A comparison of radio-frequency biomotion sensors and actigraphy versus polysomnography for the assessment of sleep in normal subjects

被引:62
作者
O'Hare, Emer [1 ]
Flanagan, David [1 ,3 ]
Penzel, Thomas [2 ,4 ]
Garcia, Carmen [4 ]
Frohberg, Daniela [2 ]
Heneghan, Conor [1 ,3 ]
机构
[1] ResMed Sensor Technol, Dublin 4, Ireland
[2] Adv Sleep Res, D-10117 Berlin, Germany
[3] Natl Univ Ireland Univ Coll Dublin, Sch Elect Elect & Commun Engn, Dublin 4, Ireland
[4] Charite, Sleep Med Ctr, D-10117 Berlin, Germany
关键词
Biomotion sensor; Actigraphy; Total sleep time; Wake-after-sleep onset; WRIST ACTIGRAPHY; APNEA PATIENTS; PARAMETERS; WAKE; VALIDATION; INSOMNIA;
D O I
10.1007/s11325-014-0967-z
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
This paper aims to compare the absolute performance of three noncontact sleep measurement devices for measuring sleep parameters in normal subjects against polysomnography and to assess their relative performance. The devices investigated were two noncontact radio-frequency biomotion sensors (SleepMinder (SM) and SleepDesign (HSL-101)) and an actigraphy-based system (Actiwatch). Overnight polysomnography measurements were carried out in 20 normal subjects, with simultaneous assessment of sleep parameters using the three devices. The parameters measured included total sleep time (TST), sleep efficiency (SE), sleep-onset latency (SOL), and wake-after-sleep onset (WASO). The per-epoch agreement level for sleep/wake distinction was evaluated. The TSTs reported by the three devices were 426 +/- 34, 434 +/- 22, and 441 +/- 16 min, for the SM, HSL-101, and Actiwatch, respectively, against polysomnogram (PSG)-reported TST of 391 +/- 49 min. The SOLs were 10 +/- 10, 5 +/- 6, and 3 +/- 2 min for the SM, HSL-101 and Actiwatch, respectively against PSG SOL of 19 +/- 13 min. The WASO times were 46 +/- 33, 43 +/- 22, and 38 +/- 17 min, as against PSG-reported 69 +/- 46 min. All three devices had a statistically significant bias to overestimate sleep time and underestimate WASO and SOL compared with PSG. The performance of the three devices was basically equivalent, with only minor interdevice differences. The overall per-epoch agreement levels were 86 % for the SM, 86 % for the HSL-101, and 85 % for the Actiwatch. Noncontact biomotion approaches to sleep measurement provided reasonable estimates of TST, but with a bias to over-estimation of sleep. The radio-frequency biomotion sensors provided similar accuracies for sleep/wake determination in normal subjects as the actigraph used in this study and slightly improved estimates of TST, SOL, and WASO.
引用
收藏
页码:91 / 98
页数:8
相关论文
共 17 条
  • [1] Berry RB, 2020, AASM MANUAL SCORING
  • [2] Comparison of sleep parameters from actigraphy and polysomnography in older women: The SOF study
    Blackwell, Terri
    Redline, Susan
    Ancoli-Israel, Sonia
    Schneider, Jennifer L.
    Surovec, Susan
    Johnson, Nathan L.
    Cauley, Jane A.
    Stone, Katie L.
    [J]. SLEEP, 2008, 31 (02) : 283 - 291
  • [3] Factors that May Influence the Classification of Sleep-Wake by Wrist Actigraphy: The MrOS Sleep Study
    Blackwell, Terri
    Ancoli-Israel, Sonia
    Redline, Susan
    Stone, Katie L.
    [J]. JOURNAL OF CLINICAL SLEEP MEDICINE, 2011, 7 (04): : 357 - 367
  • [4] Blood ML, 1997, SLEEP, V20, P388
  • [5] Sleep/wake measurement using a non-contact biomotion sensor
    de Chazal, Philip
    Fox, Niall
    O'Hare, Emer
    Heneghan, Conor
    Zaffaroni, Alberto
    Boyle, Patricia
    Smith, Stephanie
    O'Connell, Caroline
    McNicholas, Walter T.
    [J]. JOURNAL OF SLEEP RESEARCH, 2011, 20 (02) : 356 - 366
  • [6] Further validation of actigraphy for sleep studies
    de Souza, L
    Benedito-Silva, AA
    Pires, MLN
    Poyares, D
    Tufik, S
    Calil, HM
    [J]. SLEEP, 2003, 26 (01): : 81 - 85
  • [7] Wrist-actigraphic estimation of sleep time under nCPAP treatment in sleep apnoea patients
    Gagnadoux, F
    Nguyen, XL
    Rakotonanahary, D
    Vidal, S
    Fleury, B
    [J]. EUROPEAN RESPIRATORY JOURNAL, 2004, 23 (06) : 891 - 895
  • [8] A novel adaptive wrist actigraphy algorithm for sleep-wake assessment in sleep apnea patients
    Hedner, J
    Pillar, G
    Pittman, SD
    Zou, D
    Grote, L
    White, DP
    [J]. SLEEP, 2004, 27 (08): : 1560 - 1566
  • [9] Why Bland-Altman plots should use X, not (Y+X)/2 when X is a reference method
    Krouwer, Jan S.
    [J]. STATISTICS IN MEDICINE, 2008, 27 (05) : 778 - 780
  • [10] Comparison of actigraphic, polysomnographic, and subjective assessment of sleep parameters in sleep-disordered patients
    Kushida, Clete A.
    Chang, Arthur
    Gadkary, Chirag
    Guilleminault, Christian
    Carrillo, Oscar
    Dement, William C.
    [J]. SLEEP MEDICINE, 2001, 2 (05) : 389 - 396