24-Hour IOP Telemetry in the Nonhuman Primate: Implant System Performance and Initial Characterization of IOP at Multiple Timescales

被引:117
作者
Downs, J. Crawford [1 ]
Burgoyne, Claude F. [2 ]
Seigfreid, William P. [1 ,3 ]
Reynaud, Juan F. [1 ,2 ]
Strouthidis, Nicholas G. [4 ,5 ]
Sallee, Verney [6 ]
机构
[1] Devers Eye Inst, Ocular Biomech Lab, Portland, OR 97232 USA
[2] Devers Eye Inst, Opt Nerve Head Res Lab, Portland, OR 97232 USA
[3] Univ Missouri, Columbia, MO USA
[4] Moorfields Eye Hosp, NIHR Natl Inst Hlth Research Biomed Sci Ctr, London, England
[5] UCL Inst Ophthalmol, London, England
[6] RTOP Consulting, Colorado Springs, CO USA
基金
美国国家卫生研究院;
关键词
OPEN-ANGLE GLAUCOMA; LOW-TENSION GLAUCOMA; INDEPENDENT RISK-FACTOR; OCULAR PULSE AMPLITUDE; INTRAOCULAR-PRESSURE; VISUAL-FIELD; NOCTURNAL ELEVATION; JAPANESE POPULATION; BARBADOS EYE; MONKEY EYES;
D O I
10.1167/iovs.11-7955
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. IOP is the most common independent risk factor for development and progression of glaucoma, but very little is known about IOP dynamics. Continuous IOP telemetry was used in three nonhuman primates to characterize IOP dynamics at multiple time scales for multiple 24-hour periods. METHODS. An existing implantable telemetric pressure transducer system was adapted to monitoring anterior chamber IOP. The system records 500 IOP, ECG, and body temperature measurements per second and compensates for barometric pressure in real time. The continuous IOP signal was digitally filtered for noise and dropout and reported using time-window averaging for 19, 18, and 4 24-hour periods in three animals, respectively. Those data were analyzed for a nycthemeral pattern within each animal. RESULTS. Ten-minute time-window averaging for multiple 24hour periods showed that IOP fluctuated from 7 to 14 mm Hg during the day, and those changes occurred frequently and quickly. Two-hour time-window averages of IOP for multiple 24-hour periods in three animals showed a weak nycthemeral trend, but IOP was not repeatable from day-to-day within animals. CONCLUSIONS. The measured IOP was successfully measured continuously by using a new, fully implantable IOP telemetry system. IOP fluctuates as much as 10 mm Hg from day to day and hour to hour in unrestrained nonhuman primates, which indicates that snapshot IOP measurements may be inadequate to capture the true dynamic character of IOP. The distributions, magnitudes, and patterns of IOP are not reproducible from day to day within animals, but IOP tends to be slightly higher at night when IOP data are averaged across multiple 24-hour periods within animals. (Invest Ophthalmol Vis Sci. 2011;52:7365-7375) DOI:10.1167/iovs.11-7955
引用
收藏
页码:7365 / 7375
页数:11
相关论文
共 48 条
[31]  
Markert Michael, 2007, Journal of Pharmacological and Toxicological Methods, V56, P203, DOI 10.1016/j.vascn.2007.03.004
[32]  
McLaren JW, 1996, INVEST OPHTH VIS SCI, V37, P966
[33]   Comparison of effects of topical ibopamine and epinephrine on the circadian rhythm of intraocular pressure of the rabbit eye as measured by telemetry [J].
McLaren, JW ;
Bachman, LA ;
Brubaker, RF .
JOURNAL OF OCULAR PHARMACOLOGY AND THERAPEUTICS, 1999, 15 (02) :107-116
[34]   Long-term intraocular pressure fluctuations and risk of conversion from ocular hypertension to glaucoma [J].
Medeiros, Felipe A. ;
Weinreb, Robert N. ;
Zangwill, Linda M. ;
Alencar, Luciana A. ;
Sample, Pamela A. ;
Vasile, Cristiana ;
Bowd, Christopher .
OPHTHALMOLOGY, 2008, 115 (06) :934-940
[35]   Predictive factors for open-angle glaucoma among patients with ocular hypertension in the European Glaucoma Prevention Study [J].
Miglior, Stefano .
OPHTHALMOLOGY, 2007, 114 (01) :3-9
[36]   Visual Field Progression in the Collaborative Initial Glaucoma Treatment Study The Impact of Treatment and Other Baseline Factors [J].
Musch, David C. ;
Gillespie, Brenda W. ;
Lichter, Paul R. ;
Niziol, Leslie M. ;
Janz, Nancy K. .
OPHTHALMOLOGY, 2009, 116 (02) :200-207
[37]  
Musch DC., Ophthalmology
[38]   Age related changes in intraocular pressure in a large Japanese population -: A cross-sectional and longitudinal study [J].
Nomura, H ;
Shimokata, H ;
Ando, F ;
Miyake, Y ;
Kuzuya, F .
OPHTHALMOLOGY, 1999, 106 (10) :2016-2022
[39]   The relationship between age and intraocular pressure in a Japanese population: The influence of central corneal thickness [J].
Nomura, H ;
Ando, F ;
Niino, N ;
Shimokata, H ;
Miyake, Y .
CURRENT EYE RESEARCH, 2002, 24 (02) :81-85
[40]   Physiological diurnal variability and characteristics of the ocular pulse amplitude (OPA) with the dynamic contour tonometer (DCT-Pascal®) [J].
Pourjavan S. ;
Boëlle P.-Y. ;
Detry-Morel M. ;
De Potter P. .
International Ophthalmology, 2007, 27 (6) :357-360