Initial characterization of the dosimetry and radiobiology of a device for administering interstitial stereotactic radiosurgery

被引:21
作者
Hakim, R
Zervas, NT
Hakim, F
Butler, WE
Beatty, J
Yanch, JC
Biggs, PJ
Gall, KP
Sliski, AP
机构
[1] MASSACHUSETTS GEN HOSP, NEUROSURG SERV, BOSTON, MA 02114 USA
[2] MASSACHUSETTS GEN HOSP, DEPT RADIAT MED, BOSTON, MA 02114 USA
[3] PHOTOELECTRON CORP, LEXINGTON, MA USA
[4] MIT, DEPT NUCL ENGN, CAMBRIDGE, MA 02139 USA
关键词
dosimetry; interstitial radiation; radiobiology; radiosurgery; stereotaxy;
D O I
10.1097/00006123-199703000-00016
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: We report the design and initial characterization of the dosimetry and radiobiology of a novel device for interstitial stereotactic radiosurgery. INSTRUMENTATION: The device is lightweight, handheld, and battery-powered, and it emits x-ray radiation from the tip of a probe 3 mm in diameter by 10 cm in length. METHODS: The dosimetry was characterized by two independent methods: thermoluminescent dosimeters and radiochromic film. The radiobiology was characterized by in vivo irradiation of rat liver, dog liver, and dog brain. The animals were killed at varying intervals of time, and histological examinations were performed. Heat transfer from the probe to dog brain was studied in vivo by placing thermocouple sensors around the probe tip before irradiating. RESULTS: Both dosimetric methods showed a steep dose-distance fall-off relationship (proportional to the reciprocal of the cube of the distance from the probe tip). Rats and dogs that were killed weeks to months after liver irradiation tended to have sharply demarcated lesions. Liver enzyme levels, measured serially in the dogs, did not give evidence of chronic inflammation. Histological examination of the brains of dogs that were killed acutely after irradiation did not show evidence of inflammation, edema, or hemorrhage. The tissue temperature elevation 1 cm from the tip never exceeded 0.5 degrees C, thereby excluding hyperthermia as a significant contributor to the formation of lesions. CONCLUSIONS: Because this device requires relatively few supporting resources, has sharp dosimetric properties, and seems to be safe, it may be useful as a clinical tool for interstitial stereotactic radiosurgery.
引用
收藏
页码:510 / 516
页数:7
相关论文
共 11 条
[1]   A new miniature x-ray device for interstitial radiosurgery: Dosimetry [J].
Beatty, J ;
Biggs, PJ ;
Gall, K ;
Okunieff, P ;
Pardo, FS ;
Harte, KJ ;
Dalterio, MJ ;
Sliski, AP .
MEDICAL PHYSICS, 1996, 23 (01) :53-62
[2]   A new miniature x-ray source for interstitial radiosurgery: Device description [J].
Dinsmore, M ;
Harte, KJ ;
Sliski, AP ;
Smith, DO ;
Nomikos, PM ;
Dalterio, MJ ;
Boom, AJ ;
Leonard, WF ;
Oettinger, PE ;
Yanch, JC .
MEDICAL PHYSICS, 1996, 23 (01) :45-52
[3]  
GALL K, 1993, MED PHYS, V20, P925
[4]   THE RADIOBIOLOGY OF RADIOSURGERY [J].
LARSON, DA ;
FLICKINGER, JC ;
LOEFFLER, JS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (03) :557-561
[5]  
LARSSON B, 1992, RADIOSURGERY, P3
[6]   STEREOTACTIC RADIOSURGERY [J].
LEKSELL, L .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1983, 46 (09) :797-803
[7]   STEREOTACTIC RADIOSURGERY - PRINCIPLES AND COMPARISON OF TREATMENT METHODS [J].
LUXTON, G ;
PETROVICH, Z ;
JOZSEF, G ;
NEDZI, LA ;
APUZZO, MLJ .
NEUROSURGERY, 1993, 32 (02) :241-259
[8]   SENSITOMETRY OF THE RESPONSE OF A NEW RADIOCHROMIC FILM DOSIMETER TO GAMMA-RADIATION AND ELECTRON-BEAMS [J].
MCLAUGHLIN, WL ;
YUNDONG, C ;
SOARES, CG ;
MILLER, A ;
VANDYK, G ;
LEWIS, DF .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 302 (01) :165-176
[9]   A COMPARISON OF SOLID PHANTOMS WITH WATER FOR DOSIMETRY OF I125 BRACHYTHERAPY SOURCES [J].
MEIGOONI, AS ;
MELI, JA ;
NATH, R .
MEDICAL PHYSICS, 1988, 15 (05) :695-701
[10]  
Ostertag C B, 1994, Recent Results Cancer Res, V135, P105