Comparison of clinical performance of size 1.5 Supreme LMA and Proseal LMA among Asian children: a randomized controlled trial

被引:7
作者
Chaw, Sook Hui [1 ]
Shariffuddin, Ina I. [1 ]
Foo, Li Lian [1 ]
Lee, Pui Kuan [2 ]
Paran, Ramona Maya [1 ]
Cheang, Peak Chee [1 ]
Chan, Lucy [3 ]
机构
[1] Univ Malaya, Dept Anaesthesiol, Fac Med, Kuala Lumpur 50603, Malaysia
[2] Univ Malaya, Med Ctr, Dept Anaesthesiol, Kuala Lumpur, Malaysia
[3] Univ Malaya, Fac Dent, Dept Oromaxillofacial Surg & Med Sci, Kuala Lumpur, Malaysia
关键词
Supraglottic airway device; Paediatric airway management; Paediatric anaesthesia; Laryngeal mask airway; LARYNGEAL MASK AIRWAY; PEDIATRIC SURGICAL-PATIENTS; SUPRAGLOTTIC AIRWAYS; COHORT EVALUATION; INSERTION; PRESSURE; INFANTS; ADULTS;
D O I
10.1007/s10877-018-0109-4
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
To date, most of the studies on safety and efficacy of supraglottic airway devices were done in Caucasian patients, and the results may not be extrapolated to Asian patients due to the different airway anatomy. We conducted this study to compare Supreme LMA (SLMA) and Proseal LMA (PLMA) size 1.5 in anaesthetized children among an Asian population. This prospective randomized clinical trial was conducted in a tertiary teaching hospital from September 2013 until May 2016. Sixty children, weighing 5-10kg, who were scheduled for elective surgery under general anaesthesia were recruited and completed the study. Patients were randomly assigned to have either SLMA or PLMA as the airway device for general anaesthesia, and standard anaesthesia protocol was followed. The primary outcome measured was the oropharyngeal leak pressure (OLP). The rate of successful insertion, insertion time, fibreoptic view of larynx and airway complications for each device were also assessed. There were no statistically significant differences between SLMA and PLMA size 1.5 in oropharyngeal leak pressure [19.1 (+/- 5.5) cmH(2)O vs. 19.8 (+/- 4.5) cmH(2)O, p=0.68]. Secondary outcomes including time to insertion [20.8 (+/- 8.3) vs. 22.1 (+/- 8.3) s, p=0.57], first attempt success rate for device insertion, fibreoptic view of larynx, and airway complications were also comparable between the two devices. We found that all the patients who had a failed device insertion (either PLMA or SLMA) were of a smaller size (5-6.2kg). The oropharyngeal leak pressure of the SLMA 1.5 was comparable with the PLMA 1.5, and both devices were able to maintain an airway effectively without significant clinical complications in anaesthetized children from an Asian population.
引用
收藏
页码:1093 / 1099
页数:7
相关论文
共 28 条
[1]   Ease of insertion of the laryngeal mask airway in pediatric surgical patients: Predictors of failure and outcome [J].
Asida, S. M. ;
Ahmed, S. S. .
SAUDI JOURNAL OF ANAESTHESIA, 2016, 10 (03) :295-300
[2]   A multicenter study comparing the ProSeal™ and Classic™ laryngeal mask airway in anesthetized, nonparalyzed patients [J].
Brimacombe, J ;
Keller, C ;
Fullekrug, B ;
Agrò, F ;
Rosenblatt, W ;
Dierdorf, SF ;
de Lucas, EG ;
Capdevilla, X ;
Brimacombe, N .
ANESTHESIOLOGY, 2002, 96 (02) :289-295
[3]  
Chinachoti Thitima, 2016, Journal of the Medical Association of Thailand, V99, P811
[4]   Gum elastic bougie-guided insertion of the ProSeal laryngeal mask airway is superior to the digital and introducer tool techniques in patients with simulated difficult laryngoscopy using a rigid neck collar [J].
Eschertzhuber, Stephan ;
Brimacombe, Joseph ;
Hohlrieder, Matthias ;
Stadlbauer, Karl-Heinz ;
Keller, Christian .
ANESTHESIA AND ANALGESIA, 2008, 107 (04) :1253-1256
[5]   The size 11/2 ProSeal™ laryngeal mask airway in infants:: A randomized, crossover investigation with the Classic™ laryngeal mask airway [J].
Goldmann, K ;
Roettger, C ;
Wulf, H .
ANESTHESIA AND ANALGESIA, 2006, 102 (02) :405-410
[6]   A randomized crossover comparison of the size 2 1/2 laryngeal mask airway ProSeal™ versus Laryngeal Mask Airway-Classic™ in pediatric patients [J].
Goldmann, K ;
Jakob, C .
ANESTHESIA AND ANALGESIA, 2005, 100 (06) :1605-1610
[7]   RADIOLOGIC LOCALIZATION OF THE LARYNGEAL MASK AIRWAY IN CHILDREN [J].
GOUDSOUZIAN, NG ;
DENMAN, W ;
ANAES, FC ;
CLEVELAND, R ;
SHORTEN, G .
ANESTHESIOLOGY, 1992, 77 (06) :1085-1089
[8]   Comparison of craniofacial characteristics of typical Chinese and Caucasian young adults [J].
Gu, Yan ;
McNamara, James A., Jr. ;
Sigler, Lauren M. ;
Baccetti, Tiziano .
EUROPEAN JOURNAL OF ORTHODONTICS, 2011, 33 (02) :205-211
[9]   Comparison of ProSeal LMA with Supreme LMA in paediatric patients [J].
Hosten, T. ;
Gurkan, Y. ;
Kus, A. ;
Ozdamar, D. ;
Aksu, C. ;
Solak, M. ;
Toker, K. .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2013, 57 (08) :996-1001
[10]   A new supraglottic airway device: LMA-Supreme™, comparison with LMA-Proseal™ [J].
Hosten, T. ;
Gurkan, Y. ;
Ozdamar, D. ;
Tekin, M. ;
Toker, K. ;
Solak, M. .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2009, 53 (07) :852-857