摘要
Study objectives: The monitoring of cardiac output (CO) during exercise rehabilitation in patients with COPD, often including strenuous exercise, is advisable. Invasive methods (thermodilution, Fick method) are accurate, but for clinical routine use noninvasive CO estimation is required. We have shown that impedance cardiography (Physio Flow; Manatee Biomedical; Macheren, France) is reliable in COPD patients at rest and during a recumbent, light-intensity exercise. The aim of our study was to evaluate the validity of this noninvasive device in COPD patients during a maximal incremental exercise test (IET) and also during a strenuous intermittent work exercise test (IWET). Design: Prospective comparative study of the impedance cardiograph vs the direct Fick method applied to oxygen. Patients: Eight patients with moderate-to-severe COPD (59 +/- 6 years old; FEV1, 38 +/- 15% predicted; residual volume, 194 +/- 64% predicted) [mean +/- SD]. Measurements and main results: Forty-nine simultaneous measurements of CO by means of the direct Fick method (COFICK) and CO measured by the impedance cardiograph (COPF) were obtained during the IET, and 108 measurements were made during the IWET. The correlation coefficients between the two measurements were r = 0.85 and r = 0.71 for the IET and the IWET, respectively. COPF was higher than COFICK. The difference between the two methods was 3.2 +/- 2.9 L/min during the IET and 2.5 +/- 2.1 L/min during the IWET. Expressed as a percentage of the mean of the two measurements, this corresponded to 31 +/- 21% and 25 +/- 20%, respectively. Conclusions: The relatively high number of values differing by > 20% precludes the use of impedance cardiography in clinical routine in such a difficult setting (hyperinflated patients and intense exercise).
引用
收藏
页码:1122 / 1131
页数:10
相关论文
共 35 条
[21]   Pulmonary hemodynamics during a strenuous intermittent exercise in healthy subjects [J].
Lonsdorfer-Wolf, E ;
Richard, R ;
Doutreleau, S ;
Billat, VL ;
Oswald-Mammosser, M ;
Lonsdorfer, J .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2003, 35 (11) :1866-1874
[22]   CLARIFYING THE ROLE OF CONTINUOUS FICK CARDIAC-OUTPUT MEASUREMENT DURING EXERCISE [J].
LUMB, PD .
CHEST, 1993, 104 (02) :332-333
[23]   CONTINUOUS FICK CARDIAC-OUTPUT MEASUREMENT DURING EXERCISE BY MONITORING OF MIXED VENOUS OXYGEN-SATURATION AND OXYGEN-UPTAKE [J].
NAKANISHI, N ;
YOSHIOKA, T ;
OKANO, Y ;
NISHIMURA, T .
CHEST, 1993, 104 (02) :419-426
[24]   Influence of age and gender on cardiac output VO2 relationships during submaximal cycle ergometry [J].
Proctor, DN ;
Beck, KC ;
Shen, PH ;
Eickhoff, TJ ;
Halliwill, JR ;
Joyner, MJ .
JOURNAL OF APPLIED PHYSIOLOGY, 1998, 84 (02) :599-605
[25]   LUNG-VOLUMES AND FORCED VENTILATORY FLOWS - REPORT WORKING PARTY STANDARDIZATION OF LUNG-FUNCTION TESTS EUROPEAN-COMMUNITY FOR STEEL AND COAL - OFFICIAL STATEMENT OF THE EUROPEAN RESPIRATORY SOCIETY [J].
QUANJER, PH ;
TAMMELING, GJ ;
COTES, JE ;
PEDERSEN, OF ;
PESLIN, R ;
YERNAULT, JC .
EUROPEAN RESPIRATORY JOURNAL, 1993, 6 :5-40
[26]   A meta-analysis of three decades of validating thoracic impedance cardiography [J].
Raaijmakers, E ;
Faes, TJC ;
Scholten, RJPM ;
Goovaerts, HG ;
Heethaar, RM .
CRITICAL CARE MEDICINE, 1999, 27 (06) :1203-1213
[27]   Non-invasive cardiac output evaluation during a maximal progressive exercise test, using a new impedance cardiograph device [J].
Richard, R ;
Lonsdorfer-Wolf, E ;
Charloux, A ;
Doutreleau, S ;
Buchheit, M ;
Oswald-Mammosser, M ;
Lampert, E ;
Mettauer, B ;
Geny, B ;
Lonsdorfer, J .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2001, 85 (3-4) :202-207
[28]   Exercise training in chronic obstructive pulmonary disease [J].
Rochester, CL .
JOURNAL OF REHABILITATION RESEARCH AND DEVELOPMENT, 2003, 40 (05) :59-80
[29]   COMPARISON OF CARDIAC-OUTPUT ESTIMATION BY THORACIC ELECTRICAL BIOIMPEDANCE, THERMODILUTION, AND FICK METHODS [J].
SALANDIN, V ;
ZUSSA, C ;
RISICA, G ;
MICHIELON, P ;
PACCAGNELLA, A ;
CIPOLOTTI, G ;
SIMINI, G .
CRITICAL CARE MEDICINE, 1988, 16 (11) :1157-1158
[30]   Pulmonary vascular responses to hypoxia and hyperoxia in healthy volunteers and COPD patients measured by electrical impedance tomography [J].
Smit, HJ ;
Vonk-Noordegraaf, A ;
Marcus, JT ;
van der Weijden, S ;
Postmus, PE ;
de Vries, PMJM ;
Boonstra, A .
CHEST, 2003, 123 (06) :1803-1809