Purpose: To investigate the applicability of the scanning version of dual-beam partial coherence interferometry (PCI) for measuring the anterior segment and axial length of pseudophakic eyes in a clinical setting and to determine the achievable precision with this biometry technique. Setting: Department of Ophthalmology, Vienna General Hospital, and Institute of Medical Physics, University of Vienna, Austria. Methods: Partial coherence interferometry was performed in 39 pseudophakic eyes of 39 patients after implantation of a foldable acrylic intraocular lens (IOL). Results: Effective lens position (ELP), IOL thickness, and lens-capsule distance (LCD) were determined with a precision of 2 to 3 mu m; corneal thickness and axial eye length, with a precision of 0.8 and 5.0 mu m, respectively The mean ELF of the IOL was 4.093 mm +/- 0.290 (SD). In 7 eyes (18%), a positive LCD of 68 +/- 40 mu m was detected with PCI. Mean corneal thickness was 526.4 +/- 31.5 mu m; mean IOL thickness, 791.5 +/- 40.2 mu m; and mean axial length, 23.388 +/- 0.824 mm. Conclusion: The scanning version of PCI enables high precision (less than or equal to 5 mu m) and high resolution (similar to 12 mu m) biometry of pseudophakic eyes that is better than conventional ultrasound by a factor of more than 20. For the first time, positive LCD, a possible risk factor for posterior capsule opacification: could be detected;nd quantified. Furthermore, this technique offers a high degree of comfort for the patient since it is a noncontract method with no need for local anesthesia or pupil dilation and has a reduced risk of corneal infection.
引用
收藏
页码:1087 / 1093
页数:7
相关论文
共 33 条
[1]  
*AM NAT STAND I, 1986, 1361 ANSI Z
[2]  
Bamber JC., 1988, PHYS MED IMAGING, P319
[3]   SIGNAL AND RESOLUTION ENHANCEMENTS IN DUAL BEAM OPTICAL COHERENCE TOMOGRAPHY OF THE HUMAN EYE [J].
Baumgartner, Angela ;
Hitzenberger, Christoph K. ;
Sattmann, Harald ;
Drexler, Wolfgang ;
Fercher, Adolf F. .
JOURNAL OF BIOMEDICAL OPTICS, 1998, 3 (01) :45-54
[4]   Investigation of dispersion effects in ocular media by multiple wavelength partial coherence interferometry [J].
Drexler, W ;
Hitzenberger, CK ;
Baumgartner, A ;
Findl, O ;
Sattmann, H ;
Fercher, AF .
EXPERIMENTAL EYE RESEARCH, 1998, 66 (01) :25-33
[5]  
Drexler W, 1997, INVEST OPHTH VIS SCI, V38, P1304
[6]   Biometric investigation of changes in the anterior eye segment during accommodation [J].
Drexler, W ;
Baumgartner, A ;
Findl, O ;
Hitzenberger, CK ;
Fercher, AF .
VISION RESEARCH, 1997, 37 (19) :2789-2800
[7]   MEASUREMENT OF THE THICKNESS OF FUNDUS LAYERS BY PARTIAL COHERENCE TOMOGRAPHY [J].
DREXLER, W ;
HITZENBERGER, CK ;
SATTMANN, H ;
FERCHER, AF .
OPTICAL ENGINEERING, 1995, 34 (03) :701-710
[8]  
DREXLER W, IN PRESS INVEST OPHT
[9]  
DREXLER W, IN PRESS AM J OPHTHA
[10]  
Drexler W, 1993, P SOC PHOTO-OPT INS, V2083, P363