videolaryngoscopy; airway; intubation; laryngoscopy; CERVICAL-SPINE IMMOBILIZATION; MACINTOSH LARYNGOSCOPE BLADE; DIFFICULT AIRWAY MANAGEMENT; INTUBATING LARYNGEAL MASK; EARLY CLINICAL-EXPERIENCE; TRACHEAL INTUBATION; VIDEO LARYNGOSCOPE; ENDOTRACHEAL INTUBATION; MCGRATH(R) VIDEOLARYNGOSCOPE; C-MAC(R) VIDEOLARYNGOSCOPE;
D O I
10.1177/0310057X1204000407
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Due to the large number of videolaryngoscopes now available, it might be difficult for novice users to assess the various devices or use them optimally. We have collated the experiences of several airway management experts to assist in the assessment and optimal use of seven commonly used videolaryngoscopes. While all videolaryngoscopes have unique features, they can be broadly divided into those inserted via a midline approach over the tongue and those inserted laterally along the floor of the mouth. Videolaryngoscopes that are placed on the floor of the mouth displace the tongue antero-laterally and flatten the submandibular tissues. They generally require a conventional shaped bougie for tracheal intubation. Videolaryngoscopes that use the midline approach may have an in-built airway conduit for the tracheal tube or may require a 'J-shaped' stylet in the tracheal tube to negotiate the upper airway. This may cause difficulty when the tracheal tube is inserted through the glottis and the tip abuts the anterior wall of the subglottic space. Knowledge of the mechanism used by videolaryngoscopes to achieve laryngoscopy is essential for safe and successful tracheal intubation when using these devices.
引用
收藏
页码:622 / 630
页数:9
相关论文
共 42 条
[1]   An Algorithm for Difficult Airway Management, Modified for Modern Optical Devices (Airtraq Laryngoscope; LMA CTrach™) A 2-Year Prospective Validation in Patients for Elective Abdominal, Gynecologic, and Thyroid Surgery [J].
Amathieu, Roland ;
Combes, Xavier ;
Abdi, Widad ;
El Housseini, Loutfi ;
Rezzoug, Ahmed ;
Dinca, Andrei ;
Slavov, Velislav ;
Bloc, Sebastien ;
Dhonneur, Gilles .
ANESTHESIOLOGY, 2011, 114 (01) :25-33
[2]   Videolaryngoscopes Do They Truly Have Roles in Difficult Airways? [J].
Asai, Takashi .
ANESTHESIOLOGY, 2012, 116 (03) :515-517
[3]   Use of the Pentax-AWS® in 293 Patients with Difficult Airways [J].
Asai, Takashi ;
Liu, Eugene H. ;
Matsumoto, Sanae ;
Hirabayashi, Yoshihiro ;
Seo, Norimasa ;
Suzuki, Akihiro ;
Toi, Takashi ;
Yasumoto, Kazumasa ;
Okuda, Yasuhisa .
ANESTHESIOLOGY, 2009, 110 (04) :898-904
[4]   Comparative Effectiveness of the C-MAC Video Laryngoscope versus Direct Laryngoscopy in the Setting of the Predicted Difficult Airway [J].
Aziz, Michael F. ;
Dillman, Dawn ;
Fu, Rongwei ;
Brambrink, Ansgar M. .
ANESTHESIOLOGY, 2012, 116 (03) :629-636
[5]   Routine Clinical Practice Effectiveness of the Glidescope in Difficult Airway Management An Analysis of 2,004 Glidescope Intubations, Complications, and Failures from Two Institutions [J].
Aziz, Michael F. ;
Healy, David ;
Kheterpal, Sachin ;
Fu, Rongwei F. ;
Dillman, Dawn ;
Brambrink, Ansgar M. .
ANESTHESIOLOGY, 2011, 114 (01) :34-41
[6]   Improved Glottic Exposure With the Video Macintosh Laryngoscope in Adult Emergency Department Tracheal Intubations [J].
Brown, Calvin A., III ;
Bair, Aaron E. ;
Pallin, Daniel J. ;
Laurin, Erik G. ;
Walls, Ron M. .
ANNALS OF EMERGENCY MEDICINE, 2010, 56 (02) :83-88
[7]   Endotracheal tube as a guide for an Eschmann gum elastic bougie to aid tracheal intubation using the McGrath® or GlideScope® videolaryngoscopes [J].
Budde, Arne O. ;
Pott, Leonard M. .
JOURNAL OF CLINICAL ANESTHESIA, 2008, 20 (07) :560-560
[8]   The learning curve for videolaryngoscopy [J].
Butchart, A. G. ;
Young, P. .
ANAESTHESIA, 2010, 65 (11) :1145-1146
[9]  
Caplan RA, 2003, ANESTHESIOLOGY, V98, P1269
[10]   First Clinical Evaluation of the C-MAC D-Blade Videolaryngoscope During Routine and Difficult Intubation [J].
Cavus, Erol ;
Neumann, Tobias ;
Doerges, Volker ;
Moeller, Thora ;
Scharf, Edwin ;
Wagner, Klaus ;
Bein, Berthold ;
Serocki, Goetz .
ANESTHESIA AND ANALGESIA, 2011, 112 (02) :382-385