Impact of the MLC leaf-tip model in a commercial TPS: Dose calculation limitations and IROC-H phantom failures

被引:28
作者
Koger, Brandon [1 ]
Price, Ryan [1 ]
Wang, Da [1 ]
Toomeh, Dolla [1 ]
Geneser, Sarah [1 ]
Ford, Eric [1 ]
机构
[1] Univ Washington, Sch Med, Dept Radiat Oncol, Seattle, WA 98195 USA
关键词
MLC leaf tip; multi-leaf collimator; treatment planning; treatment planning system validation; POSITION ACCURACY; IMRT QA; DOSIMETRY; VERIFICATION; PERFORMANCE; VALIDATION; BEAM;
D O I
10.1002/acm2.12819
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Treatment planning system (TPS) dose calculation is sensitive to multileaf collimator (MLC) modeling, especially when treating with intensity-modulated radiation therapy (IMRT) or VMAT. This study investigates the dosimetric impact of the MLC leaf-tip model in a commercial TPS (RayStation v.6.1). The detectability of modeling errors was assessed through both measurements with an anthropomorphic head-and-neck phantom and patient-specific IMRT QA using a 3D diode array. Methods and Materials An Agility MLC (Elekta Inc.) was commissioned in RayStation. Nine IMRT and VMAT plans were optimized to treat the head-and-neck phantom from the Imaging and Radiation Oncology Core Houston branch (IROC-H). Dose distributions for each plan were re-calculated on 27 beam models, varying leaf-tip width (2.0, 4.5, and 6.5 mm) and leaf-tip offset (-2.0 to +2.0 mm) values. Doses were compared to phantom TLD measurements. Patient-specific IMRT QA was performed, and receiver-operating characteristic (ROC) analysis was performed to determine the detectability of modeling errors. Results Dose calculations were very sensitive to leaf-tip offset values. Offsets of +/- 1.0 mm resulted in dose differences up to 10% and 15% in the PTV and spinal cord TLDs respectively. Offsets of +/- 2.0 mm caused dose deviations up to 50% in the spinal cord TLD. Patient-specific IMRT QA could not reliably detect these deviations, with an ROC area under the curve (AUC) value of 0.537 for a +/- 1.0 mm change in leaf-tip offset, corresponding to >7% dose deviation. Leaf-tip width had a modest dosimetric impact with <2% and 5.6% differences in the PTV and spinal cord TLDs respectively. Conclusions Small changes in the MLC leaf-tip offset in this TPS model can cause large changes in the calculated dose for IMRT and VMAT plans that are difficult to identify through either dose curves or standard patient-specific IMRT QA. These results may, in part, explain the reported high failure rate of IROC-H phantom tests.
引用
收藏
页码:82 / 88
页数:7
相关论文
共 28 条
[1]  
[Anonymous], 2012, AM SOC RAD ONCOL
[2]   Quantifying the performance of in vivo portal dosimetry in detecting four types of treatment parameter variations [J].
Bojechko, C. ;
Ford, E. C. .
MEDICAL PHYSICS, 2015, 42 (12) :6912-6918
[3]   Requirements for leaf position accuracy for dynamic multileaf collimation [J].
Budgell, GJ ;
Mott, JHL ;
Williams, PC ;
Brown, KJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (05) :1211-1227
[4]   Dosimetric considerations for validation of a sequential IMRT process with a commercial treatment planning system [J].
Cadman, P ;
Bassalow, R ;
Sidhu, NPS ;
Ibbott, G ;
Nelson, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (16) :3001-3010
[5]   Validation of physics improvements for IMRT with a commercial treatment-planning system [J].
Cadman, Patrick ;
McNutt, Todd ;
Bzdusek, Karl .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2005, 6 (02) :74-86
[6]   ROC analysis in patient specific quality assurance [J].
Carlone, Marco ;
Cruje, Charmainne ;
Rangel, Alejandra ;
McCabe, Ryan ;
Nielsen, Michelle ;
MacPherson, Miller .
MEDICAL PHYSICS, 2013, 40 (04)
[7]   Optimizing the MLC model parameters for IMRT in the RayStation treatment planning system [J].
Chen, Shifeng ;
Yi, Byong Yong ;
Yang, Xiaocheng ;
Xu, Huijun ;
Prado, Karl L. ;
D'Souza, Warren D. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (05) :322-332
[8]  
Donaldson S., 2007, Towards Safer Radiotherapy
[9]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373
[10]   Treatment plan complexity does not predict IROC Houston anthropomorphic head and neck phantom performance [J].
Glenn, Mallory C. ;
Hernandez, Victor ;
Saez, Jordi ;
Followill, David S. ;
Howell, Rebecca M. ;
Pollard-Larkin, Julianne M. ;
Zhou, Shouhao ;
Kry, Stephen F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (20)