Good agreement between bioelectrical impedance and dual-energy X-ray absorptiometry for estimating changes in body composition during weight loss in overweight young women

被引:160
作者
Thomson, Rebecca [1 ,2 ]
Brinkworth, Grant D. [1 ]
Buckley, Jonathan D. [2 ]
Noakes, Manny [1 ]
Clifton, Peter M. [1 ]
机构
[1] CSIRO Human Nutr, Adelaide, SA 5000, Australia
[2] Univ S Australia, Sch Hlth Sci, ATN Ctr Metabol Fitness & Nutrit Physiol Res Ctr, Adelaide, SA 5001, Australia
基金
英国医学研究理事会;
关键词
body composition; body fat; weight toss; bioelectrical impedance analysis; dual X-ray absorptiometry;
D O I
10.1016/j.clnu.2007.08.003
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Aims: To compare estimations of body composition using two different methods of bioetectrical impedance analysis (BIA) with dual X-ray absorptiometry (DXA) for estimating body composition during weight loss in overweight and obese young females. Methods: Twenty-four overweight or obese females (age 29.5+/-6.1 years, BMI 36.4+/-4.3 kg/m(2)) had body composition assessed using single-frequency (Tanita Ultimate Scale (TM); SF-BIA) and multi-frequency (Impedimed SFB7(TM); MF-BIA) BIA and DXA before and after a 10-week weight toss intervention. Results: MF-BIA estimates of body composition showed good absolute agreement with DXA, as evidenced by the small biases in the estimation of fat free mass (FFM), fat mass (FM) and percentage body fat (BF%); however, the limits of agreement for each variable were wide (bias +/-1.96 standard deviation; FFM -1.6+/-6.5kg, FM 1.6+/-6.5kg, BF% 1.4 +/- 6.3%). SF-BIA exhibited a larger bias with wide limits of agreement (FFM 3.8 +/- 9.1 kg, FM -3.8+/-9.1 kg, BF% -4.37+/-10.3%). During weight toss the values provided by MF-BIA and SF-BIA were not significantly different from DXA (p >= 0.89) due to small bias and the limits of agreement were narrow (MF-BIA: FFM -0.+/-3.74 kg, FM 0.01+/-3.74 kg, BF% 0.22+/-3.40%; SF-BIA: FFM 0.40+/-3.92 kg, FM -0.40+/-3.92 kg, BF% 0.25+/-3.40%). Conclusion: Compared with DXA, both the MF-BIA and SF-BIA accurately assessed changes in body composition with weight loss but, compared with SF-BIA, MF-BIA provided superior cross-sectional estimates of body composition. (C) 2007 Elsevier Ltd and European Society for Clinical Nutrition and Metabolism. ALL rights reserved.
引用
收藏
页码:771 / 777
页数:7
相关论文
共 24 条
[1]  
Andreoli A, 2002, J SPORT MED PHYS FIT, V42, P186
[2]  
Carella MJ, 1997, OBES RES, V5, P250
[3]  
Cornish BH, 1996, EUR J CLIN NUTR, V50, P159
[4]  
Deurenberg Paul, 1996, American Journal of Clinical Nutrition, V64, p449S, DOI 10.1093/ajcn/64.3.449S
[5]   Reliability and variability of bioimpedance measures in normal adults: Effects of age, gender, and body mass [J].
Dittmar, M .
AMERICAN JOURNAL OF PHYSICAL ANTHROPOLOGY, 2003, 122 (04) :361-370
[6]   Comparison of methods to assess body composition changes during a period of weight loss [J].
Frisard, MI ;
Greenway, FL ;
DeLany, JP .
OBESITY RESEARCH, 2005, 13 (05) :845-854
[7]  
Hendel HW, 1996, INT J OBESITY, V20, P1111
[8]   Change in fat-free mass assessed by bioelectrical impedance, total body potassium and dual energy X-ray absorptiometry during prolonged weight loss [J].
Hendel, HW ;
Gotfredsen, A ;
Hojgaard, L ;
Andersen, T ;
Hilsted, J .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1996, 56 (08) :671-679
[9]   Validity of the leg-to-leg bioimpedance to estimate changes in body fat during weight loss and regain in overweight women:: a comparison with multi-compartment models [J].
Jebb, S. A. ;
Siervo, M. ;
Murgatroyd, P. R. ;
Evans, S. ;
Fruhbeck, G. ;
Prentice, A. M. .
INTERNATIONAL JOURNAL OF OBESITY, 2007, 31 (05) :756-762
[10]   ASSESSMENT OF BODY-COMPOSITION WITH USE OF DUAL-ENERGY X-RAY ABSORPTIOMETRY - EVALUATION AND COMPARISON WITH OTHER METHODS [J].
JENSEN, MD ;
KANALEY, JA ;
ROUST, LR ;
OBRIEN, PC ;
BRAUN, JS ;
DUNN, WL ;
WAHNER, HW .
MAYO CLINIC PROCEEDINGS, 1993, 68 (09) :867-873