In-situ force plate calibration: 12 years' experience with an approach for correcting the point of force application

被引:3
作者
List, Renate [1 ]
Hitz, Marina [1 ]
Angst, Michael [1 ]
Taylor, William R. [1 ]
Lorenzetti, Silvio [1 ]
机构
[1] Swiss Fed Inst Technol, Inst Biomech, Leopold Ruzicka Weg 4,HCP H21-1, CH-8093 Zurich, Switzerland
关键词
Ground reaction force plates; In-situ calibration; Point of force application; RE-CALIBRATION; ACCURACY; PRESSURE; PLATFORMS;
D O I
10.1016/j.gaitpost.2017.07.111
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Force plates are common assessment tools used in biomechanics to measure ground reaction forces during motion capture or strength exercises. While the accuracy of the resulting kinetic data is critical for accurate gait analysis, factors such as plate mounting are thought to influence the measured point of force application (PFA) of the ground reaction forces. The aim of this study was therefore to present an accurate in-situ PFA calibration method. In addition, the approach was evaluated by examining the required PFA corrections over a period of ten years, which included plate remounting and a change of foundation. The in-situ PFA calibration was performed on six plates by assessing the locations of up to 98 points of force application per plate using an instrumented pole. Application of the in- situ PFA calibration approach reduced the root mean square errors by up to approximately 60% compared to the manufacturers calculation. Correction coefficients were strongly dependent on the individual platform as well as the location of the applied force on the plate. Remounting of the plates altered the plate coefficient corrections considerably, while changes over time were notable but not as extensive. We therefore recommend that plates should be recalibrated after remounting, but also at least every 5 years, in order to ensure the preservation of an accurate PFA.
引用
收藏
页码:98 / 102
页数:5
相关论文
共 19 条
[1]   ACCURACY OF DETERMINING THE POINT OF FORCE APPLICATION WITH PIEZOELECTRIC FORCE PLATES [J].
BOBBERT, MF ;
SCHAMHARDT, HC .
JOURNAL OF BIOMECHANICS, 1990, 23 (07) :705-710
[2]   CALCULATION OF VERTICAL GROUND REACTION FORCE ESTIMATES DURING RUNNING FROM POSITIONAL DATA [J].
BOBBERT, MF ;
SCHAMHARDT, HC ;
NIGG, BM .
JOURNAL OF BIOMECHANICS, 1991, 24 (12) :1095-1105
[3]   From "first" to "last" steps in life - Pressure patterns of three generations [J].
Bosch, Kerstin ;
Nagel, Arne ;
Weigend, Lars ;
Rosenbaum, Dieter .
CLINICAL BIOMECHANICS, 2009, 24 (08) :676-681
[4]   Periodical in-situ re-calibration of force platforms:: a new method for the robust estimation of the calibration matrix [J].
Cappello, A ;
Lenzi, D ;
Chiari, L .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2004, 42 (03) :350-355
[5]   A portable system for in-situ re-calibration of force platforms:: Theoretical validation [J].
Cedraro, Andrea ;
Cappello, Angelo ;
Chiari, Lorenzo .
GAIT & POSTURE, 2008, 28 (03) :488-494
[6]   A portable system for in-situ re-calibration of force platforms: Experimental validation [J].
Cedraro, Andrea ;
Cappello, Angelo ;
Chiari, Lorenzo .
GAIT & POSTURE, 2009, 29 (03) :449-453
[7]   A simple method for calibrating force plates and force treadmills using an instrumented pole [J].
Collins, Steven H. ;
Adamczyk, Peter G. ;
Ferris, Daniel P. ;
Kuo, Arthur D. .
GAIT & POSTURE, 2009, 29 (01) :59-64
[8]  
Dettwyler M., 2005, BIOMECHANISCHE UNTER
[9]   A new testing rig for force platform calibration and accuracy tests [J].
Gill, HS ;
OConnor, JJ .
GAIT & POSTURE, 1997, 5 (03) :228-232
[10]   Static in situ calibration of force plates [J].
Hall, MG ;
Fleming, HE ;
Dolan, MJ ;
Millbank, SFD ;
Paul, JP .
JOURNAL OF BIOMECHANICS, 1996, 29 (05) :659-665