Image Definition and Defect Sizing in Lock-in Thermography: An Experimental Investigation

被引:1
作者
Sharath, D. [1 ,2 ]
Menaka, M. [3 ]
Venkatraman, B. [3 ,4 ]
机构
[1] Visvesvaraya Technol Univ, Ctr Imaging Technol, MS Ramaiah Inst Technol, Bengaluru, India
[2] Visvesvaraya Technol Univ, MS Ramaiah Inst Technol, Dept Chem, Bengaluru, India
[3] Indira Gandhi Ctr Atom Res, Safety Qual & Resources Management Grp, Kalpakkam, Tamil Nadu, India
[4] Homi Bhabha Natl Inst, Kalpakkam, Tamil Nadu, India
关键词
Lock-in Thermography; Image Definition; Defect Sizing; Edge Detection; SUBSURFACE DEFECTS; RESOLUTION;
D O I
10.1007/s40799-021-00506-9
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
In any imaging modality the apparent size and shape of the object in the image plane is influenced by the camera parameters and experimental conditions. To extract an accurate measurement of object size from the image, the understanding of the influence of these parameters on the apparent size and shape is important. Lock-in Thermography is one of the advanced imaging Non Destructive Evaluation techniques which is widely used for quantitative defect evaluation and material characterization. The important experimental parameter in Lock-in Thermography is the excitation frequency, which greatly affects both the apparent shape (image definition) and size in resultant phase and amplitude images. Hence, in this study, an attempt is made towards understanding the effect of frequency on the image definition (apparent shape) and defect sizing in Lock-in Thermography through experimental study. Stainless Steel plate with flat bottom holes of various sizes and depths is considered for the experiment. Both phase and amplitude images are considered for analysis. The Canny edge detection technique is implemented for studying the effect of frequency on image definition by extracting the edges of the flat bottom hole. The binary image obtained after edge detection is used for sizing calculation. The apparent shape and size were greatly affected by excitation frequency and depth in both amplitude and phase images. A relation between excitation frequency and depth was established which can be used to predict the excitation frequency where the sizing measurement is accurate.
引用
收藏
页码:811 / 822
页数:12
相关论文
共 50 条
  • [31] Characterization of semi-infinite delaminations using lock-in thermography: Experimental results
    Sagarduy-Marcos, David
    Perez-Arbulu, Jon
    Rodriguez-Aseguinolaza, Javier
    Mendioroz, Arantza
    Batsale, Jean-Christophe
    Salazar, Agustin
    NDT & E INTERNATIONAL, 2023, 138
  • [32] Lock-in thermography and related topics in photovoltaic research
    Breitenstein, Otwin
    DEFECTS-RECOGNITION, IMAGING AND PHYSICS IN SEMICONDUCTORS XIV, 2012, 725 : 115 - 122
  • [33] The Impact of Excitation Periods on the Outcome of Lock-In Thermography
    Sapieta, Milan
    Dekys, Vladimir
    Kopas, Peter
    Jakubovicova, Lenka
    Savrnoch, Zdenko
    MATERIALS, 2023, 16 (07)
  • [34] Development of LabVIEW Program for Lock-In Infrared Thermography
    Min, Taehoon
    Na, Hyungchul
    Kim, Nohyu
    JOURNAL OF THE KOREAN SOCIETY FOR NONDESTRUCTIVE TESTING, 2011, 31 (02) : 127 - 133
  • [35] Phase Lock-In Thermography for Metal Walls Characterization
    Quoc, S. Pham Tu
    Cheymol, G.
    Semerok, A.
    2013 3RD INTERNATIONAL CONFERENCE ON ADVANCEMENTS IN NUCLEAR INSTRUMENTATION, MEASUREMENT METHODS AND THEIR APPLICATIONS (ANIMMA), 2013,
  • [36] Applications of pulse heating thermography and lock-in thermography to quantitative nondestructive evaluations
    Sakagami, T
    Kubo, S
    INFRARED PHYSICS & TECHNOLOGY, 2002, 43 (3-5) : 211 - 218
  • [37] Research on thermal wave processing of lock-in thermography based on analyzing image sequences for NDT
    Liu Junyan
    Wang Yang
    Dai Jingmin
    INFRARED PHYSICS & TECHNOLOGY, 2010, 53 (05) : 348 - 357
  • [38] Investigation of Geometry, Frequency and Material's Effects in Lock-In Thermography Applications in Semiconductor Packages
    Kijkanjanapaiboon, Kasemsak
    Xie, Mayue
    Fan, Xuejun
    2016 IEEE 66TH ELECTRONIC COMPONENTS AND TECHNOLOGY CONFERENCE (ECTC), 2016, : 1423 - 1429
  • [39] Research on structure stress analysis on composite by lock-in thermography
    Liu Xun
    Liu Junyan
    Dai Jingmin
    ADVANCES IN COMPOSITES, PTS 1 AND 2, 2011, 150-151 : 1649 - +
  • [40] Inspection of Impact Damage on CFRP by Lock-in Induction Thermography
    Liu, Ting
    Zhang, Yu
    Netzelmann, Udo
    Guo, Jianguang
    Zhang, Wang
    ELEVENTH INTERNATIONAL CONFERENCE ON INFORMATION OPTICS AND PHOTONICS (CIOP 2019), 2019, 11209