Evaluation of cardiac assays on a benchtop chemiluminescent enzyme immunoassay analyzer, PATHFAST

被引:51
作者
Kurihara, Takashi [1 ]
Yanagida, Atsushi [1 ]
Yokoi, Hiroyuki [1 ]
Koyata, Atsushi [1 ]
Matsuya, Takeshi [1 ]
Ogawa, Junichi [1 ]
Okamura, Yoshikazu [1 ]
Miyamoto, Daishi [1 ]
机构
[1] Mitsubishi Kagaku Iatron, Dept Res & Dev, Chiba 2760046, Japan
关键词
D O I
10.1016/j.ab.2007.12.030
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Devices for cardiac immunoassay are widely available at point of care settings. The lateral flow method is the most popular solution for ease of use. But its major shortcomings, poor assay precision, and low analytical sensitivity may lead to false negative results. Therefore, confirmatory testing by a routine lab analyzer sometimes is necessary. In the current study, we evaluated the cardiac assays troponin I, MB isoenzyme of creatine kinase, myoglobin, and N-terminal pro brain natriuretic peptide on a chemiluminescent enzyme immunoassay analyzer, PATHFAST. All of the assays demonstrated correlation coefficients higher than 0.97 against the predicate devices along with good total precision (coefficients of variation <10%), and the troponin I assay showed analytical sensitivity in conformance with the guidelines of European Society of Cardiology/American College of Cardiology and National Academy of Clinical Biochemistry. We concluded that the system was rapid and easy to use without compromising analytical performance. (C) 2007 Elsevier Inc. All rights reserved.
引用
收藏
页码:144 / 146
页数:3
相关论文
共 11 条
  • [1] Myocardial infarction redefined -: A consensus Document of the Joint European Society of Cardiology/American College of Cardiology Committee for the Redefinition of Myocardial Infarction
    Alpert, JS
    Antman, E
    Apple, F
    Armstrong, PW
    Bassand, JP
    de Luna, AB
    Beller, G
    Breithardt, G
    Chaitman, BR
    Clemmensen, P
    Falk, E
    Fishbein, MC
    Galvani, M
    Garson, A
    Grines, C
    Hamm, C
    Jaffe, A
    Katus, H
    Kjekshus, J
    Klein, W
    Klootwijk, P
    Lenfant, C
    Levy, D
    Levy, RI
    Luepker, R
    Marcus, F
    Näslund, U
    Ohman, M
    Pahlm, O
    Poole-Wilson, P
    Popp, R
    Alto, P
    Pyörälä, K
    Ravkilde, J
    Rehnquist, N
    Roberts, W
    Roberts, R
    Roelandt, J
    Rydén, L
    Sans, S
    Simoons, ML
    Thygesen, K
    Tunstall-Pedoe, H
    Underwood, R
    Uretsky, BF
    Van de Werf, F
    Voipio-Pulkki, LM
    Wagner, G
    Wallentin, L
    Wijns, W
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 36 (03) : 959 - 969
  • [2] Serum and plasma cardiac troponin I 99th percentile reference values for 3 2nd-generation assays
    Apple, Fred S.
    Murakarni, MaryAnn M.
    [J]. CLINICAL CHEMISTRY, 2007, 53 (08) : 1558 - 1560
  • [3] Clinical and Laboratory Standards Institute (CLSI), 2004, EP5A2 CLSI
  • [4] Point-of-care testing and cardiac biomarkers: The standard of care and vision for chest pain centers
    Kost, GJ
    Tran, NK
    [J]. CARDIOLOGY CLINICS, 2005, 23 (04) : 467 - +
  • [5] National Academy of Clinical Biochemistry Laboratory Medicine Practice Guidelines: Clinical characteristics and utilization of biochemical markers in acute coronary syndromes
    Morrow, David A.
    Cannon, Christopher P.
    Jesse, Robert L.
    Newby, L. Kristin
    Ravkilde, Jan
    Storrow, Alan B.
    Wu, Alan H. B.
    Christenson, Robert H.
    Apple, Fred S.
    Francis, Gary
    Tang, Wilson
    [J]. CLINICAL CHEMISTRY, 2007, 53 (04) : 552 - 574
  • [6] NCCLS, 2004, EP17A NCCLS
  • [7] Development of a novel method for operating magnetic particles, magtration technology, and its use for automating nucleic acid purification
    Obata, K
    Segawa, O
    Yakabe, M
    Ishida, Y
    Kuroita, T
    Ikeda, K
    Kawakami, B
    Kawamura, Y
    Yohda, M
    Matsunaga, T
    Tajima, H
    [J]. JOURNAL OF BIOSCIENCE AND BIOENGINEERING, 2001, 91 (05) : 500 - 503
  • [8] Evaluation of imprecision for cardiac troponin assays at low-range concentrations
    Panteghini, M
    Pagani, F
    Yeo, KTJ
    Apple, FS
    Christenson, RH
    Dati, F
    Mair, J
    Ravkilde, J
    Wu, AHB
    [J]. CLINICAL CHEMISTRY, 2004, 50 (02) : 327 - 332