone concentration could be obtained. Using known acetone concentrations in vitro, we determined the instrument's accuracy, inter-measurement variability and dynamic response time. Further, we determined the effects of expiratory flow and volume on acetone concentration in the breath of 12 volunteers and calculated intra-individual coefficients of variation (CVs). At acetone concentrations of 600-3000 ppb on 30 days over 2 months there was an instrument measurement bias of 8% that did not change over time, inter-day and intra-day CVs were 5.6% and 0.0%, respectively, and the 10-90% response time was 500 +/- 50 ms (mean +/- SE). Acetone concentrations at exhalation flows of 193 +/- 18 (mean +/- SD) and 313 +/- 32 ml s(-1) were 619 +/- 1.83 (geometric mean +/- logSD) and 618 +/- 1.82 ppb in the fraction 70-85% by volume of exhaled vital capacity (V70-85%) and 636 +/- 1.82 (geometric mean +/- logSD) and 631 +/- 1.83 ppb in V85-100%. A difference was observed between acetone concentrations in the V70-85% and V85-100% fractions (p < 0.01), but flow had no effect. Median intra-individual CVs were 1.6-2.6%. On-line SIFT-MS measurement of acetone concentration in a single exhalation requires control of exhaled volume but not flow, and yields low intra-individual CVs and is potentially useful in approximating blood glucose and monitoring metabolic stress.
引用
收藏
页数:8
相关论文
共 24 条
[1]   Measuring airway exchange of endogenous acetone using a single-exhalation breathing maneuver [J].
Anderson, JC ;
Lamm, WJE ;
Hlastala, MP .
JOURNAL OF APPLIED PHYSIOLOGY, 2006, 100 (03) :880-889
[2]   CORRECTING FOR THE DYNAMIC-RESPONSE OF A RESPIRATORY MASS-SPECTROMETER [J].
BATES, JHT ;
PRISK, GK ;
TANNER, TE ;
MCKINNON, AE .
JOURNAL OF APPLIED PHYSIOLOGY, 1983, 55 (03) :1015-1022
[3]  
DUMMER J, 2007, AM J RESP CRIT CARE, V175, pA683
[4]   Defining the role of exhaled nitric oxide measurements in COPD [J].
Dummer, Jack F. ;
Taylor, D. R. .
RESPIROLOGY, 2010, 15 (03) :385-386
[5]   Hydrogen Cyanide as a Biomarker for Pseudomonas Aeruginosa in the Breath of Children With Cystic Fibrosis [J].
Enderby, Beth ;
Smith, David ;
Carroll, W. ;
Lenney, W. .
PEDIATRIC PULMONOLOGY, 2009, 44 (02) :142-147
[6]   GeoVOC: A SIFT-MS method for the analysis of small linear hydrocarbons of relevance to oil exploration [J].
Francis, G. J. ;
Wilson, P. F. ;
Milligan, D. B. ;
Langford, V. S. ;
McEwan, M. T. .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2007, 268 (01) :38-46
[7]  
Galassetti Pietro R, 2005, Diabetes Technol Ther, V7, P115, DOI 10.1089/dia.2005.7.115
[8]   The impact of breathing pattern and lung size on the alcohol breath test [J].
Hlastala, Michael P. ;
Anderson, Joseph C. .
ANNALS OF BIOMEDICAL ENGINEERING, 2007, 35 (02) :264-272
[9]  
KUNDU SK, 1993, CLIN CHEM, V39, P87
[10]   Standardisation of spirometry [J].
Miller, MR ;
Hankinson, J ;
Brusasco, V ;
Burgos, F ;
Casaburi, R ;
Coates, A ;
Crapo, R ;
Enright, P ;
van der Grinten, CPM ;
Gustafsson, P ;
Jensen, R ;
Johnson, DC ;
MacIntyre, N ;
McKay, R ;
Navajas, D ;
Pedersen, OF ;
Pellegrino, R ;
Viegi, G ;
Wanger, J .
EUROPEAN RESPIRATORY JOURNAL, 2005, 26 (02) :319-338