A new X-ray detector for magnetic circular dichroism experiments

被引:4
作者
Bateman, JE [1 ]
Derbyshire, GE
Dudzik, E
van der Laan, G
Lipp, JD
Smith, AD
Stephenson, R
机构
[1] Rutherford Appleton Lab, CLRC, Instumentat Div, Didcot OX11 0QX, Oxon, England
[2] CLRC, Daresbury Lab, Warrington WA4 4AD, Cheshire, England
[3] Hahn Meitner Inst Kernforsch Berlin GmbH, D-14109 Berlin, Germany
关键词
gas microstrip; X-ray detectors; X-ray magnetic circular dichroism;
D O I
10.1016/S0168-9002(01)00566-6
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
X-ray magnetic circular dichroism (XMCD) studies of magnetic 3d transition metal samples require the recording of high quality absorption scans in high magnetic fields using circularly polarised soft X-rays of energies in the range 0.5-1 keV. A Gas Microstrip Detector is described which permits the option of using the X-ray fluorescence signal instead of the usual electron yield signal. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1140 / 1143
页数:4
相关论文
共 7 条
[1]  
[Anonymous], SYNCHR RAD NEWS
[2]  
Bateman J.E., RALTR1999057
[3]   THE DEVELOPMENT OF GAS MICROSTRIP DETECTORS FOR HIGH-ENERGY PHYSICS APPLICATIONS [J].
BATEMAN, JE ;
CONNOLLY, JF ;
STEPHENSON, R ;
EDWARDS, M ;
THOMPSON, JC .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 348 (2-3) :372-377
[4]  
BATEMAN JE, 1999, IN PRESS NUCL INSTR
[6]  
SAULI F, 7709 CERN, P31
[7]   STRONG MAGNETIC-X-RAY DICHROISM IN 2P ABSORPTION-SPECTRA OF 3D TRANSITION-METAL IONS [J].
VANDERLAAN, G ;
THOLE, BT .
PHYSICAL REVIEW B, 1991, 43 (16) :13401-13411