Soft tissue attenuation patterns in stress myocardial perfusion SPECT images: A comparison between supine and upright acquisition systems

被引:25
作者
Chawla, David [1 ]
Rahaby, Mouyyad
Amin, Amit P.
Vashistha, Raj [2 ]
Alyousef, Tareq
Martinez, Hector X. [3 ]
Doukky, Rami [1 ]
机构
[1] Rush Univ, Cardiol Sect, Med Ctr, Chicago, IL 60612 USA
[2] John H Stroger Jr Hosp Cook Cty, Dept Med, Chicago, IL 60612 USA
[3] Consultants Cardiovasc Imaging SC, Elk Grove Village, IL USA
关键词
Attenuation and scatter correction; instrumentation: SPECT; myocardial perfusion imaging: SPECT; SPECT; image artifacts; EMISSION COMPUTED-TOMOGRAPHY; DISEASE;
D O I
10.1007/s12350-010-9336-x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Soft tissue attenuation patterns and their interaction with body habitus and gender in Single Photon Emission Computed Tomography (SPECT)-myocardial perfusion imaging (MPI) of upright patient-position acquisition systems are not well described. In a retrospective cross-sectional study, we compared the prevalence and patterns of soft tissue attenuation in two groups of normal SPECT-MPI studies acquired by supine (n = 263) vs upright (n = 212) acquisition systems. Attenuation patterns observed in the study population were: anterior (22.3%), inferior (51.6%) and lateral (18.1%). Anterior attenuation was significantly less in those imaged upright (6.1% vs 35.4%), P < .001; particularly among women (9.5% vs 50.7%), P < .001. Inferior attenuation was more common among women imaged upright (49.5% vs 13.5%), P < .001; but was not affected by imaging position among men. Lateral attenuation was more prevalent in the upright group (24.1% vs 13.3%), P = .002; and had a strong association with female gender (P < .001) and BMI a parts per thousand yen 30 (P < .001). Upright SPECT-MPI acquisition is associated with a unique attenuation pattern which is vastly different than the supine position. Female gender strongly impacted this attenuation pattern, particularly obese women. Our study is the first to describe, in details, the attenuation patterns with upright imaging and is critical for the accurate interpretation of SPECT-MPI acquired with upright systems.
引用
收藏
页码:281 / 290
页数:10
相关论文
共 21 条
  • [11] Hendel RC, 2005, Q J NUCL MED MOL IM, V49, P30
  • [12] HENDEL RC, 2004, NUCL CARDIOLOGY PRAC, P157
  • [13] A Multicenter Assessment of the Use of Single-Photon Emission Computed Tomography Myocardial Perfusion Imaging With Appropriateness Criteria
    Hendel, Robert C.
    Cerqueira, Manuel
    Douglas, Pamela S.
    Caruth, Karen C.
    Allen, Joseph M.
    Jensen, Neil C.
    Pan, Wenqin
    Brindis, Ralph
    Wolk, Michael
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (02) : 156 - 162
  • [14] Henzlova Milena J, 2006, J Nucl Cardiol, V13, pe80
  • [15] *IMV MED INF DIV, 2008, IMV MED MARK SUMM RE
  • [16] Diagnostic accuracy of supine and prone thallium-201 stress myocardial perfusion single-photon emission computed tomography to detect coronary artery disease in inferior wall of left ventricle
    Katayama, Takuji
    Ogata, Nobuhiko
    Tsuruya, Yoshio
    [J]. ANNALS OF NUCLEAR MEDICINE, 2008, 22 (04) : 317 - 321
  • [17] Maddahi J, 2009, EUR HEART J SUPPL, V11, pS48
  • [18] Nishina H, 2006, J NUCL MED, V47, P51
  • [19] Attenuation artifact, attenuation correction, and the future of myocardial perfusion SPELT
    Singh, Bhupinder
    Bateman, Timothy M.
    Case, James A.
    Heller, Gary
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2007, 14 (02) : 153 - 164
  • [20] Changes in the appearance of attenuation artefacts due to change in posture in myocardial perfusion imaging
    Tonge, Christine M.
    Armstrong, Ian S.
    Arumugam, Parthiban
    James, Jacqueline M.
    Al-Bahrani, Ghada I.
    Lawson, Richard S.
    Shields, Robert A.
    Prescott, Mary C.
    [J]. NUCLEAR MEDICINE COMMUNICATIONS, 2008, 29 (05) : 441 - 447