Automation of μ-SPE (Smart-SPE) and Liquid-Liquid Extraction Applied for the Analysis of Chemical Warfare Agents

被引:5
作者
Althoff, Marc Andre [1 ]
Bertsch, Andreas [2 ]
Metzulat, Manfred [2 ]
机构
[1] LMU, Dept Chem, Butenandtstr 5-13,Haus D, D-81377 Munich, Germany
[2] Chem Def Safety & Environm Protect Sch, Sci Dept, Chem Sect, Muhlenweg 12, D-87527 Sonthofen, Germany
关键词
automation; sample handling; sample preparation; chemical warfare agent; SPE; LLE; WATER;
D O I
10.3390/separations6040049
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Existing autosamplers are frequently applied only for subjecting the samples to the instruments for injection. In our study, we have set up a TriPlusRSH autosampler mounted on a GC-FID-MS/MS system using the new Method Composer and Script Editor software to automatize all necessary sample preparation steps and subsequent injection of samples in the field of chemical disarmament. Those include but are not limited to: liquid-liquid extraction, drying steps, solvent exchange, and mu-SPE. Tedious and error prone off-line steps are eliminated. In particular, when investigating highly toxic substances like chemical warfare agents or anticancer drugs, automation can help to minimize health risks for lab personnel. The setup engaged features brand new prototype equipment, e.g., a centrifuge to assist in phase separation for liquid-liquid extraction. Efficiency and accuracy of the automated methods were carefully evaluated and proven to outperform the respective manual steps after optimization, e.g., the processing time is up to 60% faster and recovery rates are doubled. The developed workflows can easily be adapted to other sample preparation protocols, e.g., determination of octanol/water partition coefficients, and be used amongst different instruments and chromatography data handling systems.
引用
收藏
页数:15
相关论文
共 17 条
[1]   Application of headspace and direct immersion solid-phase microextraction in the analysis of organothiophosphates related to the Chemical Weapons Convention from water and complex matrices [J].
Althoff, Marc Andre ;
Bertsch, Andreas ;
Metzulat, Manfred ;
Klapoetke, Thomas M. ;
Karaghiosoff, Konstantin L. .
TALANTA, 2017, 174 :295-300
[2]   New aspects of the detection and analysis of organo(thio)phosphates related to the chemical weapons convention [J].
Althoff, Marc Andre ;
Bertsch, Andreas ;
Metzulat, Manfred ;
Kalthoff, Oliver ;
Karaghiosoff, Konstantin .
PHOSPHORUS SULFUR AND SILICON AND THE RELATED ELEMENTS, 2017, 192 (02) :149-156
[3]  
[Anonymous], 2018, TRINKWASSERANLAYSE S, V2018, P1
[4]  
[Anonymous], 2011, HDB SOLID PHASE MICR
[5]  
Armbruster David A, 2014, Clin Biochem Rev, V35, P143
[6]   Efficient Flattening Algorithm for Hierarchical and Dynamic Structure Discrete Event Models [J].
Bae, Jang Won ;
Bae, Sang Won ;
Moon, Il-Chul ;
Kim, Tag Gon .
ACM TRANSACTIONS ON MODELING AND COMPUTER SIMULATION, 2016, 26 (04)
[7]  
Cairns S., 2013, REPORT UN MISSION IN
[8]  
Chui M., 2016, Where machines could replace humans - and where they can't (yet)
[9]   Newly Introduced Sample Preparation Techniques: Towards Miniaturization [J].
Costa, Rosaria .
CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 2014, 44 (04) :299-310
[10]  
HAKKINEN VMA, 1991, HRC-J HIGH RES CHROM, V14, P811