Performance evaluation of the new fully automated urine particle analyser OF-5000 compared to the reference method of the Fuchs-Rosenthal chamber

被引:61
作者
Previtali, Giulia [1 ]
Ravasio, Rudi [1 ]
Seghezzi, Michela [1 ]
Buoro, Sabrina [1 ]
Alessio, Maria Grazia [1 ]
机构
[1] Papa Giovanni XXIII Hosp, Dept Lab Med, Clin Chem Lab, Piazza OMS 1, Bergamo, Italy
关键词
UF-5000; Automated urine analysis; Urine particles; Fuchs Rosenthal chamber; MICROSCOPY; SEDIMENT;
D O I
10.1016/j.cca.2017.07.028
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background" OF-5000 is the new fully automated urine particle analyser. We validated its performance. Methods: 736 urines were analysed and results were compared by two pathologists on uncentrifuged samples, using Fuchs-Rosenthal chamber. Results: AUC of ROC curve ranged between 0.86 and 0.99. Sensitivity was > 0.90 for all the elements and similar for RBC and yeasts. Specificity ranged between 0.74 and 0.89 for total cast, epithelial/non-squamous/renal-tubular cells and RBC. For all the other parameters specificity was > 0.90. Comparison with Fuchs Rosenthal chamber was very good for all the parameters; r ranged between 0.52 and 0.99 except for pathological cast because of the lack of the pathological samples in medium and higher ranges. Linearity performance (R-2) was 1.00, 1.00 and 0.99 respectively for RBC, WBC and epithelial cells (EC). No carry-over was observed. The within-run imprecision was 25.42%,13.81%,1.36% for RBC; 37.50%,10.16%,1.41% for WBC and 35.25%, 17.85%,6.30% for EC at low, near the cut off level and high concentrations, respectively. The between-run imprecision was 6.90%,1.60% for RBC, 4.10%,1.90% for WBC and 7.60%,7.30% for EC, using low and high positive quality controls, respectively. Conclusion: OF-5000 is an analyser of great interest to detect urine particle related to pathological process of kidney and urinary tract.
引用
收藏
页码:123 / 130
页数:8
相关论文
共 25 条
[1]  
[Anonymous], EV PREC PERF CLIN CH
[2]  
[Anonymous], 2000, SCAND J CLIN LAB I S, V231, P1, DOI [10.1080/00365513.2000.12056993, DOI 10.1080/00365513.2000.12056993]
[3]   Comparison of Cobas 6500 and Iris IQ200 fully-automated urine analyzers to manual urine microscopy [J].
Bakan, Ebubekir ;
Ozturk, Nurinnisa ;
Baygutalp, Nurcan Kilic ;
Polat, Elif ;
Akpinar, Kadriye ;
Dorman, Emrullah ;
Polat, Harun ;
Bakan, Nuri .
BIOCHEMIA MEDICA, 2016, 26 (03) :365-375
[4]   Advances in Urine Microscopy [J].
Becker, Gavin J. ;
Garigali, Giuseppe ;
Fogazzi, Giovanni B. .
AMERICAN JOURNAL OF KIDNEY DISEASES, 2016, 67 (06) :954-964
[5]   REVISED SCHEME FOR EVALUATION OF AUTOMATIC INSTRUMENTS FOR USE IN CLINICAL CHEMISTRY [J].
BROUGHTON, PM ;
GOWENLOCK, AH ;
MCCORMACK, JJ ;
NEILL, DW .
ANNALS OF CLINICAL BIOCHEMISTRY, 1974, 11 (NOV) :207-218
[6]   Evaluation of the Sysmex UF1000i flow cytometer for ruling out bacterial urinary tract infection [J].
De Rosa, Rita ;
Grosso, Shamanta ;
Bruschetta, Graziano ;
Avolio, Manuela ;
Stano, Paola ;
Modolo, Maria Luisa ;
Camporese, Alessandro .
CLINICA CHIMICA ACTA, 2010, 411 (15-16) :1137-1142
[7]   The role of automated urine particle flow cytometry in clinical practice [J].
Delanghe, JR ;
Kouri, TT ;
Huber, AR ;
Hannemann-Pohl, K ;
Guder, WG ;
Lun, A ;
Sinha, P ;
Stamminger, G ;
Beier, L .
CLINICA CHIMICA ACTA, 2000, 301 (1-2) :1-18
[8]   Clinical value of crystalluria and quantitative morphoconstitutional analysis of urinary calculi [J].
Frochot, Vincent ;
Daudon, Michel .
INTERNATIONAL JOURNAL OF SURGERY, 2016, 36 :624-632
[10]  
Ince Fatma Demet, 2016, Pract Lab Med, V5, P14, DOI 10.1016/j.plabm.2016.03.002