The ACURATE neo transcatheter aortic valve system

被引:17
作者
Choudhury, Tawfiq [1 ]
Solomonica, Amir [1 ]
Bagur, Rodrigo [1 ,2 ]
机构
[1] London Hlth Sci Ctr, London, ON, Canada
[2] Univ Keele, Inst Primary Care & Hlth Sci, Ctr Prognosis Res, Keele Cardiovasc Res Grp, Stoke On Trent, Staffs, England
关键词
TAVI; TAVR; self-expanding; aortic stenosis; transfemoral; transapical; SAVI-TF REGISTRY; OUTCOMES; BIOPROSTHESIS; IMPLANTATION; EXPERIENCE;
D O I
10.1080/17434440.2018.1526675
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Introduction: The field of transcatheter aortic valve implantation (TAVI) has seen an exponential expansion in TAVI devices for the treatment of severe symptomatic aortic stenosis.Areas covered: The ACURATE neo (Boston Scientific, MA, USA) is a second-generation self-expanding transcatheter aortic valve system. The valve presents a supra-annular design, and comprises three stabilization arches for axial alignment, an upper-crown for capping the aortic annulus, and a lower-crown that is opened over the native aortic valve for full deployment. This device minimally protrudes into the left ventricular outflow-tract, thereby minimizing the risk of conduction abnormalities. The ACURATE neo can be delivered by transfemoral and transapical approaches with a unique top-down' deployment. Information was gathered from published manuscripts obtained using standard databases (MEDLINE) and trial data presented at conferences.Expert commentary: Data from observational series show that TAVI with the ACURATEneo is associated with low rates of all-cause mortality, paravalvular leak, and permanent pacemaker requirement. Randomized data comparing the ACURATE neo valve with contemporary valves and long-term durability data are awaited. Continuing modifications in device and delivery system design are anticipated with a view to further reduction in paravalvular leak.
引用
收藏
页码:693 / 699
页数:7
相关论文
共 15 条
[1]   Transcatheter aortic valve implantation with the repositionable and fully retrievable Lotus Valve System™ [J].
Bagur, Rodrigo ;
Choudhury, Tawfiq ;
Mamas, Mamas A. .
JOURNAL OF THORACIC DISEASE, 2017, 9 (09) :2798-2803
[2]   Importance of the valve durability-life expectancy ratio in selection of a prosthetic aortic valve [J].
Bagur, Rodrigo ;
Pibarot, Philippe ;
Otto, Catherine M. .
HEART, 2017, 103 (22) :1756-1759
[3]   First North American experience with the transfemoral ACURATE-neoTM self-expanding transcatheter aortic bioprosthesis [J].
Bagur, Rodrigo ;
Teefy, Patrick J. ;
Kiaii, Bob ;
Diamantouros, Pantelis ;
Chu, Michael W. A. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2017, 90 (01) :130-138
[4]   Transfemoral transcatheter ACURATE-neo™ aortic valve replacement in a patient with a previous mechanical mitral valve [J].
Bagur, Rodrigo ;
Pestrichella, Vincenzo ;
Montesanti, Rosamaria ;
Alemanni, Rossella ;
Cassese, Mauro .
JOURNAL OF CARDIAC SURGERY, 2017, 32 (06) :358-360
[5]   Transcatheter Aortic Valve Implantation With or Without Preimplantation Balloon Aortic Valvuloplasty: A Systematic Review and Meta-Analysis [J].
Bagur, Rodrigo ;
Kwok, Chun Shing ;
Nombela-Franco, Luis ;
Ludman, Peter F. ;
de Belder, Mark A. ;
Sponga, Sandro ;
Gunning, Mark ;
Nolan, James ;
Diamantouros, Pantelis ;
Teefy, Patrick J. ;
Kiaii, Bob ;
Chu, Michael W. A. ;
Mamas, Mamas A. .
JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2016, 5 (06)
[6]   Early clinical outcomes of a novel self-expanding transapical transcatheter aortic valve bioprosthesis [J].
Chu, Michael W. A. ;
Bagur, Rodrigo ;
Losenno, Katie L. ;
Jones, Philip M. ;
Diamantouros, Pantelis ;
Teefy, Patrick ;
Gelinas, Jill J. ;
Kiaii, Bob .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2017, 153 (04) :810-818
[7]   Prospective multicentre evaluation of a novel, low-profile transapical delivery system for self-expandable transcatheter aortic valve implantation: 6-month outcomes [J].
Conradi, Lenard ;
Hilker, Michael ;
Kempfert, Joerg ;
Boergermann, Jochen ;
Treede, Hendrik ;
Holzhey, David M. ;
Schroefel, Holger ;
Kim, Won-Keun ;
Schaefer, Ulrich ;
Walther, Thomas .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2018, 54 (04) :762-767
[8]   Structural valve deterioration after transcatheter aortic valve implantation [J].
Foroutan, Farid ;
Guyatt, Gordon H. ;
Otto, Catherine M. ;
Siemieniuk, Reed A. ;
Schandelmaier, Stefan ;
Agoritsas, Thomas ;
Vandvik, Per O. ;
Bhagra, Sai ;
Bagur, Rodrigo .
HEART, 2017, 103 (23) :1899-1905
[9]   Updated standardized endpoint definitions for transcatheter aortic valve implantation: the Valve Academic Research Consortium-2 consensus document (VARC-2) [J].
Kappetein, Arie Pieter ;
Head, Stuart J. ;
Genereux, Philippe ;
Piazza, Nicolo ;
van Mieghem, Nicolas M. ;
Blackstone, Eugene H. ;
Brott, Thomas G. ;
Cohen, David J. ;
Cutlip, Donald E. ;
van Es, Gerrit-Anne ;
Hahn, Rebecca T. ;
Kirtane, Ajay J. ;
Krucoff, Mitchell W. ;
Kodali, Susheel ;
Mack, Michael J. ;
Mehran, Roxana ;
Rodes-Cabau, Josep ;
Vranckx, Pascal ;
Webb, John G. ;
Windecker, Stephan ;
Serruys, Patrick W. ;
Leon, Martin B. .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2012, 42 (05) :S45-S60
[10]   The SAVI-TF Registry 1-Year Outcomes of the European Post-Market Registry Using the ACURATE neo Transcatheter Heart Valve Under Real-World Conditions in 1,000 Patients [J].
Kim, Won-Keun ;
Hengstenberg, Christian ;
Hilker, Michael ;
Kerber, Sebastian ;
Schaefer, Ulrich ;
Rudolph, Tanja ;
Linke, Axel ;
Franz, Norbert ;
Kuntze, Thomas ;
Nef, Holger ;
Kappert, Utz ;
Zembala, Michael O. ;
Toggweiler, Stefan ;
Walther, Thomas ;
Moellmann, Helge .