Combined Electrocardiography- and Respiratory-Triggered CT of the Lung to Reduce Respiratory Misregistration Artifacts between Imaging Slabs in Free-Breathing Children: Initial Experience

被引:19
作者
Goo, Hyun Woo [1 ,2 ]
Allmendinger, Thomas [3 ]
机构
[1] Univ Ulsan, Coll Med, Asan Med Ctr, Dept Radiol, 88 Olymp Ro,43 Gil, Seoul 05505, South Korea
[2] Univ Ulsan, Coll Med, Asan Med Ctr, Res Inst Radiol, 88 Olymp Ro,43 Gil, Seoul 05505, South Korea
[3] Siemens Healthcare GmbH, Computed Tomog Div, D-91301 Forchheim, Germany
关键词
Lung CT; ECG triggering; Respiratory triggering; Children; Motion artifacts; THIN-SECTION CT; HIGH-RESOLUTION CT; CHEST CT; MOTION ARTIFACTS; YOUNG-CHILDREN; INFANTS; FEASIBILITY; ACQUISITION; SCAN; MDCT;
D O I
10.3348/kjr.2017.18.5.860
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: Cardiac and respiratory motion artifacts degrade the image quality of lung CT in free-breathing children. The aim of this study was to evaluate the effect of combined electrocardiography (ECG) and respiratory triggering on respiratory misregistration artifacts on lung CT in free-breathing children. Materials and Methods: In total, 15 children (median age 19 months, range 6 months-8 years; 7 boys), who underwent free-breathing ECG-triggered lung CT with and without respiratory-triggering were included. A pressure-sensing belt of a respiratory gating system was used to obtain the respiratory signal. The degree of respiratory misregistration artifacts between imaging slabs was graded on a 4-point scale (1, excellent image quality) on coronal and sagittal images and compared between ECG-triggered lung CT studies with and without respiratory triggering. A p value < 0.05 was considered significant. Results: Lung CT with combined ECG and respiratory triggering showed significantly less respiratory misregistration artifacts than lung CT with ECG triggering only (1.1 +/- 0.4 vs. 2.2 +/- 1.0, p = 0.003). Conclusion: Additional respiratory-triggering reduces respiratory misregistration artifacts on ECG-triggered lung CT in free-breathing children.
引用
收藏
页码:860 / 866
页数:7
相关论文
共 28 条
[1]   Peak contrast enhancement in CT and MR angiography: When does it occur and why? Pharmacokinetic study in a porcine model [J].
Bae, KT .
RADIOLOGY, 2003, 227 (03) :809-816
[2]   Intraindividual comparison of image quality using retrospective and prospective respiratory gating for the acquisition of thin sliced four dimensional multidetector CT of the thorax in a porcine model [J].
Behzadi, Cyrus ;
Groth, Michael ;
Henes, Frank Oliver ;
Schwarz, Dorothee ;
Deibele, Andre ;
Begemann, Philipp G. C. ;
Adam, Gerhard ;
Regier, Marc .
EXPERIMENTAL LUNG RESEARCH, 2015, 41 (09) :489-498
[3]   Current trends in cardiac CT in children [J].
Goo, Hyun Woo .
ACTA RADIOLOGICA, 2013, 54 (09) :1055-1062
[4]   Advanced functional thoracic imaging in children: from basic concepts to clinical applications [J].
Goo, Hyun Woo .
PEDIATRIC RADIOLOGY, 2013, 43 (03) :262-268
[5]   State-of-the-art pediatric chest imaging [J].
Goo, Hyun Woo .
PEDIATRIC RADIOLOGY, 2013, 43 (03) :261-261
[6]   CT Radiation Dose Optimization and Estimation: an Update for Radiologists [J].
Goo, Hyun Woo .
KOREAN JOURNAL OF RADIOLOGY, 2012, 13 (01) :1-11
[7]   Individualized volume CT dose index determined by cross-sectional area and mean density of the body to achieve uniform image noise of contrast-enhanced pediatric chest CT obtained at variable kV levels and with combined tube current modulation [J].
Goo, Hyun Woo .
PEDIATRIC RADIOLOGY, 2011, 41 (07) :839-847
[8]   Coronary artery visibility in free-breathing young children with congenital heart disease on cardiac 64-slice CT: dual-source ECG-triggered sequential scan vs. single-source non-ECG-synchronized spiral scan [J].
Goo, Hyun Woo ;
Yang, Dong Hyun .
PEDIATRIC RADIOLOGY, 2010, 40 (10) :1670-1680
[9]   State-of-the-Art CT Imaging Techniques for Congenital Heart Disease [J].
Goo, Hyun Woo .
KOREAN JOURNAL OF RADIOLOGY, 2010, 11 (01) :4-18
[10]   Effects of high-resolution CT of the lung using partial versus full reconstruction on motion artifacts and image noise [J].
Ha, Hong Il ;
Goo, Hyun Woo ;
Seo, Joon Beom ;
Song, Jae-Woo ;
Lee, Jin Seong .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 187 (03) :618-622