Forces applied to the maxillary incisors by video laryngoscopes and the Macintosh laryngoscope

被引:46
作者
Lee, R. A. [2 ]
van Zundert, A. A. J. [1 ,3 ,4 ]
Maassen, R. L. J. G. [1 ,4 ]
Wieringa, P. A. [2 ]
机构
[1] Catharina Hosp, Brabant Med Sch, Dept Anesthesiol Intens Care & Pain Therapy, NL-5623 EJ Eindhoven, Netherlands
[2] Delft Univ Technol, Dept Biomech Engn, Delft, Netherlands
[3] Univ Ghent Hosp, Dept Anesthesiol, Fac Med, Ghent, Belgium
[4] Univ Hosp Maastricht, Dept Anesthesiol, Maastricht, Netherlands
关键词
PERIANESTHETIC DENTAL INJURIES; TRACHEAL INTUBATION; OBESE-PATIENTS; NORMAL AIRWAYS; STYLET USE; VIDEOLARYNGOSCOPE; GLIDESCOPE(R); ANESTHESIA; BLADE;
D O I
10.1111/j.1399-6576.2011.02541.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background Modern video laryngoscopes (VLSs) provide a superior view of the glottis, facilitating easier intubations. This study evaluates the forces applied to the maxillary incisors when using various VLSs and a Macintosh blade. Methods Fifty consecutive surgery patients were randomly assigned to receive laryngoscopy from a pair of four blades investigated in the study - the VLS GlideScope ® (Verathon Inc., Bothell, WA, USA), V-Mac™ Storz ® (Karl Storz, Tuttlingen, Germany), and McGrath™ (Aircraft Medical, Edinburgh, United Kingdom); and the classic Macintosh blade also from Storz ® (Karl Storz). An endotracheal tube (ETT) was brought into position anterior to the vocal cords, with actual intubation carried out only with the second of the laryngoscopes. Sensors measured the forces directly applied to the patient's maxillary incisors while inserting the ETT. Other common metrics of intubation difficulty (e.g. Mallampati grade, Cormack-Lehane grade, and time) were also recorded. Results Only one patient was not intubated within the standard study parameters and was converted to the hospital protocols for difficult intubations. The forces applied to the maxillary incisors were significantly greater with the Macintosh blade compared with all VLSs. There were no differences between the VLSs with regard to the forces. Patient characteristics, including Mallampati grade, were not predictive of the forces applied. Conclusions All VLSs considered were safer for the patient than was the Macintosh blade in terms of the forces applied to the maxillary teeth, time, number of insertion attempts, and view achieved of the glottic arch. There is a small, but significant, difference in the time and number of insertion attempts required during laryngoscopy with the different VLSs. There was no difference in the forces applied. The geometry of the respective blades may be an important component in the ease of laryngoscopy. © 2011 The Acta Anaesthesiologica Scandinavica Foundation.
引用
收藏
页码:224 / 229
页数:6
相关论文
共 19 条
[1]   Evaluation of the Video Intubation Unit in morbid obese patients [J].
Bathory, I. ;
Granges, J. -C ;
Frascarolo, P. ;
Magnusson, L. .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2010, 54 (01) :55-58
[2]   Transverse forces exerted on the maxillary incisors during laryngoscopy [J].
Bucx, MJL ;
vanderVegt, MH ;
Snijders, CJ ;
Stijnen, T ;
Wesselink, PR .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1996, 43 (07) :665-671
[3]  
Bucx MJL, 1996, ANESTH ANALG, V83, P1130, DOI 10.1097/00000539-199611000-00047
[4]   FORCES ACTING ON THE MAXILLARY INCISOR TEETH DURING LARYNGOSCOPY USING THE MACINTOSH LARYNGOSCOPE [J].
BUCX, MJL ;
SNIJDERS, CJ ;
VANGEEL, RTM ;
ROBERS, C ;
VANDEGIESSEN, H ;
ERDMANN, W ;
STIJNEN, T .
ANAESTHESIA, 1994, 49 (12) :1064-1070
[5]   Complications associated with the use of the GlideScope® videolaryngoscope [J].
Cooper, Richard M. .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2007, 54 (01) :54-57
[6]  
Cooper RM, 2005, CAN J ANAESTH, V52, P191, DOI 10.1007/BF03027728
[7]   Comparison of the GlideScope® video laryngoscope vs. the intubating laryngeal mask for females with normal airways [J].
Fun, W. L. L. ;
Lim, Y. ;
Teoh, W. H. L. .
EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2007, 24 (06) :486-491
[8]   Perianesthetic dental injuries: Analysis of incident reports [J].
Givol, N ;
Gershtansky, Y ;
Halamish-Shani, T ;
Taicher, S ;
Perel, A ;
Segal, E .
JOURNAL OF CLINICAL ANESTHESIA, 2004, 16 (03) :173-176
[9]   Forces Applied to the Maxillary Incisors During Video-Assisted Intubation [J].
Lee, Ruben A. ;
van Zundert, Andre A. J. ;
Maassen, Ralph L. J. G. ;
Willems, Remi J. ;
Beeke, Leon P. ;
Schaaper, Jan N. ;
van Dobbelsteen, Johan ;
Wieringa, Peter A. .
ANESTHESIA AND ANALGESIA, 2009, 108 (01) :187-191
[10]   A Comparison of Three Videolaryngoscopes: The Macintosh Laryngoscope Blade Reduces, but Does Not Replace, Routine Stylet Use for Intubation in Morbidly Obese Patients [J].
Maassen, Ralph ;
Lee, Ruben ;
Hermans, Boukje ;
Marcus, Marco ;
van Zundert, Andre .
ANESTHESIA AND ANALGESIA, 2009, 109 (05) :1560-1565